LifeScan brand OneTouch¿ Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.

Class I - Dangerous
🏥 Medical Devices Recalled: February 25, 2013 Lifescan Diagnostic Equipment Nationwide

What Should You Do?

  1. Check if you have this product:
    2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Lifescan Inc
Reason for Recall:
When OneTouch¿ Ultra and OneTouch¿ Select test strips are tested with Ultra and Select control solution the meter may display a result outside the control solution test range printed on the strip vial label even though the system is working properly.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

LifeScan brand OneTouch¿ Select Control Solution, Part Number/Model #: 02168902. Testing of Blood Glucose Levels.

Product Codes/Lot Numbers:

2AA2101, Exp. Date: 9-30-2014; 2AA2102, Exp. Date: 10-31-2-14

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1130-2013

Related Recalls