CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Serial No. 13201, 11125, 11519, 11320, 11115.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biosense Webster, Inc.
Reason for Recall:
Biosense Webster has initiated a recall of the CartoUnivu Module within the Carto 3 System (V3.2.2 and V3.2.3) when used with Siemens Axiom Artis VB Fluoroscopy systems in particular procedural configurations only. If the table rotation is not returned to zero point, there is a misalignment between the Carto 3 System map display and the fluoroscopic capture.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

CartoUnivu Module, Model: Carto 3 Software versions 3.2.2 and 3.2.3; Catalog No. KT-5400-124 WITH THE CARTOUNIVU MODULE, FLUOROSCOPIC CAPTURES (IMAGES AND CINE CLIPS) CAN BE IMPORTED TO AND DISPLAYED IN THE CART0 3 SYSTEM. THESE IMAGES ARE NOT FOR DIAGNOSTIC PURPOSES. WHEN FLUOROSCOPIC CAPTURES ARE INTEGRATED INTO THE CART0 3 SYSTEM MAP VIEWERS (MAP VIEWER AND ADDITIONAL VIEW), CART0 3 MAPS, MRl OR CT IMAGES, AND CATHETER VISUALIZATION ARE SUPERIMPOSED ON THE FLUOROSCOPIC CAPTURE.

Product Codes/Lot Numbers:

Serial No. 13201, 11125, 11519, 11320, 11115.

Distribution:

Distributed in: CA, KY, NJ, MA, UT

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1045-2014

Related Recalls

Due to an observed trend of neurovascular events

Jan 5, 2025 Surgical Instruments View Details →

Due to a manufacturing issue, device under process validation phase were inadvertently mixed into the main manufacturing process.