CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Sensation 7 fr.34 cc IAB Demonstration Unit 0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469 Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470 Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units 0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568 Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576 Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Maquet Datascope Corp - Cardiac Assist Division
- Reason for Recall:
- Ferrule Cap was not removed on some fiber-optic Sensation and Sensation Plus IABC demonstration units. If fiber-optic connector is inserted into the Maquet CS300 or CARDIOSAVE IABP, cap will prevent connection of unit and may become dislodged and remain in IABPs fiber-optic receptacle.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
CARDIOSAVE Intra-Aortic Balloon Pump and CS300 Intra-Aortic Balloon Pump; and, non-sterile, not for human use, SENSATION and SENSATION PLUS IABC demonstration units. Part Number 0998-00-0800-xx; 0998-00-3023-xx; 0998-UC-0800-xx; 0998-UC-3023-xx An electromechanical system used to inflate and deflate intra-aortic balloons.
Product Codes/Lot Numbers:
Sensation 7 fr.34 cc IAB Demonstration Unit 0684-DC-0433 0684-DO-0433 0684-DC-0469 0684-DO-0469 Sensation 7 Fr. 40 cc IAB Demonstration Unit 0684-DC-0434 0684-DO-0434 0684-DC-0470 0684-DO-0470 Sesnsation Plus 7.5 Fr. 40cc IAB Demonstration Units 0684-DC-0567 0684-DO-0567 0684-DC-0568 0684-DO-0568 Sensation Plus 7.5 Fr. 50cc IAB Demonstration Units 0684-DC-0575 0684-DO-0575 0684-DC-0576 0684-DO-0576 Sensation Plus 8 Fr. 50 cc IAB Demonstration Unit 0684-DC-0272 0684-DO-0272
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1062-2015
Related Recalls
Cardiosave Hybrid IABP, Part Number 0998-00-0800-53
Maquet Datascope Corp - Cardiac Assist Division
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
Cardiosave Rescue IABP, Part Number 0998-UC-0800-83
Maquet Datascope Corp - Cardiac Assist Division
There is a potential for interruption and/or inability to start therapy to the patient prior to or during use of Cardiosave IABP for users who are at altitudes above 3200 Ft.
CARDIOSAVE Hybrid Intra-Aortic Balloon Pump (IABP), Part Number 0998-00-0800-XX & 0998-UC-0800-XX (excluding 0998-00-0800-83, 0998-UC-0800-83 & 0998-00-0800-75) and cart 0997-00-1179
Maquet Datascope Corp - Cardiac Assist Division
Ingress of fluids into the device can affect various electronic circuit boards which would prevent initiation or continuation of therapy.