Atellica IM Folate 700 test kit (REF)-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 11206252
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI: (01)00630414293653(10)16097337(17)20230320 (01)00630414293653(10) 44861343 (17)20230630 (01)00630414293653(10) 63869345 (17)20230726 Lot Numbers: 16097337, 44861343, 63869345 and higher. Affects all current and future lots of the Atellica IM Folate assay until the Instructions for Use are updated.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Negative bias occurred when whole blood calibration (Atellica IM Fol or ADVIA Centaur FolateBA/FolBA) used to test serum samples.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Atellica IM Folate 700 test kit (REF)-For in vitro diagnostic use in the quantitative determination of folate in human serum and red blood cells. Siemens Material Number (SMN): 11206252
Product Codes/Lot Numbers:
UDI: (01)00630414293653(10)16097337(17)20230320 (01)00630414293653(10) 44861343 (17)20230630 (01)00630414293653(10) 63869345 (17)20230726 Lot Numbers: 16097337, 44861343, 63869345 and higher. Affects all current and future lots of the Atellica IM Folate assay until the Instructions for Use are updated.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1151-2023
Related Recalls
Siemens Dimension Tacrolimus Flex Reagent Cartridge (TAC), REF DF207, SMN 10700795; enzyme immunoassay
Siemens Healthcare Diagnostics
A potential positive bias in quality control (QC) and patient sample results in three lots of the Dimension Tacrolimus (TAC) Flex reagent cartridge.
Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.
Siemens Healthcare Diagnostics
Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results
Atellica CH Enzymatic Hemoglobin A1c (A1c_E); Siemens Material Number (SMN): 11097536;
Siemens Healthcare Diagnostics
A potential for depressed results for Atellica CH A1c_E when processing Atellica CH RCRP on the same analyzer. Customers processing A1c_E and RCRP on separate analyzers are not impacted by the recall. The frequency and magnitude of the negative bias and imprecision increase over time and are not consistent with the assay Instructions For Use (IFU). When this issue is present, Quality Control (QC) results demonstrate a negative trend and increased imprecision. The clinical impact of a negative (A1c_E) bias may lead to a delay in diagnosis for diabetes or lead to less-than-optimal treatment for existing diabetes.