ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- ulrich medical USA Inc
- Reason for Recall:
- Certain Torque Limiting Handles in the field are beyond their 3 year calibration life.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
ULRICH Torque Limiting Handle for use with a Set Screw Driver to perform the final tightening of set screws in a spinal implant construct.
Product Codes/Lot Numbers:
Model #s CS 7130 Serial #s 538436-02_70 (69 total) UDI Description 4052536123704, Model #s CS 7134 Serial #s 538438-02_70 (69 total) UDI Description 4052536123728, Model #s CS 7146 Serial #s 04-14-022-18_23 (6 total) UDI Description4052536078301, Model #s CS 7148 Serial #s 11-12-035-01_09, 03-12-006-24, and 03-12-006-25 (11 total) UDI Description 4052536078318, and Model #s CS 8049 Serial #s J04192-31_40, J04192-056, J04192-064, H12015-031_6, and J01276-031_3, (21 total) UDI Description 4052536022076.
Distribution:
Distributed in: AZ, IL, LA, MI, MO, NV, OH, PA, TN, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1126-2019
Related Recalls
Ulrich medical neon3 OCT spinal stabilization, REF CS 3917-12, Locking screw for crosslink, screw-to-screw
ulrich medical USA
Due to a production error, the weld seam of the "Locking screw for crosslink, screw-to-screw" cannot reliably transmit the required torque of 3.5 Nm during installation.
Surgical technique was revised to prevent static plate screws blocking mechanisms from disassociating.