Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Class I - DangerousWhat Should You Do?
- Check if you have this product: UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Beckman Coulter, Inc.
- Reason for Recall:
- A software error caused an increase in tray gripper motion errors that causes intermittent hard stops and a delay in patient results.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- completed
Product Information
Full Description:
Beckman Coulter DxI 9000 Access Immunoassay Analyzer, catalog number C11137; in-vitro diagnostic instrument
Product Codes/Lot Numbers:
UDI/DI 15099590732103, serial numbers: 300208, 300201, 300184, 300185, 300169, 300225, 300175, 300172, 300221, 300209, 300211, 300212, 300213, 300214, 300215, 300210, 300174, 300226, 300204, 300216, 300217, 300170, 300220, 300176, 300188, 300199, 300200, 300177, 300222, 300224, 300173, 300181, 300202, 300191, 300223, 300171, 300186, 300190, 300203, 300178, 300195, 300189, 300198, 300227, 300228, 300168, 300194, 300205, 300206, 300207, 300197, 300218, 300179, 300182, 300192, 300219, 300193, 300183, 300180, 300187, 300196.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1115-2025
Related Recalls
It has been determined that certain lots of Access 2 Reaction Vessels may contain manufacturing deformities. The bottom (round end) of the Reaction Vessel may have a small protrusion or hair like deformity. This could cause causing instrument errors (e.g., pipetting, rake, wash carousel errors) that could delay patient results.
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
Beckman Coulter UniCel DxI 600, Part Numbers: A71460, A30260; Photometric Chemistry Analyzer
Beckman Coulter
An internal investigation has determined that due to defective sub-components within the gantry assembly, a z-axis CAD movement motion error may occur and cause an analyzer to enter a Not Ready or Paused state, leading to a delay in obtaining results.