VITEK 2 Systems and VITEK 2 with MYLA.
Class I - DangerousWhat Should You Do?
- Check if you have this product: VITEK 2 - Versions 8.01, 9.01, 9.02 (with or without 9MR2), 9.03 (with or without #9MR2). VITEK 2 with MYLA - Versions 4.8 and 4.9.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- bioMerieux, Inc.
- Reason for Recall:
- Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
VITEK 2 Systems and VITEK 2 with MYLA.
Product Codes/Lot Numbers:
VITEK 2 - Versions 8.01, 9.01, 9.02 (with or without 9MR2), 9.03 (with or without #9MR2). VITEK 2 with MYLA - Versions 4.8 and 4.9.
Distribution:
Distributed in: NC
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1059-2022
Related Recalls
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.