IntelliVue Patient Monitor MX500. Model Number: 866064
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model Number: 866064. UDI-DI: (01)00884838038776(21); Serial Numbers: DE7580A1R5, DE7580A1R4, DE7580A1R1, DE7580A1R0, DE7580A1PM, DE7580A1PL, DE7580A1PK, DE7580A1PJ, DE7580A1P5, DE351C8913, DE351C8912, DE351C8908, DE351C8911, DE351C8910, DE351C8909, DE32300881, DE32300882, DE32300883, DE32300937, DE32300936, DE32301099, DE35114607, DE35114606, DE671T0786, DE671K1669, DE671V9213, DE35133301, DE671N0625, DE671E0967, DE671E0964, DE671E0965, DE671E0968, DE671E0978, DE671E1009, DE671E1008, DE671E1007, DE671E1006, DE671E1005, DE671E1004, DE671E0977, DE671E1003, DE671E0979, DE671E1015, DE671E0966, DE671E1059.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips North America Llc
- Reason for Recall:
- Potential failure to alarm for "No Device Data" INOP. The device to alarm after a Hamilton ventilator is disconnected from the EC10 Module or IntelliBridge I/O board.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
IntelliVue Patient Monitor MX500. Model Number: 866064
Product Codes/Lot Numbers:
Model Number: 866064. UDI-DI: (01)00884838038776(21); Serial Numbers: DE7580A1R5, DE7580A1R4, DE7580A1R1, DE7580A1R0, DE7580A1PM, DE7580A1PL, DE7580A1PK, DE7580A1PJ, DE7580A1P5, DE351C8913, DE351C8912, DE351C8908, DE351C8911, DE351C8910, DE351C8909, DE32300881, DE32300882, DE32300883, DE32300937, DE32300936, DE32301099, DE35114607, DE35114606, DE671T0786, DE671K1669, DE671V9213, DE35133301, DE671N0625, DE671E0967, DE671E0964, DE671E0965, DE671E0968, DE671E0978, DE671E1009, DE671E1008, DE671E1007, DE671E1006, DE671E1005, DE671E1004, DE671E0977, DE671E1003, DE671E0979, DE671E1015, DE671E0966, DE671E1059.
Distribution:
Distributed in: US
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1109-2025
Related Recalls
Azurion R3.0. Model Number 722234. Commercial Name: Azurion 7 M20. INTERVENTIONAL FLUOROSCOPIC XRAY SYSTEM
Philips North America
Between July 2024 and June 2025, Philips Azurion R3.0 systems with a FlexArm stand were shipped without the Source-to-Skin Distance (SSD) Spacer. The SSD spacer is necessary to meet the FDA minimum source-skin distance requirement of 38 cm for standard fluoroscopy applications.
IntelliVue MP90. Product Number: M8010A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.
IntelliVue MP30. Product Number: M8002A.
Philips North America
Potential issue where the IntelliVue monitors did not alarm.