Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    161028133653JH, 161003111544KH, and 161007105036JG
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Immuno-Mycologics, Inc
Reason for Recall:
The affected products have approximately a 91% specificity while the non-affected lots have approximately a 99% specificity. The recalling firm is aware of three customers who reported false positive results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Cryptococcal Antigen Lateral Flow Assay (CrAg LFA), Reference # CR2003

Product Codes/Lot Numbers:

161028133653JH, 161003111544KH, and 161007105036JG

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1109-2017

Related Recalls

The firm found immunoassay products used to detect cryptococcal antigen in serum and cerebrospinal fluid to have reduced specificity (90% now versus 99% before). As a result, a small number of samples with positive test results may be false positives, which may cause some patients to initiate unnecessary anti-fungal therapy. Samples with negative test results are NOT affected.

Jan 3, 2022 Infusion Pumps Nationwide View Details →