Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    10110770 11030805 11030806 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Villa Radiology Systems LLC
Reason for Recall:
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Juno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.

Product Codes/Lot Numbers:

10110770 11030805 11030806 11070845 11070848 11070849 11090864 11110900 12020978 12020982 12030999 12041003 12061014 12061018 12061020 12081040 12091074 12101104 12121186 12121187 13031215 13101263 14021301 14031313 14041321 14051332 14071353 14081359 14091368 14091371 14121400 14121403 15021426 15031440 15041444 15091473 15091478 15101484 15101487

Distribution:

Nationwide distribution

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1063-2017

Related Recalls

The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.

Feb 19, 2016 Other Medical Devices Nationwide View Details →

The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.

Feb 19, 2016 Other Medical Devices Nationwide View Details →

The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.

Feb 19, 2016 Other Medical Devices Nationwide View Details →