BladderScan BVI 9600, Model Numbers: 0270-0451, 0270-0452, 0270-0754, and U270-0451. AortaScan AMI 9700, Model Numbers: 0270-0636, 0270-0639, and U270-0636.
Class I - DangerousWhat Should You Do?
- Check if you have this product: SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE USA **************** A7001001, A7001002, A7001005, A7001007, A7001011, A7001022, A7001024, A7001031, A7001034, A7001036, A7001040, A7001041, A7001042, A7001043, A7001044, A7001045, A7001046, A7001047, A7001049, A7001050, A7001051, A7001052, A7001053, A7001054, A7001055, A7001056, A7001057, A7001060, A7001062, A7001065, A7001067, A7001068, A7001079, A7001080, A7001086, A7001088, A7001093, A7001094, A7001098, A7001101, A7001116, A7001117, A7001118, A7001119, A7001120, A7001121, A7001122, A7001123, A7001124, A7001125, A7001126, A7001127, A7001128, A7001129, A7001130, A7001131, A7001132, A7001138, A7001139, A7001140, A7001141, A7001143, A7001144, A7001146, A7001147, A7001148, A7001150, A7001151, A7001152, A7001153, A7001155, A7001156, A7001157, A7001158, A7001161, A7001162, A7001163, A7001164, A7001165, A7001168, A7001169, A7001170, A7001171, A7001172, A7001173, A7001174, A7001175, A7001176, A7001177, A7001178, A7001179, A7001180, A7001181, A7001182, A7001183, A7001184, A7001185, A7001186, A7001187, A7001188, A7001189, A7001190, A7001191, A7001192, A7001194, A7001195, A7001196, A7001197, A7001198, A7001199, A7001201, A7001202, A7001204, A7001205, A7001206, A7001207, A7001209, A7001210, A7001211, A7001212, A7001216, A7001217, A7001218, A7001219, A7001220, A7001221, A7001222, A7001223, A7001224, A7001226, A7001227, A7001228, A7001229, A7001231, A7001232, A7001233, A7001234, A7001235, A7001236, A7001237, A7001238, A7001239, A7001240, A7001241, A7001242, A7001243, A7001244, A7001246, A7001247, A7001248, A7001249, A7001250, A7001251, A7001252, A7001253, A7001254, A7001255, A7001256, A7001257, A7001258, A7001260, A7001261, A7001262, A7001264, A7001265, A7001266, A7001268, A7001269, A7001270, A7001271, A7001272, A7001273, A7001275, A7001276, A7001277, A7001278, A7001279, A7001281, A7001283, A7001284, A7001285, A7001286, A7001287, A7001288, A7001289, A7001290, A7001291, A7001292, A7001293, A7001294, A7001295, A7001296, A7001297, A7001298, A7001299, A7001300, A7001301, A7001302, A7001303, A7001305, A7001306, A7001307, A7001308, A7001312, A7001313, A7001314, A7001315, A7001316, A7001317, A7001318, A7001319, A7001320, A7001321, A7001322, A7001323, A7001324, A7001325, A7001326, A7001327, A7001328, A7001329, A7001330, A7001331, A7001332, A7001333, A7001334, A7001335, A7001336, A7001337, A7001338, A7001339, A7001340, A7001341, A7001342, A7001343, A7001344, A7001345, A7001347, A7001348, A7001349, A7001350, A7001351, A7001352, A7001353, A7001354, A7001355, A7001356, A7001357, A7001358, A7001359, A7001360, A7001361, A7001362, A7001363, A7001364, A7001365, A7001366, A7001367, A7001369, A7001370, A7001371, A7001372, A7001373, A7001375, A7001376, A7001377, A7001378, A7001380, A7001381, A7001382, A7001383, A7001384, A7001385, A7001386, A7001387, A7001388, A7001389, A7001390, A7001391, A7001392, A7001393, A7001394, A7001395, A7001396, A7001397, A7001398, A7001399, A7001400, A7001401, A7001402, A7001403, A7001404, A7001405, A7001406, A7001408, A7001409, A7001410, A7001411, A7001412, A7001413, A7001414, A7001415, A7001416, A7001418, A7001419, A7001420, A7001421, A7001422, A7001423, A7001424, A7001425, A7001426, A7001427, A7001428, A7001429, A7001430, A7001431, A7001432, A7001433, A7001434, A7001435, A7001436, A7001437, A7001438, A7001439, A7001440, A7001441, A7001445, A7001446, A7001447, A7001448, A7001449, A7001450, A7001451, A7001452, A7001453, A7001454, A7001455, A7001456, A7001457, A7001458, A7001459, A7001460, A7001461, A7001462, A7001463, A7001464, A7001465, A7001466, A7001467, A7001468, A7001469, A7001470, A7001471, A7001472, A7001473, A7001474, A7001475, A7001476, A7001477, A7001478, A7001479, A7001480, A7001482, A7001483, A7001484, A7001485, A7001486, A7001487, A7001488, A7001489, A7001490, A7001491, A7001492, A7001493, A7001494, A7001495, A7001496, A7001498, A7001499, A7001500, A7001502, A7001503, A7001504, A7001505, A7001506, A7001507, A7001508, A7001509, A7001510, A7001511, A7001512, A7001513, A7001514, A7001515, A7001516, A7001517, A7001518, A7001519, A7001520, A7001521, A7001522, A7001523, A7001524, A7001525, A7001526, A7001527, A7001528, A7001529, A7001530, A7001531, A7001532, A7001533, A7001534, A7001535, A7001536, A7001537, A7001538, A7001539, A7001540, A7001541, A7001542, A7001543, A7001544, A7001545, A7001546, A7001547, A7001548, A7001549, A7001550, A7001551, A7001552, A7001553, A7001554, A7001555, A7001556, A7001557, A7001558, A7001559, A7001560, A7001561, A7001562, A7001563, A7001564, A7001565, A7001566, A7001567, A7001568, A7001569, A7001570, A7001571, A7001572, A7001573, A7001574, A7001575, A7001576, A7001577, A7001578, A7001579, A7001580, A7001581, A7001582, A7001583, A7001584, A7001585, A7001586, A7001587, A7001588, A7001589, A7001590, A7001591, A7001592, A7001593, A7001594, A7001595, A7001596, A7001597, A7001598, A7001599, A7001600, A7001601, A7001602, A7001603, A7001604, A7001605, A7001606, A7001607, A7001608, A7001609, A7001610, A7001611, A7001612, A7001613, A7001614, A7001615, A7001616, A7001617, A7001618, A7001619, A7001620, A7001621, A7001622, A7001623, A7001624, A7001625, A7001626, A7001628, A7001629, A7001630, A7001631, A7001632, A7001633, A7001634, A7001635, A7001636, A7001637, A7001638, A7001639, A7001640, A7001641, A7001642, A7001643, A7001644, A7001645, A7001646, A7001647, A7001648, A7001649, A7001650, A7001651, A7001652, A7001653, A7001654, A7001655, A7001656, A7001657, A7001658, A7001659, A7001660, A7001661, A7001662, A7001663, A7001664, A7001665, A7001666, A7001667, A7001668, A7001669, A7001670, A7001671, A7001672, A7001673, A7001674, A7001675, A7001676, A7001677, A7001678, A7001679, A7001681, A7001683, A7001684, A7001685, A7001686, A7001687, A7001688, A7001689, A7001690, A7001691, A7001692, A7001693, A7001694, A7001695, A7001697, A7001698, A7001699, A7001700, A7001701, A7001702, A7001703, A7001704, A7001705, A7001706, A7001707, A7001708, A7001709, A7001710, A7001711, A7001712, A7001713, A7001714, A7001715, A7001716, A7001717, A7001718, A7001719, A7001721, A7001722, A7001723, A7001724, A7001725, A7001726, A7001727, A7001729, A7001730, A7001732, A7001733, A7001734, A7001735, A7001736, A7001737, A7001738, A7001739, A7001740, A7001741, A7001742, A7001743, A7001744, A7001745, A7001746, A7001747, A7001748, A7001749, A7001752, A7001753, A7001754, A7001755, A7001756, A7001757, A7001758, A7001759, A7001760, A7001761, A7001762, A7001763, A7001765, A7001766, A7001767, A7001768, A7001769, A7001770, A7001771, A7001780, A7001781, A7001782, A7001783, A7001784, A7001785, A7001786, A7001787, A7001788, A7001789, A7001790, A7001791, A7001792, A7001793, A7001794, A7001795, A7001800, A7001801, A7001802, A7001803, A7001804, A7001805, A7001806, A7001807, A7001808, A7001809, A7001810, A7001811, A7001812, A7001813, A7001814, A7001815, A7001816, A7001817, A7001819, A7001820, A7001821, A7001822, A7001823, A7001824, A7001825, A7001826, A7001827, A7001828, A7001836, A7001837, A7001841, A7001842, A7001843, A7001844, A7001845, A7001846, A7001847, A7001848, A7001857, A7001858, A7001859, A7001860, A7001861, A7001862, A7001863, A7001864, A7001865, A7001866, A7001867, A7001868, A7001869, A7001870, A7001871, A7001872, A7001873, A7001874, A7001876, A7001877, A7001878, A7001879, A7001880, A7001881, A7001882, A7001883, A7001884, A7001885, A7001886, A7001887, A7001888, A7001889, A7001890, A7001891, A7001892, A7001893, A7001894, A7001895, A7001897, A7001898, A7001899, A7001900, A7001901, A7001902, A7001903, A7001904, A7001905, A7001906, A7001907, A7001908, A7001909, A7001910, A7001911, A7001912, A7001913, A7001914, A7001915, A7001916, A7001917, A7001918, A7001919, A7001920, A7001921, A7001922, A7001923, A7001924, A7001925, A7001927, A7001928, A7001929, A7001930, A7001931, A7001932, A7001933, A7001934, A7001935, A7001936, A7001937, A7001938, A7001939, A7001940, A7001941, A7001942, A7001943, A7001944, A7001945, A7001946, A7001947, A7001948, A7001949, A7001950, A7001951, A7001952, A7001953, A7001954, A7001955, A7001956, A7001957, A7001958, A7001959, A7001960, A7001961, A7001962, A7001963, A7001964, A7001965, A7001966, A7001967, A7001968, A7001969, A7001970, A7001971, A7001972, A7001973, A7001981, A7300000, A7300001, A7300002, A7300003, A7300004, A7300005, A7300006, A7300007, A7300008, A7300012, B6001016, B6001020, B6001023, B6001024, B6001025, B6001026, B6001030, B6001040, B6001041, B6001043, B6001047, B6001049, B6001060, B6001065, B6001066, B6001068, B6001069, B6001070, B6001071, B6001073, B6001080, B6001082, B6001084, B6001085, B6001086, B6001088, B6001090, B6001092, B6001093, B6001094, B6001095, B6001096, B6001097, B6001098, B6001100, B6001102, B6001103, B6001104, B6001107, B6001108, B6001109, B6001110, B6001111, B6001112, B6001113, B6001123, B6001124, B6001130, B6001131, B6001133, B6001138, B6001140, B6001141, B6001143, B6001148, B6001157, B6001160, B6001164, B6001165, B6001166, B6001167, B6001168, B6001169, B6001170, B6001171, B6001172, B6001173, B6001174, B6001175, B6001176, B6001177, B6001178, B6001179, B6001180, B6001181, B6001183, B6001184, B6001185, B6001186, B6001187, B6001198, B6001199, B6001202, B6001203, B6001204, B6001205, B6001206, B6001207, B6001208, B6001209, B6001212, B6001213, B6001216, B6001217, B6001218, B6001219, B6001220, B6001221, B6001223, B6001225, B6001227, B6001231, B6001232, B6001234, B6001237, B6001238, B6001239, B6001241, B6001242, B6001243, B6001244, B6001245, B6001246, B6001254, B6001256, B6001257, B6001258, B6001259, B6001260, B6001261, B6001263, B6001265, B6001267, B6001269, B6001270, B6001271, B6001272, B6001273, B6001276, B6001277, B6001278, B6001279, B6001280, B6001281, B6001282, B6001283, B6001284, B6001285, B6001286, B6001287, B6001288, B6001298, B6001299, B6001305, B6001306, B6001307, B6001308, B6001309, B6001310, B6001311, B6001312, B6001313, B6001314, B6001315, B6001316, B6001317, B6001318, B6001319, B6001320, B6001321, B6001322, B6001323, B6001324, B6001325, B6001326, B6001327, B6001328, B6001329, B6001330, B6001331, B6001332, B6001339, B6001341, B6001342, B6001343, B6001345, B6001346, B6001348, B6001349, B6001350, B6001351, B6001354, B6001355, B6001356, B6001357, B6001358, B6001359, B6001360, B6001361, B6001362, B6001363, B6001364, B6001365, B6001366, B6001367, B6001369, B6001370, B6001372, B6001373, B6001374, B6001375, B6001377, B6001378, B6001379, B6001381, B6001382, B6001384, B6001386, B6001387, B6001388, B6001389, B6001390, B6001396, B6001397, B6001398, B6001399, B6001401, B6001402, B6001404, B6001406, B6001407, B6001408, B6001409, B6001410, B6001411, B6001419, B6001420, B6001421, B6001422, B6001423, B6001424, B6001425, B6001426, B6001427, B6001428, B6001429, B6001431, B6001435, B6001436, B6001437, B6001438, B6001439, B6001440, B6001441, B6001442, B6001443, B6001444, B6001445, B6001448, B6001449, B6001450, B6001451, B6001452, B6001453, B6001455, B6001456, B6001457, B6001459, B6001461, B6001462, B6001467, B6001468, B6001469, B6001470, B6001471, B6001472, B6001473, B6001474, B6001477, B6001478, B6001479, B6001480, B6001481, B6001482, B6001483, B6001484, B6001485, B6001486, B6001487, B6001488, B6001489, B6001490, B6001491, B6001492, B6001493, B6001494, B6001495, B6001496, B6001497, B6001498, B6001499, B6001500, B6001501, B6001502, B6001503, B6001504, B6001505, B6001506, B6001507, B6001508, B6001509, B6001510, B6001511, B6001512, B6001514, B6001515, B6001516, B6001518, B6001519, B6001520, B6001521, B6001522, B6001523, B6001524, B6001525, B6001526, B6001529, B6001530, B6001531, B6001532, B6001533, B6001535, B6001537, B6001541, B6001542, B6001544, B6001545, B6001546, B6001547, B6001549, B6001550, B6001553, B6001554, B6001557, B6001558, B6001559, B6001560, B6001561, B6001562, B6001563, B6001564, B6001565, B6001567, B6001568, B6001569, B6001570, B6001571, B6001572, B6001573, B6001574, B6001575, B6001576, B6001577, B6001578, B6001579, B6001580, B6001581, B6001582, B6001583, B6001584, B6001585, B6001586, B6001587, B6001588, B6001589, B6001590, B6001591, B6001592, B6001593, B6001594, B6001595, B6001596, B6001597, B6001598, B6001599, B6001600, B6001601, B6001602, B6001603, B6001604, B6001605, B6001606, B6001608, B6001609, B6001610, B6001611, B6001612, B6001613, B6001614, B6001615, B6001616, B6001617, B6001618, B6001619, B6001620, B6001621, B6001622, B6001623, B6001624, B6001627, B6001628, B6001629, B6001630, B6001631, B6001632, B6001633, B6001634, B6001635, B6001636, B6001637, B6001638, B6001639, B6001640, B6001641, B6001642, B6001643, B6001644, B6001645, B6001646, B6001647, B6001648, B6001660, B6001661, B6001662, B6001663, B6001664, B6001665, B6001666, B6001667, B6001668, B6001669, B6001670, B6001672, B6001673, B6001674, B6001675, B6001676, B6001677, B6001678, B6001679, B6001680, B6001681, B6001682, B6001683, B6001685, B6001686, B6001687, B6001688, B6001689, B6001691, B6001692, B6001693, B6001694, B6001695, B6001696, B6001697, B6001698, B6001699, B6001700, B6001701, B6001702, B6001703, B6001704, B6001705, B6001707, B6001708, B6001709, B6001710, B6001711, B6001712, B6001713, B6001715, B6001716, B6001717, B6001718, B6001719, B6001720, B6001721, B6001722, B6001723, B6001724, B6001725, B6001726, B6001727, B6001728, B6001729, B6001730, B6001731, B6001733, B6001734, B6001735, B6001736, B6001738, B6001739, B6001740, B6001741, B6001742, B6001743, B6001744, B6001746, B6001747, B6001748, B6001750, B6001751, B6001752, B6001753, B6001754, B6001755, B6001756, B6001757, B6001758, B6001759, B6001761, B6001762, B6001764, B6001765, B6001766, B6001767, B6001768, B6001769, B6001770, B6001771, B6001772, B6001773, B6001774, B6001776, B6001777, B6001778, B6001779, B6001780, B6001781, B6001782, B6001783, B6001787, B6001788, B6001790, B6001791, B6001793, B6001795, B6001796, B6001797, B6001798, B6001799, B6001800, B6001801, B6001803, B6001805, B6001806, B6001807, B6001808, B6001809, B6001810, B6001811, B6001812, B6001813, B6001814, B6001815, B6001816, B6001817, B6001818, B6001819, B6001820, B6001821, B6001822, B6001824, B6001825, B6001826, B6001827, B6001828, B6001829, B6001830, B6001831, B6001832, B6001833, B6001834, B6001835, B6001836, B6001838, B6001839, B6001840, B6001841, B6001842, B6001843, B6001844, B6001845, B6001846, B6001848, B6001849, B6001850, B6001851, B6001852, B6001853, B6001854, B6001855, B6001856, B6001857, B6001858, B6001859, B6001860, B6001861, B6001862, B6001863, B6001864, B6001865, B6001866, B6001867, B6001868, B6001869, B6001870, B6001871, B6001872, B6001873, B6001874, B6001875, B6001876, B6001877, B6001878, B6001879, B6001880, B6001881, B6001882, B6001883, B6001884, B6001885, B6001887, B6001888, B6001889, B6001890, B6001891, B6001892, B6001893, B6001894, B6001895, B6001896, B6001897, B6001898, B6001899, B6001900, B6001901, B6001902, B6001903, B6001904, B6001905, B6001906, B6001907, B6001908, B6001909, B6001910, B6001911, B6001912, B6001913, B6001914, B6001915, B6001916, B6001917, B6001918, B6001919, B6001920, B6001921, B6001922, B6001923, B6001924, B6001925, B6001926, B6001927, B6001928, B6001930, B6001931, B6001932, B6001934, B6001935, B6001936, B6001937, B6001938, B6001939, B6001940, B6001941, B6001942, B6001943, B6001944, B6001945, B6001946, B6001947, B6001948, B6001949, B6001950, B6001951, B6001952, B6001953, B6001954, B6001955, B6001956, B6001957, B6001958, B6001959, B6001960, B6001961, B6001962, B6001963, B6001964, B6001965, B6001966, B6001967, B6001968, B6001969, B6001972, B6001973, B6001974, B6001975, B6001976, B6001977, B6001978, B6001979, B6001980, B6001981, B6001982, B6001983, B6001984, B6001986, B6001987, B6001988, B6001989, B6001990, B6001991, B6001992, B6001993, B6001994, B6001995, B6001996, B6001997, B6001998, B6001999, B6002000, B6002001, B6002002, B6002003, B6002004, B6002005, B6002007, B6002008, B6002009, B6002010, B6002011, B6002012, B6002013, B6002014, B6002015, B6002016, B6002017, B6002018, B6002019, B6002020, B6002021, B6002022, B6002023, B6002024, B6002025, B6002026, B6002027, B6002028, B6002029, B6002030, B6002031, B6002032, B6002033, B6002034, B6002035, B6002036, B6002037, B6002038, B6002039, B6002040, B6002041, B6002042, B6002043, B6002044, B6002045, B6002047, B6002048, B6002049, B6002050, B6002051, B6002052, B6002054, B6002055, B6002057, B6002058, B6002059, B6002060, B6002061, B6002062, B6002063, B6002064, B6002065, B6002066, B6002067, B6002068, B6002069, B6002070, B6002071, B6002072, B6002073, B6002074, B6002075, B6002076, B6002077, B6002078, B6002079, B6002080, B6002081, B6002082, B6002083, B6002084, B6002085, B6002086, B6002087, B6002088, B6002089, B6002090, B6002091, B6002092, B6002093, B6002094, B6002096, B6002097, B6002098, B6002099, B6002100, B6002101, B6002102, B6002103, B6002104, B6002105, B6002106, B6002107, B6002108, B6002109, B6002110, B6002111, B6002112, B6002113, B6002114, B6002115, B6002116, B6002117, B6002118, B6002119, B6002120, B6002121, B6002122, B6002123, B6002124, B6002125, B6002126, B6002127, B6002128, B6002129, B6002130, B6002131, B6002132, B6002133, B6002134, B6002135, B6002136, B6002137, B6002138, B6002139, B6002140, B6002141, B6002142, B6002143, B6002145, B6002147, B6002148, B6002149, B6002150, B6002151, B6002152, B6002153, B6002154, B6002155, B6002156, B6002157, B6002158, B6002159, B6002160, B6002161, B6002162, B6002163, B6002164, B6002168, B6002169, B6002170, B6002171, B6002172, B6002173, B6002174, B6002175, B6002176, B6002177, B6002178, B6002179, B6002180, B6002181, B6002182, B6002183, B6002184, B6002185, B6002186, B6002187, B6002188, B6002189, B6002190, B6002191, B6002192, B6002193, B6002195, B6002196, B6002197, B6002198, B6002199, B6002200, B6002201, B6002202, B6002203, B6002204, B6002205, B6002206, B6002207, B6002208, B6002209, B6002210, B6002211, B6002212, B6002213, B6002214, B6002215, B6002216, B6002217, B6002218, B6002219, B6002220, B6002221, B6002222, B6002223, B6002224, B6002225, B6002226, B6002227, B6002228, B6002229, B6002230, B6002231, B6002232, B6002233, B6002234, B6002235, B6002236, B6002237, B6002238, B6002239, B6002240, B6002241, B6002242, B6002243, B6002244, B6002245, B6002246, B6002247, B6002248, B6002251, B6002252, B6002253, B6002254, B6002255, B6002256, B6002257, B6002258, B6002259, B6002260, B6002261, B6002262, B6002263, B6002264, B6002265, B6002266, B6002267, B6002268, B6002269, B6002270, B6002271, B6002272, B6002273, B6002274, B6002275, B6002276, B6002277, B6002278, B6002279, B6002280, B6002281, B6002282, B6002283, B6002284, B6002285, B6002286, B6002287, B6002288, B6002289, B6002292, B6002293, B6002294, B6002295, B6002296, B6002297, B6002298, B6002299, B6002300, B6002301, B6002302, B6002303, B6002304, B6002305, B6002306, B6002307, B6002308, B6002309, B6002310, B6002311, B6002313, B6002314, B6002315, B6002316, B6002317, B6002318, B6002319, B6002320, B6002321, B6002322, B6002323, B6002324, B6002325, B6002326, B6002327, B6002328, B6002329, B6002331, B6002332, B6002334, B6002335, B6002336, B6002337, B6002338, B6002339, B6002340, B6002341, B6002342, B6002343, B6002345, B6002346, B6002347, B6002348, B6002349, B6002350, B6002351, B6002352, B6002353, B6002354, B6002356, B6002357, B6002358, B6002359, B6002360, B6002361, B6002362, B6002363, B6002364, B6002365, B6002366, B6002367, B6002368, B6002369, B6002370, B6002371, B6002373, B6002374, B6002375, B6002376, B6002377, B6002378, B6002379, B6002381, B6002382, B6002383, B6002384, B6002385, B6002386, B6002387, B6002388, B6002389, B6002390, B6002391, B6002392, B6002393, B6002394, B6002395, B6002396, B6002398, B6002399, B6002401, B6002402, B6002403, B6002404, B6002405, B6002406, B6002407, B6002408, B6002411, B6002412, B6002413, B6002414, B6002415, B6002416, B6002417, B6002418, B6002419, B6002420, B6002421, B6002422, B6002423, B6002424, B6002425, B6002426, B6002427, B6002428, B6002429, B6002430, B6002431, B6002432, B6002433, B6002434, B6002435, B6002436, B6002437, B6002438, B6002439, B6002440, B6002441, B6002442, B6002443, B6002444, B6002445, B6002446, B6002447, B6002448, B6002449, B6002451, B6002452, B6002457, B6002458, B6002459, B6002460, B6002461, B6002462, B6002463, B6002469, B6002470, B6002471, B6002472, B6002473, B6002474, B6002475, B6002476, B6002477, B6002478, B6002479, B6002480, B6002481, B6002482, B6002483, B6002484, B6002485, B6002486, B6002487, B6002488, B6002489, B6002490, B6002491, B6002492, B6002493, B6002494, B6002495, B6002496, B6002504, B6002505, B6002506, B6002507, B6002508, B6002509, B6002510, B6002511, B6002512, B6002513, B6002514, B6002515, B6002519, B6002520, B6002521, B6002522, B6002524, B6002525, B6002526, B6002528, B6002529, B6002530, B6002531, B6002534, B6002535, B6002536, B6002542, B6002543, B6002544, B6002545, B6002546, B6002547, B6002548, B6002549, B6002550, B6002551, B6002553, B6002554, B6002555, B6002556, B6002557, B6002558, B6002559, B6002560, B6002561, B6002562, B6002563, B6002564, B6002565, B6002566, B6002602, B6002667, B6002668, B6002669, B6002670, B6002671, B6002672, B6002673, B6002674, B6002675, B6002676, B6002677, B6002678, B6002679, B6002680, B6002681, B6002682, B6002683, B6002684, B6002685, B6002686, B6002687, B6002688, B6002689, B6002690, B6002691, B6002692, B6002693, B6002694, B6002695, B6002696, B6002697, B6002698, B6002700, B6002701, B6002702, B6002703, B6002704, B6002705, B6002707, B6002708, B6002709, B6002710, B6002711, B6002712, B6002713, B6002714, B6002715, B6002716, B6002717, B6002718, B6002719, B6002720, B6002722, B6002723, B6002724, B6002725, B6002726, B6002727, B6002728, B6002729, B6002730, B6002732, B6002733, B6002734, B6002735, B6002736, B6002737, B6002746, B6002747, B6002748, B6002749, B6002750, B6002751, B6002752, B6002753, B6002754, B6002755, B6002756, B6002757, B6002758, B6002759, B6002760, B6002761, B6002762, B6002763, B6002764, B6002765, B6002766, B6002767, B6002768, B6002769, B6002770, B6002771, B6002772, B6002773, B6002774, B6002775, B6002777, B6002778, B6002779, B6002781, B6002782, B6002783, B6002784, B6002785, B6002786, B6002787, B6002788, B6002789, B6002790, B6002791, B6002792, B6002793, B6002794, B6002795, B6002796, B6002797, B6002798, B6002799, B6002800, B6002801, B6002803, B6002805, B6002806, B6002807, B6002808, B6002809, B6002810, B6002811, B6002812, B6002813, B6002815, B6002816, B6002817, B6002818, B6002819, B6002820, B6002821, B6002822, B6002823, B6002824, B6002825, B6002826, B6002827, B6002828, B6002829, B6002830, B6002831, B6002832, B6002833, B6002834, B6002835, B6002836, B6002837, B6002838, B6002839, B6002840, B6002841, B6002842, B6002843, B6002844, B6002845, B6002849, B6002857, B6002858, B6002859, B6002860, B6002861, B6002862, B6002863, B6002864, B6002865, B6002866, B6002867, B6002868, B6002869, B6002870, B6002871, B6002872, B6002873, B6002874, B6002875, B6002876, B6002878, B6002879, B6002880, B6002881, B6002882, B6002883, B6002884, B6002885, B6002886, B6002888, B6002889, B6002890, B6002891, and B6002892. ****SERIAL NUMBERS OF UNITS DISTRIBUTED OUTSIDE THE USA***** 00001001, 00001007, 00001008, A7001048, A7001069, A7001071, A7001089, A7001090, A7001091, A7001092, A7001095, A7001096, A7001097, A7001100, A7001102, A7001103, A7001106, A7001107, A7001108, A7001109, A7001110, A7001111, A7001112, A7001113, A7001114, A7001115, A7001133, A7001136, A7001263, A7001280, A7001282, A7001442, A7001443, A7001444, A7001720, A7001796, A7001797, A7001798, A7001799, A7001829, A7001830, A7001831, A7001832, A7001833, A7001834, A7001835, A7001838, A7001839, A7001840, A7001849, A7001850, A7001851, A7001852, A7001853, A7001854, A7001855, A7001856, B6001063, B6001064, B6001075, B6001079, B6001081, B6001188, B6001189, B6001211, B6001224, B6001233, B6001251, B6001289, B6001294, B6001296, B6001300, B6001337, B6001347, B6001353, B6001368, B6001371, B6001380, B6001383, B6001391, B6001393, B6001394, B6001395, B6001405, B6001412, B6001413, B6001416, B6001418, B6001432, B6001446, B6001447, B6001454, B6001465, B6001466, B6001475, B6001476, B6001517, B6001534, B6001539, B6001540, B6001548, B6001551, B6001552, B6001556, B6001607, B6001625, B6001671, B6001684, B6001690, B6001714, B6001732, B6001745, B6001749, B6001760, B6001763, B6001775, B6001784, B6001785, B6001786, B6001789, B6001792, B6001802, B6001847, B6001886, B6001929, B6001933, B6001970, B6001971, B6001985, B6002053, B6002056, B6002166, B6002167, B6002194, B6002250, B6002290, B6002291, B6002330, B6002333, B6002344, B6002372, B6002409, B6002410, B6002450, B6002453, B6002454, B6002455, B6002456, B6002523, B6002532, B6002533, B6002552, B6002567, B6002568, B6002569, B6002570, B6002571, B6002572, B6002573, B6002574, B6002575, B6002576, B6002577, B6002578, B6002579, B6002580, B6002581, B6002582, B6002583, B6002584, B6002585, B6002586, B6002587, B6002588, B6002589, B6002590, B6002591, B6002592, B6002593, B6002594, B6002595, B6002596, B6002597, B6002598, B6002599, B6002600, B6002601, B6002603, B6002604, B6002605, B6002606, B6002607, B6002608, B6002609, B6002610, B6002611, B6002612, B6002613, B6002614, B6002615, B6002616, B6002617, B6002618, B6002619, B6002620, B6002621, B6002622, B6002623, B6002624, B6002625, B6002626, B6002627, B6002628, B6002629, B6002630, B6002631, B6002632, B6002633, B6002634, B6002635, B6002636, B6002637, B6002638, B6002639, B6002640, B6002641, B6002642, B6002643, B6002644, B6002645, B6002646, B6002647, B6002648, B6002649, B6002650, B6002651, B6002652, B6002653, B6002654, B6002655, B6002656, B6002657, B6002658, B6002659, B6002660, B6002661, B6002662, B6002663, B6002664, B6002665, B6002666, B6002706, B6002731, B6002738, B6002739, B6002740, B6002741, B6002742, B6002743, B6002744, B6002802, B6002804, and B6002877.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Verathon, Inc.
- Reason for Recall:
- The firm is providing customers with an updated Operations and Maintenance Manual for each of the BladderScan BVI 9600 and AortaScan AMI 9700 devices. The updated Operations and Maintenance Manuals clarify that these devices should not be used for the screening, detection, or diagnosis of abdominal aortic aneurysms (AAAs).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BladderScan BVI 9600, Model Numbers: 0270-0451, 0270-0452, 0270-0754, and U270-0451. AortaScan AMI 9700, Model Numbers: 0270-0636, 0270-0639, and U270-0636.
Product Codes/Lot Numbers:
SERIAL NUMBERS OF UNITS DISTRIBUTED IN THE USA **************** A7001001, A7001002, A7001005, A7001007, A7001011, A7001022, A7001024, A7001031, A7001034, A7001036, A7001040, A7001041, A7001042, A7001043, A7001044, A7001045, A7001046, A7001047, A7001049, A7001050, A7001051, A7001052, A7001053, A7001054, A7001055, A7001056, A7001057, A7001060, A7001062, A7001065, A7001067, A7001068, A7001079, A7001080, A7001086, A7001088, A7001093, A7001094, A7001098, A7001101, A7001116, A7001117, A7001118, A7001119, A7001120, A7001121, A7001122, A7001123, A7001124, A7001125, A7001126, A7001127, A7001128, A7001129, A7001130, A7001131, A7001132, A7001138, A7001139, A7001140, A7001141, A7001143, A7001144, A7001146, A7001147, A7001148, A7001150, A7001151, A7001152, A7001153, A7001155, A7001156, A7001157, A7001158, A7001161, A7001162, A7001163, A7001164, A7001165, A7001168, A7001169, A7001170, A7001171, A7001172, A7001173, A7001174, A7001175, A7001176, A7001177, A7001178, A7001179, A7001180, A7001181, A7001182, A7001183, A7001184, A7001185, A7001186, A7001187, A7001188, A7001189, A7001190, A7001191, A7001192, A7001194, A7001195, A7001196, A7001197, A7001198, A7001199, A7001201, A7001202, A7001204, A7001205, A7001206, A7001207, A7001209, A7001210, A7001211, A7001212, A7001216, A7001217, A7001218, A7001219, A7001220, A7001221, A7001222, A7001223, A7001224, A7001226, A7001227, A7001228, A7001229, A7001231, A7001232, A7001233, A7001234, A7001235, A7001236, A7001237, A7001238, A7001239, A7001240, A7001241, A7001242, A7001243, A7001244, A7001246, A7001247, A7001248, A7001249, A7001250, A7001251, A7001252, A7001253, A7001254, A7001255, A7001256, A7001257, A7001258, A7001260, A7001261, A7001262, A7001264, A7001265, A7001266, A7001268, A7001269, A7001270, A7001271, A7001272, A7001273, A7001275, A7001276, A7001277, A7001278, A7001279, A7001281, A7001283, A7001284, A7001285, A7001286, A7001287, A7001288, A7001289, A7001290, A7001291, A7001292, A7001293, A7001294, A7001295, A7001296, A7001297, A7001298, A7001299, A7001300, A7001301, A7001302, A7001303, A7001305, A7001306, A7001307, A7001308, A7001312, A7001313, A7001314, A7001315, A7001316, A7001317, A7001318, A7001319, A7001320, A7001321, A7001322, A7001323, A7001324, A7001325, A7001326, A7001327, A7001328, A7001329, A7001330, A7001331, A7001332, A7001333, A7001334, A7001335, A7001336, A7001337, A7001338, A7001339, A7001340, A7001341, A7001342, A7001343, A7001344, A7001345, A7001347, A7001348, A7001349, A7001350, A7001351, A7001352, A7001353, A7001354, A7001355, A7001356, A7001357, A7001358, A7001359, A7001360, A7001361, A7001362, A7001363, A7001364, A7001365, A7001366, A7001367, A7001369, A7001370, A7001371, A7001372, A7001373, A7001375, A7001376, A7001377, A7001378, A7001380, A7001381, A7001382, A7001383, A7001384, A7001385, A7001386, A7001387, A7001388, A7001389, A7001390, A7001391, A7001392, A7001393, A7001394, A7001395, A7001396, A7001397, A7001398, A7001399, A7001400, A7001401, A7001402, A7001403, A7001404, A7001405, A7001406, A7001408, A7001409, A7001410, A7001411, A7001412, A7001413, A7001414, A7001415, A7001416, A7001418, A7001419, A7001420, A7001421, A7001422, A7001423, A7001424, A7001425, A7001426, A7001427, A7001428, A7001429, A7001430, A7001431, A7001432, A7001433, A7001434, A7001435, A7001436, A7001437, A7001438, A7001439, A7001440, A7001441, A7001445, A7001446, A7001447, A7001448, A7001449, A7001450, A7001451, A7001452, A7001453, A7001454, A7001455, A7001456, A7001457, A7001458, A7001459, A7001460, A7001461, A7001462, A7001463, A7001464, A7001465, A7001466, A7001467, A7001468, A7001469, A7001470, A7001471, A7001472, A7001473, A7001474, A7001475, A7001476, A7001477, A7001478, A7001479, A7001480, A7001482, A7001483, A7001484, A7001485, A7001486, A7001487, A7001488, A7001489, A7001490, A7001491, A7001492, A7001493, A7001494, A7001495, A7001496, A7001498, A7001499, A7001500, A7001502, A7001503, A7001504, A7001505, A7001506, A7001507, A7001508, A7001509, A7001510, A7001511, A7001512, A7001513, A7001514, A7001515, A7001516, A7001517, A7001518, A7001519, A7001520, A7001521, A7001522, A7001523, A7001524, A7001525, A7001526, A7001527, A7001528, A7001529, A7001530, A7001531, A7001532, A7001533, A7001534, A7001535, A7001536, A7001537, A7001538, A7001539, A7001540, A7001541, A7001542, A7001543, A7001544, A7001545, A7001546, A7001547, A7001548, A7001549, A7001550, A7001551, A7001552, A7001553, A7001554, A7001555, A7001556, A7001557, A7001558, A7001559, A7001560, A7001561, A7001562, A7001563, A7001564, A7001565, A7001566, A7001567, A7001568, A7001569, A7001570, A7001571, A7001572, A7001573, A7001574, A7001575, A7001576, A7001577, A7001578, A7001579, A7001580, A7001581, A7001582, A7001583, A7001584, A7001585, A7001586, A7001587, A7001588, A7001589, A7001590, A7001591, A7001592, A7001593, A7001594, A7001595, A7001596, A7001597, A7001598, A7001599, A7001600, A7001601, A7001602, A7001603, A7001604, A7001605, A7001606, A7001607, A7001608, A7001609, A7001610, A7001611, A7001612, A7001613, A7001614, A7001615, A7001616, A7001617, A7001618, A7001619, A7001620, A7001621, A7001622, A7001623, A7001624, A7001625, A7001626, A7001628, A7001629, A7001630, A7001631, A7001632, A7001633, A7001634, A7001635, A7001636, A7001637, A7001638, A7001639, A7001640, A7001641, A7001642, A7001643, A7001644, A7001645, A7001646, A7001647, A7001648, A7001649, A7001650, A7001651, A7001652, A7001653, A7001654, A7001655, A7001656, A7001657, A7001658, A7001659, A7001660, A7001661, A7001662, A7001663, A7001664, A7001665, A7001666, A7001667, A7001668, A7001669, A7001670, A7001671, A7001672, A7001673, A7001674, A7001675, A7001676, A7001677, A7001678, A7001679, A7001681, A7001683, A7001684, A7001685, A7001686, A7001687, A7001688, A7001689, A7001690, A7001691, A7001692, A7001693, A7001694, A7001695, A7001697, A7001698, A7001699, A7001700, A7001701, A7001702, A7001703, A7001704, A7001705, A7001706, A7001707, A7001708, A7001709, A7001710, A7001711, A7001712, A7001713, A7001714, A7001715, A7001716, A7001717, A7001718, A7001719, A7001721, A7001722, A7001723, A7001724, A7001725, A7001726, A7001727, A7001729, A7001730, A7001732, A7001733, A7001734, A7001735, A7001736, A7001737, A7001738, A7001739, A7001740, A7001741, A7001742, A7001743, A7001744, A7001745, A7001746, A7001747, A7001748, A7001749, A7001752, A7001753, A7001754, A7001755, A7001756, A7001757, A7001758, A7001759, A7001760, A7001761, A7001762, A7001763, A7001765, A7001766, A7001767, A7001768, A7001769, A7001770, A7001771, A7001780, A7001781, A7001782, A7001783, A7001784, A7001785, A7001786, A7001787, A7001788, A7001789, A7001790, A7001791, A7001792, A7001793, A7001794, A7001795, A7001800, A7001801, A7001802, A7001803, A7001804, A7001805, A7001806, A7001807, A7001808, A7001809, A7001810, A7001811, A7001812, A7001813, A7001814, A7001815, A7001816, A7001817, A7001819, A7001820, A7001821, A7001822, A7001823, A7001824, A7001825, A7001826, A7001827, A7001828, A7001836, A7001837, A7001841, A7001842, A7001843, A7001844, A7001845, A7001846, A7001847, A7001848, A7001857, A7001858, A7001859, A7001860, A7001861, A7001862, A7001863, A7001864, A7001865, A7001866, A7001867, A7001868, A7001869, A7001870, A7001871, A7001872, A7001873, A7001874, A7001876, A7001877, A7001878, A7001879, A7001880, A7001881, A7001882, A7001883, A7001884, A7001885, A7001886, A7001887, A7001888, A7001889, A7001890, A7001891, A7001892, A7001893, A7001894, A7001895, A7001897, A7001898, A7001899, A7001900, A7001901, A7001902, A7001903, A7001904, A7001905, A7001906, A7001907, A7001908, A7001909, A7001910, A7001911, A7001912, A7001913, A7001914, A7001915, A7001916, A7001917, A7001918, A7001919, A7001920, A7001921, A7001922, A7001923, A7001924, A7001925, A7001927, A7001928, A7001929, A7001930, A7001931, A7001932, A7001933, A7001934, A7001935, A7001936, A7001937, A7001938, A7001939, A7001940, A7001941, A7001942, A7001943, A7001944, A7001945, A7001946, A7001947, A7001948, A7001949, A7001950, A7001951, A7001952, A7001953, A7001954, A7001955, A7001956, A7001957, A7001958, A7001959, A7001960, A7001961, A7001962, A7001963, A7001964, A7001965, A7001966, A7001967, A7001968, A7001969, A7001970, A7001971, A7001972, A7001973, A7001981, A7300000, A7300001, A7300002, A7300003, A7300004, A7300005, A7300006, A7300007, A7300008, A7300012, B6001016, B6001020, B6001023, B6001024, B6001025, B6001026, B6001030, B6001040, B6001041, B6001043, B6001047, B6001049, B6001060, B6001065, B6001066, B6001068, B6001069, B6001070, B6001071, B6001073, B6001080, B6001082, B6001084, B6001085, B6001086, B6001088, B6001090, B6001092, B6001093, B6001094, B6001095, B6001096, B6001097, B6001098, B6001100, B6001102, B6001103, B6001104, B6001107, B6001108, B6001109, B6001110, B6001111, B6001112, B6001113, B6001123, B6001124, B6001130, B6001131, B6001133, B6001138, B6001140, B6001141, B6001143, B6001148, B6001157, B6001160, B6001164, B6001165, B6001166, B6001167, B6001168, B6001169, B6001170, B6001171, B6001172, B6001173, B6001174, B6001175, B6001176, B6001177, B6001178, B6001179, B6001180, B6001181, B6001183, B6001184, B6001185, B6001186, B6001187, B6001198, B6001199, B6001202, B6001203, B6001204, B6001205, B6001206, B6001207, B6001208, B6001209, B6001212, B6001213, B6001216, B6001217, B6001218, B6001219, B6001220, B6001221, B6001223, B6001225, B6001227, B6001231, B6001232, B6001234, B6001237, B6001238, B6001239, B6001241, B6001242, B6001243, B6001244, B6001245, B6001246, B6001254, B6001256, B6001257, B6001258, B6001259, B6001260, B6001261, B6001263, B6001265, B6001267, B6001269, B6001270, B6001271, B6001272, B6001273, B6001276, B6001277, B6001278, B6001279, B6001280, B6001281, B6001282, B6001283, B6001284, B6001285, B6001286, B6001287, B6001288, B6001298, B6001299, B6001305, B6001306, B6001307, B6001308, B6001309, B6001310, B6001311, B6001312, B6001313, B6001314, B6001315, B6001316, B6001317, B6001318, B6001319, B6001320, B6001321, B6001322, B6001323, B6001324, B6001325, B6001326, B6001327, B6001328, B6001329, B6001330, B6001331, B6001332, B6001339, B6001341, B6001342, B6001343, B6001345, B6001346, B6001348, B6001349, B6001350, B6001351, B6001354, B6001355, B6001356, B6001357, B6001358, B6001359, B6001360, B6001361, B6001362, B6001363, B6001364, B6001365, B6001366, B6001367, B6001369, B6001370, B6001372, B6001373, B6001374, B6001375, B6001377, B6001378, B6001379, B6001381, B6001382, B6001384, B6001386, B6001387, B6001388, B6001389, B6001390, B6001396, B6001397, B6001398, B6001399, B6001401, B6001402, B6001404, B6001406, B6001407, B6001408, B6001409, B6001410, B6001411, B6001419, B6001420, B6001421, B6001422, B6001423, B6001424, B6001425, B6001426, B6001427, B6001428, B6001429, B6001431, B6001435, B6001436, B6001437, B6001438, B6001439, B6001440, B6001441, B6001442, B6001443, B6001444, B6001445, B6001448, B6001449, B6001450, B6001451, B6001452, B6001453, B6001455, B6001456, B6001457, B6001459, B6001461, B6001462, B6001467, B6001468, B6001469, B6001470, B6001471, B6001472, B6001473, B6001474, B6001477, B6001478, B6001479, B6001480, B6001481, B6001482, B6001483, B6001484, B6001485, B6001486, B6001487, B6001488, B6001489, B6001490, B6001491, B6001492, B6001493, B6001494, B6001495, B6001496, B6001497, B6001498, B6001499, B6001500, B6001501, B6001502, B6001503, B6001504, B6001505, B6001506, B6001507, B6001508, B6001509, B6001510, B6001511, B6001512, B6001514, B6001515, B6001516, B6001518, B6001519, B6001520, B6001521, B6001522, B6001523, B6001524, B6001525, B6001526, B6001529, B6001530, B6001531, B6001532, B6001533, B6001535, B6001537, B6001541, B6001542, B6001544, B6001545, B6001546, B6001547, B6001549, B6001550, B6001553, B6001554, B6001557, B6001558, B6001559, B6001560, B6001561, B6001562, B6001563, B6001564, B6001565, B6001567, B6001568, B6001569, B6001570, B6001571, B6001572, B6001573, B6001574, B6001575, B6001576, B6001577, B6001578, B6001579, B6001580, B6001581, B6001582, B6001583, B6001584, B6001585, B6001586, B6001587, B6001588, B6001589, B6001590, B6001591, B6001592, B6001593, B6001594, B6001595, B6001596, B6001597, B6001598, B6001599, B6001600, B6001601, B6001602, B6001603, B6001604, B6001605, B6001606, B6001608, B6001609, B6001610, B6001611, B6001612, B6001613, B6001614, B6001615, B6001616, B6001617, B6001618, B6001619, B6001620, B6001621, B6001622, B6001623, B6001624, B6001627, B6001628, B6001629, B6001630, B6001631, B6001632, B6001633, B6001634, B6001635, B6001636, B6001637, B6001638, B6001639, B6001640, B6001641, B6001642, B6001643, B6001644, B6001645, B6001646, B6001647, B6001648, B6001660, B6001661, B6001662, B6001663, B6001664, B6001665, B6001666, B6001667, B6001668, B6001669, B6001670, B6001672, B6001673, B6001674, B6001675, B6001676, B6001677, B6001678, B6001679, B6001680, B6001681, B6001682, B6001683, B6001685, B6001686, B6001687, B6001688, B6001689, B6001691, B6001692, B6001693, B6001694, B6001695, B6001696, B6001697, B6001698, B6001699, B6001700, B6001701, B6001702, B6001703, B6001704, B6001705, B6001707, B6001708, B6001709, B6001710, B6001711, B6001712, B6001713, B6001715, B6001716, B6001717, B6001718, B6001719, B6001720, B6001721, B6001722, B6001723, B6001724, B6001725, B6001726, B6001727, B6001728, B6001729, B6001730, B6001731, B6001733, B6001734, B6001735, B6001736, B6001738, B6001739, B6001740, B6001741, B6001742, B6001743, B6001744, B6001746, B6001747, B6001748, B6001750, B6001751, B6001752, B6001753, B6001754, B6001755, B6001756, B6001757, B6001758, B6001759, B6001761, B6001762, B6001764, B6001765, B6001766, B6001767, B6001768, B6001769, B6001770, B6001771, B6001772, B6001773, B6001774, B6001776, B6001777, B6001778, B6001779, B6001780, B6001781, B6001782, B6001783, B6001787, B6001788, B6001790, B6001791, B6001793, B6001795, B6001796, B6001797, B6001798, B6001799, B6001800, B6001801, B6001803, B6001805, B6001806, B6001807, B6001808, B6001809, B6001810, B6001811, B6001812, B6001813, B6001814, B6001815, B6001816, B6001817, B6001818, B6001819, B6001820, B6001821, B6001822, B6001824, B6001825, B6001826, B6001827, B6001828, B6001829, B6001830, B6001831, B6001832, B6001833, B6001834, B6001835, B6001836, B6001838, B6001839, B6001840, B6001841, B6001842, B6001843, B6001844, B6001845, B6001846, B6001848, B6001849, B6001850, B6001851, B6001852, B6001853, B6001854, B6001855, B6001856, B6001857, B6001858, B6001859, B6001860, B6001861, B6001862, B6001863, B6001864, B6001865, B6001866, B6001867, B6001868, B6001869, B6001870, B6001871, B6001872, B6001873, B6001874, B6001875, B6001876, B6001877, B6001878, B6001879, B6001880, B6001881, B6001882, B6001883, B6001884, B6001885, B6001887, B6001888, B6001889, B6001890, B6001891, B6001892, B6001893, B6001894, B6001895, B6001896, B6001897, B6001898, B6001899, B6001900, B6001901, B6001902, B6001903, B6001904, B6001905, B6001906, B6001907, B6001908, B6001909, B6001910, B6001911, B6001912, B6001913, B6001914, B6001915, B6001916, B6001917, B6001918, B6001919, B6001920, B6001921, B6001922, B6001923, B6001924, B6001925, B6001926, B6001927, B6001928, B6001930, B6001931, B6001932, B6001934, B6001935, B6001936, B6001937, B6001938, B6001939, B6001940, B6001941, B6001942, B6001943, B6001944, B6001945, B6001946, B6001947, B6001948, B6001949, B6001950, B6001951, B6001952, B6001953, B6001954, B6001955, B6001956, B6001957, B6001958, B6001959, B6001960, B6001961, B6001962, B6001963, B6001964, B6001965, B6001966, B6001967, B6001968, B6001969, B6001972, B6001973, B6001974, B6001975, B6001976, B6001977, B6001978, B6001979, B6001980, B6001981, B6001982, B6001983, B6001984, B6001986, B6001987, B6001988, B6001989, B6001990, B6001991, B6001992, B6001993, B6001994, B6001995, B6001996, B6001997, B6001998, B6001999, B6002000, B6002001, B6002002, B6002003, B6002004, B6002005, B6002007, B6002008, B6002009, B6002010, B6002011, B6002012, B6002013, B6002014, B6002015, B6002016, B6002017, B6002018, B6002019, B6002020, B6002021, B6002022, B6002023, B6002024, B6002025, B6002026, B6002027, B6002028, B6002029, B6002030, B6002031, B6002032, B6002033, B6002034, B6002035, B6002036, B6002037, B6002038, B6002039, B6002040, B6002041, B6002042, B6002043, B6002044, B6002045, B6002047, B6002048, B6002049, B6002050, B6002051, B6002052, B6002054, B6002055, B6002057, B6002058, B6002059, B6002060, B6002061, B6002062, B6002063, B6002064, B6002065, B6002066, B6002067, B6002068, B6002069, B6002070, B6002071, B6002072, B6002073, B6002074, B6002075, B6002076, B6002077, B6002078, B6002079, B6002080, B6002081, B6002082, B6002083, B6002084, B6002085, B6002086, B6002087, B6002088, B6002089, B6002090, B6002091, B6002092, B6002093, B6002094, B6002096, B6002097, B6002098, B6002099, B6002100, B6002101, B6002102, B6002103, B6002104, B6002105, B6002106, B6002107, B6002108, B6002109, B6002110, B6002111, B6002112, B6002113, B6002114, B6002115, B6002116, B6002117, B6002118, B6002119, B6002120, B6002121, B6002122, B6002123, B6002124, B6002125, B6002126, B6002127, B6002128, B6002129, B6002130, B6002131, B6002132, B6002133, B6002134, B6002135, B6002136, B6002137, B6002138, B6002139, B6002140, B6002141, B6002142, B6002143, B6002145, B6002147, B6002148, B6002149, B6002150, B6002151, B6002152, B6002153, B6002154, B6002155, B6002156, B6002157, B6002158, B6002159, B6002160, B6002161, B6002162, B6002163, B6002164, B6002168, B6002169, B6002170, B6002171, B6002172, B6002173, B6002174, B6002175, B6002176, B6002177, B6002178, B6002179, B6002180, B6002181, B6002182, B6002183, B6002184, B6002185, B6002186, B6002187, B6002188, B6002189, B6002190, B6002191, B6002192, B6002193, B6002195, B6002196, B6002197, B6002198, B6002199, B6002200, B6002201, B6002202, B6002203, B6002204, B6002205, B6002206, B6002207, B6002208, B6002209, B6002210, B6002211, B6002212, B6002213, B6002214, B6002215, B6002216, B6002217, B6002218, B6002219, B6002220, B6002221, B6002222, B6002223, B6002224, B6002225, B6002226, B6002227, B6002228, B6002229, B6002230, B6002231, B6002232, B6002233, B6002234, B6002235, B6002236, B6002237, B6002238, B6002239, B6002240, B6002241, B6002242, B6002243, B6002244, B6002245, B6002246, B6002247, B6002248, B6002251, B6002252, B6002253, B6002254, B6002255, B6002256, B6002257, B6002258, B6002259, B6002260, B6002261, B6002262, B6002263, B6002264, B6002265, B6002266, B6002267, B6002268, B6002269, B6002270, B6002271, B6002272, B6002273, B6002274, B6002275, B6002276, B6002277, B6002278, B6002279, B6002280, B6002281, B6002282, B6002283, B6002284, B6002285, B6002286, B6002287, B6002288, B6002289, B6002292, B6002293, B6002294, B6002295, B6002296, B6002297, B6002298, B6002299, B6002300, B6002301, B6002302, B6002303, B6002304, B6002305, B6002306, B6002307, B6002308, B6002309, B6002310, B6002311, B6002313, B6002314, B6002315, B6002316, B6002317, B6002318, B6002319, B6002320, B6002321, B6002322, B6002323, B6002324, B6002325, B6002326, B6002327, B6002328, B6002329, B6002331, B6002332, B6002334, B6002335, B6002336, B6002337, B6002338, B6002339, B6002340, B6002341, B6002342, B6002343, B6002345, B6002346, B6002347, B6002348, B6002349, B6002350, B6002351, B6002352, B6002353, B6002354, B6002356, B6002357, B6002358, B6002359, B6002360, B6002361, B6002362, B6002363, B6002364, B6002365, B6002366, B6002367, B6002368, B6002369, B6002370, B6002371, B6002373, B6002374, B6002375, B6002376, B6002377, B6002378, B6002379, B6002381, B6002382, B6002383, B6002384, B6002385, B6002386, B6002387, B6002388, B6002389, B6002390, B6002391, B6002392, B6002393, B6002394, B6002395, B6002396, B6002398, B6002399, B6002401, B6002402, B6002403, B6002404, B6002405, B6002406, B6002407, B6002408, B6002411, B6002412, B6002413, B6002414, B6002415, B6002416, B6002417, B6002418, B6002419, B6002420, B6002421, B6002422, B6002423, B6002424, B6002425, B6002426, B6002427, B6002428, B6002429, B6002430, B6002431, B6002432, B6002433, B6002434, B6002435, B6002436, B6002437, B6002438, B6002439, B6002440, B6002441, B6002442, B6002443, B6002444, B6002445, B6002446, B6002447, B6002448, B6002449, B6002451, B6002452, B6002457, B6002458, B6002459, B6002460, B6002461, B6002462, B6002463, B6002469, B6002470, B6002471, B6002472, B6002473, B6002474, B6002475, B6002476, B6002477, B6002478, B6002479, B6002480, B6002481, B6002482, B6002483, B6002484, B6002485, B6002486, B6002487, B6002488, B6002489, B6002490, B6002491, B6002492, B6002493, B6002494, B6002495, B6002496, B6002504, B6002505, B6002506, B6002507, B6002508, B6002509, B6002510, B6002511, B6002512, B6002513, B6002514, B6002515, B6002519, B6002520, B6002521, B6002522, B6002524, B6002525, B6002526, B6002528, B6002529, B6002530, B6002531, B6002534, B6002535, B6002536, B6002542, B6002543, B6002544, B6002545, B6002546, B6002547, B6002548, B6002549, B6002550, B6002551, B6002553, B6002554, B6002555, B6002556, B6002557, B6002558, B6002559, B6002560, B6002561, B6002562, B6002563, B6002564, B6002565, B6002566, B6002602, B6002667, B6002668, B6002669, B6002670, B6002671, B6002672, B6002673, B6002674, B6002675, B6002676, B6002677, B6002678, B6002679, B6002680, B6002681, B6002682, B6002683, B6002684, B6002685, B6002686, B6002687, B6002688, B6002689, B6002690, B6002691, B6002692, B6002693, B6002694, B6002695, B6002696, B6002697, B6002698, B6002700, B6002701, B6002702, B6002703, B6002704, B6002705, B6002707, B6002708, B6002709, B6002710, B6002711, B6002712, B6002713, B6002714, B6002715, B6002716, B6002717, B6002718, B6002719, B6002720, B6002722, B6002723, B6002724, B6002725, B6002726, B6002727, B6002728, B6002729, B6002730, B6002732, B6002733, B6002734, B6002735, B6002736, B6002737, B6002746, B6002747, B6002748, B6002749, B6002750, B6002751, B6002752, B6002753, B6002754, B6002755, B6002756, B6002757, B6002758, B6002759, B6002760, B6002761, B6002762, B6002763, B6002764, B6002765, B6002766, B6002767, B6002768, B6002769, B6002770, B6002771, B6002772, B6002773, B6002774, B6002775, B6002777, B6002778, B6002779, B6002781, B6002782, B6002783, B6002784, B6002785, B6002786, B6002787, B6002788, B6002789, B6002790, B6002791, B6002792, B6002793, B6002794, B6002795, B6002796, B6002797, B6002798, B6002799, B6002800, B6002801, B6002803, B6002805, B6002806, B6002807, B6002808, B6002809, B6002810, B6002811, B6002812, B6002813, B6002815, B6002816, B6002817, B6002818, B6002819, B6002820, B6002821, B6002822, B6002823, B6002824, B6002825, B6002826, B6002827, B6002828, B6002829, B6002830, B6002831, B6002832, B6002833, B6002834, B6002835, B6002836, B6002837, B6002838, B6002839, B6002840, B6002841, B6002842, B6002843, B6002844, B6002845, B6002849, B6002857, B6002858, B6002859, B6002860, B6002861, B6002862, B6002863, B6002864, B6002865, B6002866, B6002867, B6002868, B6002869, B6002870, B6002871, B6002872, B6002873, B6002874, B6002875, B6002876, B6002878, B6002879, B6002880, B6002881, B6002882, B6002883, B6002884, B6002885, B6002886, B6002888, B6002889, B6002890, B6002891, and B6002892. ****SERIAL NUMBERS OF UNITS DISTRIBUTED OUTSIDE THE USA***** 00001001, 00001007, 00001008, A7001048, A7001069, A7001071, A7001089, A7001090, A7001091, A7001092, A7001095, A7001096, A7001097, A7001100, A7001102, A7001103, A7001106, A7001107, A7001108, A7001109, A7001110, A7001111, A7001112, A7001113, A7001114, A7001115, A7001133, A7001136, A7001263, A7001280, A7001282, A7001442, A7001443, A7001444, A7001720, A7001796, A7001797, A7001798, A7001799, A7001829, A7001830, A7001831, A7001832, A7001833, A7001834, A7001835, A7001838, A7001839, A7001840, A7001849, A7001850, A7001851, A7001852, A7001853, A7001854, A7001855, A7001856, B6001063, B6001064, B6001075, B6001079, B6001081, B6001188, B6001189, B6001211, B6001224, B6001233, B6001251, B6001289, B6001294, B6001296, B6001300, B6001337, B6001347, B6001353, B6001368, B6001371, B6001380, B6001383, B6001391, B6001393, B6001394, B6001395, B6001405, B6001412, B6001413, B6001416, B6001418, B6001432, B6001446, B6001447, B6001454, B6001465, B6001466, B6001475, B6001476, B6001517, B6001534, B6001539, B6001540, B6001548, B6001551, B6001552, B6001556, B6001607, B6001625, B6001671, B6001684, B6001690, B6001714, B6001732, B6001745, B6001749, B6001760, B6001763, B6001775, B6001784, B6001785, B6001786, B6001789, B6001792, B6001802, B6001847, B6001886, B6001929, B6001933, B6001970, B6001971, B6001985, B6002053, B6002056, B6002166, B6002167, B6002194, B6002250, B6002290, B6002291, B6002330, B6002333, B6002344, B6002372, B6002409, B6002410, B6002450, B6002453, B6002454, B6002455, B6002456, B6002523, B6002532, B6002533, B6002552, B6002567, B6002568, B6002569, B6002570, B6002571, B6002572, B6002573, B6002574, B6002575, B6002576, B6002577, B6002578, B6002579, B6002580, B6002581, B6002582, B6002583, B6002584, B6002585, B6002586, B6002587, B6002588, B6002589, B6002590, B6002591, B6002592, B6002593, B6002594, B6002595, B6002596, B6002597, B6002598, B6002599, B6002600, B6002601, B6002603, B6002604, B6002605, B6002606, B6002607, B6002608, B6002609, B6002610, B6002611, B6002612, B6002613, B6002614, B6002615, B6002616, B6002617, B6002618, B6002619, B6002620, B6002621, B6002622, B6002623, B6002624, B6002625, B6002626, B6002627, B6002628, B6002629, B6002630, B6002631, B6002632, B6002633, B6002634, B6002635, B6002636, B6002637, B6002638, B6002639, B6002640, B6002641, B6002642, B6002643, B6002644, B6002645, B6002646, B6002647, B6002648, B6002649, B6002650, B6002651, B6002652, B6002653, B6002654, B6002655, B6002656, B6002657, B6002658, B6002659, B6002660, B6002661, B6002662, B6002663, B6002664, B6002665, B6002666, B6002706, B6002731, B6002738, B6002739, B6002740, B6002741, B6002742, B6002743, B6002744, B6002802, B6002804, and B6002877.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1091-2016
Related Recalls
Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.
Due to software issues, their is the potential for loss of image or a degradation of the image when using the monitor with other connectable devices.