Recall Details
- Company:
- Mevion Medical Systems, Inc.
- Reason for Recall:
- An error can occur causing Delta corrections to be lost when one setup field is closed and another is opened
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
MEVION S250-Proton Radiation Therapy Product Usage: Proton Radiation Therapy
Product Codes/Lot Numbers:
S250-0002 and S250-0003
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1122-2017
Related Recalls
Mevion S250 and S250i Proton Therapy Systems hand pendant (P/N 10263 or P/N 37808)
Mevion Medical Systems
Malfunction of the hand pendant controlling movement of the therapy couch and nozzle may result in unintentional motion. Patient injury is possible if they come in contact with the nozzle or other objects in the treatment room.
Treatment beam information disappears on Treatment Console screen while beam delivery continues potential for harm to a patient could occur
MEVION S250 Proton Radiation Treatment System; intended to deliver proton radiation treatment to patients
Mevion Medical Systems
The perforated screen that divides the inside of the process water tank has a very sharp edge across the top of the Heat Exchanger leading to potential injury for servicer.