Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model: Catalog #CT -99 Part number: 50137 Lot number(s): 13380, 13390 and 13418.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cincinnati Sub-Zero Products Inc
- Reason for Recall:
- Seals in the water path of the CT-99 cold therapy pad may separate during use, which could lead to a water leak.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cold Therapy Pad - cold water circulating pad, one unit per package. The pads are used to provide a patient with local cold therapy by circulating chilled water through the pad that has been positioned on the patient.
Product Codes/Lot Numbers:
Model: Catalog #CT -99 Part number: 50137 Lot number(s): 13380, 13390 and 13418.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1129-2015
Related Recalls
Due to FDA's safety communication relating to nontuberculous mycobacteria (NTM) infections related to heater/cooler devices, an updated Instructions For Use includes a more robust cleaning & disinfection procedure.
Hemotherm CE, model 400CE, Heater-Cooling Device
Cincinnati Sub-Zero Products
Improperly rated fuses were used in the manufacture of the device.
The Hemotherm 400CE 115v, containing power board p/n 39945
Cincinnati Sub-Zero Products
The device may not maintain water flow or temperature control.