Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    REF: 10401008. Serial Numbers: 1324, 1325, 1431, 1432, 1084, 1087, 1367, 1362,1363,1615, 1616, 1358, 1359, 1453, 1354, 1357, 1599, 1176, 1178, 1353, 1356, 1576, 1577, 1578, 1322, 1323, 1454, 1594, 1598, 1288, 1290, 1688, 1690, 1691, 1694, 1798, 1799, 1800, 1801, 1135, 1151, 1326, 1327, 1477, 1478, 1479, 1480, 1619, 1620, 1617, 1618, 1489, 1501, 1687, 1573, 1574, 1144, 1145, 1247, 1249 ,1264, 1277, 1265, 1277, 1278, 1381, 1150, 1286, 1148, 1149, 1254, 1260, 1397, 1542, 1458, 1517, 1231, 1238, 1282, 1284, 1389
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Accelerate Diagnostics Inc
Reason for Recall:
Rare isolates of Enterobacteriaceae may generate a susceptible meropenem Minimum Inhibitory Concentration (MIC) using the affected device and a resistant result by broth microdilution. In order to mitigate any risk associated with these rare isolates, the firm has implemented an Expert Rule that suppresses the meropenem result for isolates that display a profile consistent with those that generated a false-susceptible result. The firm has responded to this risk by implementing a set of suppression rules, in a software update, to prevent incorrect results.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Accelerate Pheno system, with Accelerate Pheno System Software 1.2.1 - 1.3.1, PSW000012B Product Usage: The Accelerate Pheno system is intended to measure signal intensity of fluorescent probes bound to nucleic acid in target and non-target organisms and to take time-lapse dark-field images of immobilized growing bacterial cells when used with Accelerate PhenoTest kits.

Product Codes/Lot Numbers:

REF: 10401008. Serial Numbers: 1324, 1325, 1431, 1432, 1084, 1087, 1367, 1362,1363,1615, 1616, 1358, 1359, 1453, 1354, 1357, 1599, 1176, 1178, 1353, 1356, 1576, 1577, 1578, 1322, 1323, 1454, 1594, 1598, 1288, 1290, 1688, 1690, 1691, 1694, 1798, 1799, 1800, 1801, 1135, 1151, 1326, 1327, 1477, 1478, 1479, 1480, 1619, 1620, 1617, 1618, 1489, 1501, 1687, 1573, 1574, 1144, 1145, 1247, 1249 ,1264, 1277, 1265, 1277, 1278, 1381, 1150, 1286, 1148, 1149, 1254, 1260, 1397, 1542, 1458, 1517, 1231, 1238, 1282, 1284, 1389

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1065-2019

Related Recalls

Class I - Dangerous

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

Oct 17, 2022 Diagnostic Equipment Nationwide View Details →
Class I - Dangerous

Due improperly formulated and released bulk lot of fluorescence in situ hybridization (FISH) probes resulting in false positive identification results for Staphylococcus aureus and Streptococcus spp.

Oct 17, 2022 Diagnostic Equipment Nationwide View Details →