NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) -Filling of pre-implant defects in conjunction with products intended for Guided Bone Regeneration
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot BMCXU11P1. BMCXU11H3 Lot BMCXU11P1, BMCXU12P1
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Collagen Matrix Inc
- Reason for Recall:
- During the 36 month real time shelf life testing of NuOss XC Sinus (BMCU-Umbrella) to verify the 3 year shelf life designated for the device, 3 of 13 products did not pass the recovery test, which requires that the compressed product fully expand within 1 minute of hydration time.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
NuOss XC Sinus Reference No 509-3001 (BMCU17, size 17 mm diameter x 10 mm height) Reference No 509-3002 (BMCU22, size 22 mm diameter x 12 m height) Manufacturer:- Collagen Matrix, Distributor:- ACE Surgical Supply Co, Inc, Intended use for dental surgery such as -augmentation or reconstructive treatment of alveoloar ridge -filling of infrabony periodontal defects -filling of defects after root resection, apicoectomy and cystectomy -filling of extraction sockets to enhance preservation of the alveolar ridge -elevation of maxillary sinus floor -filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR) -Filling of pre-implant defects in conjunction with products intended for Guided Bone Regeneration
Product Codes/Lot Numbers:
Lot BMCXU11P1. BMCXU11H3 Lot BMCXU11P1, BMCXU12P1
Distribution:
Distributed in: US, MA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1097-2016
Related Recalls
The packaging may not be sealed, which could compromise sterility.
DuraMatrix-Onlay Collagen Dura Substitute Membrane 4 inches x 5 inches (10.0cm x 12.5cm) REF: CDSLM45
Collagen Matrix
Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery
RESODURA matrix onlay 10.0 cm x 12.5 cm - indicated as dural substitutes for the repair of dura matter. REF: DUMO10125
Collagen Matrix
Outer packaging (outer pouch) is not sealed compromising sterility and result in potential risk of patient infection, which could lead to revision surgery