AFX Endovascular AAA System, Endoleak Type: IIIB Common Name: AFX Bifurcated and Accessory Stent Grafts Classification Name: System, Endovascular Graft, Aortic Aneurysm Treatment The AFX Endovascular AAA System is intended for endovascular treatment of patients with abdominal aortic or aortoiliac aneurysms (AAA).
Class I - DangerousWhat Should You Do?
- Check if you have this product: "AFX2 Duraply Lot Codes (July 2015 - Nov. 2016)" Lot Codes 1298056012 1354736007 1354737005 1354738002 1354739002 1354739008 1354739020 1354739027 1354740011 1354745005 1354798002 1354798008 1354798014 1354799004 1354800001 1354800008 1354800014 1354801006 1354801013 1354802004 1354802010 1354804006 1354806001 1354806007 1354806019 1354806026 1354807003 1354807009 1354808003 1354808009 1354809009 1354809015 1354810008 1354810015 1354810022 1354859002 1354859009 1354859016 1354859025 1354860005 1354861001 1354861008 1354861014 1354861021 1354861027 1354862004 1354862011 1354862017 1354862027 1354863007 1354865009 1354865017 1354865025 1354867004 1354867012 1354867020 1354867028 1354921006 1354922002 1354922010 1354922018 1354922024 1354923002 1354923009 1354923015 1354923021 1354924003 1354924012 1354924018 1354924026 1354925005 1354925012 1354925018 1354925024 1354926005 1354926013 1354926020 1354926026 1354927006 1354927019 1354927029 1354928007 1354929005 1354929011 1354980004 1354981001 1354981008 1354982003 1354982010 1354983015 1355041003 1355041011 1355041020 1355042001 1355042008 1355042015 1355042021 1355042027 1355043010 1355044004 1355044013 1355044023 1355047007 1355102002 1355102008 1355102014 1355103009 1355103015 1355103021 1355103029 1355104006 1355104016 1355104024 1355105007 1355106001 1355106007 1355106013 1355162005 1355163005 1355163011 1355164007 1355165006 1355165019 1355224004 1355224011 1355224018 1355224027 1355225006 1355225018 1355285001 1355285008 1355344002 1355344010 1355345007 1355345013 1355346005 1355346011 1355347004 1355347010 1355404009 1355404017 1355404024 1355405005 1355405019 1355405028 1355407005 1355407015 1355407023 1355467003 1355467010 1355468002 1355468009 1355469006 1355469014 1355469021 1355470002 1355470009 1355470016 1355470024 1355471002 1355471009 1355472003 1355472012 1355472021 1355473001 1355473008 1355473015 1355473026 1355474007 1355475003 1355475012 1355526005 1355527004 1355528003 1355528011 1355528021 1355588002 1355588008 1355589009 1355589017 1355590001 1355590009 1355591001 1355591008 1355591014 1355592012 1355649003 1355649009 1355710003 1355710009 1355710017 1355710024 1355711004 1355771003 1355771013 1355830004 1355830010 1355830016 1355830025 1355831011 1355831020 1355831030 1355832008 1355832015 1355891005 1355929003 1355930001 1355930007 1355930014 1355931007 1355931013 1355931020 1355931027 1355932005 1355932014 1355932023 1355933002 1355933011 1355934005 1355934011 1355934022 1355989005 1355989012 1355989023 1355990002 1355990014 1356048009 1356048015 1356048023 1356048030 1356049007 1356049013 1356049020 1356050001 1356050007 1356050014 1356053006 1356053013 1356054007 1356054015 1356055006 1356056002 1356056009 1356109003 1356109010 1356110001 1356110008 1356169003 1356220002 1356220009 1356220017 1356220024 1356221004 1356221010 1356221017 1356221024 1356222001 1356222009 1356222018 1356222027 1356281005 1356281011 1356282004 1356282013 1356282020 1356282026 1356283003 1356283011 1356498001 1356498007 1356498013 1356499006 1356499013 1356501009 1356560006 1356560015 1356560025 1356561005 1356562003 1356562009 1356562019 1356562025 1356563003 1356563009 1356564002 1356564009 1356564018 1356564025 1356565003 1356565010 1356565016 1356565023 1356565029 1356566007 1356566016 1356566023 1356566031 1356619010 1356619019 1356619029 1356619040 1356621001 1356621007 1356622011 1356623010 1356623018 1356681006 1356681012 1356681018 1356681025 1356682004 1356682010 1356683001 1356683007 1356683013 1356683019 1356683026 1356684004 1356684011 1356685003 1356685011 1356685019 1356742001 1356742008 1356803007 1356803014 1356804005 1356805002 1356805009 1356805015 1356805023 1356805029 1356806006 1356806013 1356806025 1356807002 1356807011 1356807017 1356807025 1356808006 1356808013 1356808019 1356808026 1356863005 1356863013 1356863024 1356864004 1356864011 1356926002 1356926008 1356926015 1356926021 1356927005 1356927011 1356927020 1356928006 1356928016 1356928023 1356928029 1356929011 1356929019 1356929029 1356987007 1356987014 1356988009 1356989001 1356989008 1463073003 1463073010 1463074008 1463075005 1463076001 1463076010 1463077007 1469330005 1469330015 1469330024 1472915007 1472915014 1472916005 1472916013 1472917004 1472917011 1472917018 1472917025 1472921007 1472921017 1472921028 1472922009 1472924005 1472927002 1472928002 1472928009 1472928015 1473420007 1473420014 1473420020 1473690001 1473690007 1473690014 1473690020 1473690026 1473692005 1473692011 1473692017 1473692023 1473996004 1473996011 1474001002 1474001010 1474667001 1474667012 1474668005 1474668013 1476187005 1476187014 1476188008 1477297001 1477298004 1481015009 1481015015 1484273008 1484273014 1484277008 1484284001 1484284009 1484286001 1484286010 1484291005 1484291014 1484293005 1484293015 1484293024 1484295006 1484295018 1484295027 1484320006 1484320020 1484322001 1484322010 1484323009 1484323015 1484323023 1484324003 1484324011 1484325006 1484325014 1484338015 1484339010 1484340004 1484340013 1484344003 1484345002 1484345008 1488074001 1488074013 1488074028 1488577011 1488577018 1488577026 1488599002 1488599013 1488599022 1491487004 1491487011 1491488006 1491488016 1491488022 1491488029 1491489005 1491489011 1491491008 1491499009 1491499021 1491499027 1491501005 1491501013 1494974002 1494983003 1494985003 1494985009 1496165001 1496165008 1497834011 1497834019 1497834027 1498104009 1498104015 1498872002 1498872010 1499080001 1499080009 1499615003 1499615011 1499615022 1499830005 1499830014 1499830021 1501763004 1514717001 1514717009 1514898012 1515154010 1515186002 1515186013 1515334007 1515334015 1515334022 1515687005 1515687012 1516141008 1516141015 1516141022 1516141030 1516141041 1516630008 1516630018 1516641001 1516641013 1517063013 1517063026 1517160001 1517160011 1517329010 1517412008 1517435007 1517845005 1517849003 1517849009 1518008008 1518580004 1518580012 1519185002 1519220002 1519220011 1519260002 1519368002 1519368011 1519369008 1519370003 1519685003 1519685013 1519685022 1519685029 1519908005 1519908015 1519918009 1519942007 1519966007 1519966017 1520117003 1521503010 1521994002 1521994011 1522138010 1522145002 1522145013 1522160005 1522160011 1522241009 1522447004 1522567002 1522567011 1524364014 1524365007 1524365014 1524366011 1524556011 1526681002 1526681010 1526844003 1526848003 1528890002 1531324002 1538230009 1545444005 1547526005 1547526013 1547629012 1551806004 1551806013 1553666004 1553712004 1553712011 1553758009 1553760005 1556365012 1560273007 1560273015 1562392001 1562392008 1562393006 1562573002 1562573009 1562589006 1562591002 1564482002 1564482014 1568632004 1568633003 1568633009 1575125003 1575315010 1575320003 1577574005 1577575001 1577575013 1578069005 1578332002 1578333001 1578333015 1578544001 1578544009 1578810005 1578810012 1579488008 1579613014 1579615011 1579618006 1579619005 1579619015 1579781002 1579967001 1579972013 1580162007 1580289012 1580540010 1580601009 1580675002 1581019001 1581294008 1581296003 1581296013 1581549005 1581549015 1581647015 1585827007 1585831003 1586082003 1586090007 1586134005 1586286014 1586982005 1587881006 1587884002 1588001005 1588006015 1588493009 1588526012 1588687009 1588774010 1588778013 1590360012 1590565001 1590587013 1590751011 1590769010 1591097009 1591279005 1591279013 1591670013 1592469R004 1593048R020 1593190r015 1593387R012 1593407R017 1593517R017 1594191R017 1594391011 1594558004 1594769001 1594920006 1595287001 1595477002 1595571001 1595675007 1595767002 1596271006 1596421008 1596540011 1596669010 1599015003 1599354003 1601730005 1601733019 1602400012 1602401014 1604607022 1604740013 1607095005 Lot Codes 1354736001 1354736008 1354737006 1354738003 1354739003 1354739009 1354739022 1354740001 1354740012 1354745006 1354798003 1354798009 1354798015 1354799005 1354800002 1354800009 1354801001 1354801007 1354801014 1354802005 1354803001 1354804009 1354806002 1354806008 1354806020 1354806027 1354807004 1354807010 1354808004 1354809001 1354809010 1354810002 1354810009 1354810016 1354810023 1354859003 1354859010 1354859018 1354859026 1354860006 1354861002 1354861009 1354861015 1354861022 1354861029 1354862005 1354862012 1354862019 1354862029 1354863008 1354865010 1354865018 1354865026 1354867005 1354867013 1354867021 1354867029 1354921007 1354922003 1354922011 1354922019 1354922025 1354923003 1354923010 1354923016 1354923022 1354924004 1354924013 1354924019 1354924028 1354925007 1354925013 1354925019 1354925025 1354926006 1354926014 1354926021 1354926027 1354927012 1354927020 1354927030 1354928009 1354929006 1354929012 1354980006 1354981002 1354981010 1354982004 1354982011 1354985002 1355041006 1355041013 1355041021 1355042002 1355042009 1355042016 1355042022 1355043003 1355043011 1355044005 1355044015 1355044024 1355047009 1355102003 1355102009 1355103001 1355103010 1355103016 1355103022 1355103030 1355104007 1355104018 1355104025 1355105008 1355106002 1355106008 1355106014 1355162006 1355163006 1355163012 1355164008 1355165007 1355165020 1355224005 1355224013 1355224019 1355224028 1355225008 1355225021 1355285003 1355285009 1355344003 1355345001 1355345008 1355345014 1355346006 1355346012 1355347005 1355404002 1355404011 1355404018 1355404025 1355405007 1355405021 1355405030 1355407007 1355407016 1355407024 1355467004 1355467011 1355468003 1355469001 1355469007 1355469015 1355469022 1355470004 1355470010 1355470019 1355470025 1355471003 1355471011 1355472004 1355472014 1355472023 1355473002 1355473010 1355473016 1355473027 1355474010 1355475004 1355475013 1355526006 1355527006 1355528004 1355528014 1355528022 1355588003 1355589002 1355589010 1355589019 1355590003 1355590010 1355591002 1355591009 1355592005 1355592013 1355649004 1355649010 1355710004 1355710010 1355710019 1355710026 1355711005 1355771004 1355772010 1355830005 1355830011 1355830017 1355831002 1355831012 1355831021 1355832001 1355832009 1355890001 1355891010 1355929005 1355930002 1355930009 1355930015 1355931008 1355931014 1355931021 1355931028 1355932006 1355932015 1355932024 1355933003 1355933014 1355934006 1355934013 1355934026 1355989006 1355989014 1355989025 1355990006 1355990015 1356048010 1356048017 1356048024 1356049001 1356049008 1356049014 1356049021 1356050002 1356050008 1356050015 1356053007 1356053015 1356054009 1356054016 1356055007 1356056003 1356056011 1356109004 1356109011 1356110003 1356110009 1356169008 1356220003 1356220010 1356220018 1356220026 1356221005 1356221011 1356221018 1356221025 1356222002 1356222010 1356222019 1356222028 1356281006 1356281013 1356282005 1356282014 1356282021 1356282027 1356283004 1356283012 1356498002 1356498008 1356498014 1356499007 1356499014 1356501013 1356560008 1356560018 1356560026 1356561006 1356562004 1356562010 1356562020 1356562026 1356563004 1356563010 1356564003 1356564011 1356564019 1356564026 1356565004 1356565011 1356565018 1356565024 1356565030 1356566008 1356566017 1356566025 1356619001 1356619011 1356619021 1356619031 1356620001 1356621002 1356621008 1356622012 1356623011 1356623020 1356681007 1356681013 1356681019 1356681026 1356682005 1356682011 1356683002 1356683008 1356683014 1356683020 1356683027 1356684005 1356684012 1356685005 1356685012 1356685020 1356742002 1356803001 1356803008 1356803015 1356804006 1356805003 1356805010 1356805016 1356805024 1356806001 1356806007 1356806017 1356806026 1356807004 1356807012 1356807018 1356807026 1356808007 1356808014 1356808020 1356808027 1356863006 1356863016 1356863027 1356864005 1356864012 1356926003 1356926009 1356926016 1356926022 1356927006 1356927014 1356928001 1356928008 1356928017 1356928024 1356928030 1356929013 1356929020 1356987001 1356987008 1356988002 1356988010 1356989002 1356989011 1463073004 1463074001 1463074009 1463075006 1463076004 1463077001 1463077008 1469330007 1469330016 1472915001 1472915009 1472915015 1472916006 1472916014 1472917005 1472917012 1472917019 1472917026 1472921008 1472921019 1472921029 1472922010 1472924006 1472927004 1472928003 1472928010 1473420001 1473420009 1473420015 1473420021 1473690002 1473690008 1473690015 1473690021 1473690027 1473692006 1473692012 1473692018 1473692025 1473996005 1473996012 1474001003 1474001011 1474667002 1474667013 1474668006 1474668014 1476187007 1476188001 1476188010 1477297002 1481015001 1481015010 1484273001 1484273009 1484277001 1484277009 1484284003 1484284010 1484286002 1484286011 1484291006 1484291015 1484293006 1484293017 1484293025 1484295008 1484295019 1484295028 1484320007 1484320021 1484322002 1484322015 1484323010 1484323017 1484323024 1484324004 1484324012 1484325009 1484338003 1484339003 1484339012 1484340005 1484340014 1484344004 1484345003 1484345010 1488074003 1488074022 1488577001 1488577012 1488577019 1488577027 1488599003 1488599014 1488599023 1491487005 1491487013 1491488007 1491488017 1491488023 1491488030 1491489006 1491489012 1491499003 1491499010 1491499022 1491499028 1491501006 1491501014 1494974003 1494983004 1494985004 1494985011 1496165002 1497834001 1497834013 1497834020 1497834028 1498104010 1498104016 1498872004 1498872011 1499080002 1499080011 1499615005 1499615012 1499615027 1499830006 1499830015 1499830022 1501763005 1514717003 1514717011 1515154001 1515154011 1515186005 1515186014 1515334008 1515334017 1515334023 1515687007 1515687014 1516141009 1516141016 1516141023 1516141031 1516630001 1516630009 1516630021 1516641003 1516641014 1517063015 1517063027 1517160002 1517160012 1517329011 1517435002 1517435008 1517845006 1517849004 1517849011 1518008009 1518580005 1518580013 1519185004 1519220003 1519220012 1519260003 1519368003 1519368012 1519369010 1519370004 1519685004 1519685014 1519685023 1519893001 1519908006 1519918001 1519918010 1519942009 1519966008 1519966018 1521503001 1521503012 1521994003 1521994013 1522138011 1522145003 1522145014 1522160006 1522160012 1522241011 1522447007 1522567003 1522567013 1524365001 1524365008 1524365015 1524556003 1524556012 1526681003 1526681011 1526844007 1526848005 1528890007 1531324004 1543262001 1545444006 1547526006 1547526014 1547629014 1551806005 1551806015 1553666005 1553712005 1553712013 1553758010 1553760006 1556365013 1560273008 1560274001 1562392002 1562392009 1562393012 1562573003 1562573012 1562589007 1562591003 1564482004 1566566004 1568632007 1568633004 1575122002 1575125006 1575315011 1575320005 1577574007 1577575005 1577575014 1578069008 1578332003 1578333002 1578340013 1578544002 1578544010 1578810006 1578810013 1579488012 1579615001 1579615013 1579618008 1579619008 1579770003 1579781003 1579967002 1579972015 1580162012 1580289015 1580540011 1580601013 1580675003 1581019004 1581294009 1581296004 1581296015 1581549006 1581647001 1585766007 1585827011 1585831005 1586082004 1586090014 1586134006 1586297002 1586982009 1587881009 1587884004 1588001007 1588487008 1588493013 1588526013 1588774001 1588774012 1590356003 1590360013 1590565006 1590587014 1590769003 1590769012 1591097015 1591279006 1591279014 1592089R006 1592469R010 1593107R002 1593190R019 1593387R014 1593517R001 1593517R018 1594192R001 1594391015 1594558005 1594769002 1594920007 1595287007 1595477004 1595571006 1595698003 1595767009 1596271012 1596421009 1596540013 1596669012 1599015008 1599354004 1601731004 1601735004 1602400014 1604378002 1604607028 1606029005 1607095008 Lot Codes 1354736003 1354737001 1354737007 1354738004 1354739004 1354739012 1354739023 1354740003 1354745001 1354745008 1354798004 1354798010 1354798016 1354799006 1354800003 1354800010 1354801002 1354801009 1354801015 1354802006 1354803002 1354804010 1354806003 1354806009 1354806021 1354806028 1354807005 1354807011 1354808005 1354809002 1354809011 1354810003 1354810010 1354810017 1354810024 1354859005 1354859011 1354859019 1354860001 1354860007 1354861003 1354861010 1354861016 1354861023 1354861030 1354862006 1354862013 1354862021 1354863002 1354863010 1354865013 1354865019 1354865028 1354867006 1354867014 1354867022 1354867030 1354921008 1354922004 1354922012 1354922020 1354922026 1354923004 1354923011 1354923017 1354923023 1354924005 1354924014 1354924020 1354924029 1354925008 1354925014 1354925020 1354925028 1354926007 1354926016 1354926022 1354927001 1354927013 1354927023 1354928002 1354928010 1354929007 1354929013 1354980008 1354981003 1354981011 1354982006 1354982012 1354985004 1355041007 1355041015 1355041022 1355042003 1355042010 1355042017 1355042023 1355043004 1355043012 1355044007 1355044016 1355044025 1355047011 1355102004 1355102010 1355103003 1355103011 1355103017 1355103023 1355104001 1355104010 1355104019 1355104026 1355105009 1355106003 1355106009 1355106015 1355163001 1355163007 1355163014 1355164009 1355165009 1355165022 1355224006 1355224014 1355224020 1355224029 1355225009 1355225022 1355285004 1355285011 1355344004 1355345002 1355345009 1355345015 1355346007 1355346013 1355347006 1355404004 1355404012 1355404019 1355405001 1355405010 1355405023 1355407001 1355407009 1355407017 1355407025 1355467005 1355467012 1355468005 1355469002 1355469008 1355469016 1355469023 1355470005 1355470012 1355470020 1355470026 1355471005 1355471012 1355472005 1355472015 1355472024 1355473003 1355473011 1355473019 1355474001 1355474011 1355475007 1355475014 1355526009 1355527008 1355528005 1355528015 1355528023 1355588004 1355589004 1355589011 1355589022 1355590004 1355590011 1355591003 1355591010 1355592006 1355592014 1355649005 1355650002 1355710005 1355710011 1355710020 1355710027 1355711006 1355771006 1355772013 1355830006 1355830012 1355830019 1355831004 1355831013 1355831023 1355832003 1355832010 1355890002 1355892001 1355929006 1355930003 1355930010 1355931002 1355931009 1355931015 1355931022 1355931029 1355932008 1355932018 1355932025 1355933004 1355934001 1355934007 1355934014 1355934027 1355989008 1355989018 1355989026 1355990007 1356048003 1356048011 1356048018 1356048025 1356049002 1356049009 1356049016 1356049024 1356050003 1356050009 1356053001 1356053009 1356054002 1356054010 1356054017 1356055008 1356056004 1356056013 1356109005 1356109012 1356110004 1356110010 1356169012 1356220005 1356220011 1356220019 1356220027 1356221006 1356221012 1356221019 1356221026 1356222004 1356222011 1356222020 1356281001 1356281007 1356281014 1356282006 1356282015 1356282022 1356282028 1356283005 1356283015 1356498003 1356498009 1356499002 1356499008 1356501002 1356560001 1356560009 1356560020 1356560027 1356561007 1356562005 1356562011 1356562021 1356562027 1356563005 1356563011 1356564004 1356564013 1356564020 1356564027 1356565005 1356565012 1356565019 1356565025 1356566001 1356566010 1356566018 1356566027 1356619002 1356619013 1356619022 1356619033 1356620002 1356621003 1356621009 1356622014 1356623013 1356681001 1356681008 1356681014 1356681020 1356681027 1356682006 1356682012 1356683003 1356683009 1356683015 1356683021 1356683028 1356684006 1356684013 1356685006 1356685014 1356685021 1356742003 1356803002 1356803009 1356804001 1356804007 1356805005 1356805011 1356805017 1356805025 1356806002 1356806008 1356806018 1356806027 1356807005 1356807013 1356807021 1356808001 1356808008 1356808015 1356808021 1356808028 1356863007 1356863017 1356863029 1356864006 1356864013 1356926004 1356926010 1356926017 1356927001 1356927007 1356927015 1356928002 1356928009 1356928018 1356928025 1356929002 1356929015 1356929021 1356987002 1356987010 1356988003 1356988011 1356989003 1356989013 1463073005 1463074002 1463075001 1463075007 1463076005 1463077002 1463077010 1469330008 1469330017 1472915002 1472915010 1472916001 1472916007 1472916015 1472917006 1472917013 1472917020 1472921001 1472921010 1472921024 1472922004 1472922011 1472924008 1472927005 1472928004 1472928011 1473420002 1473420010 1473420016 1473420022 1473690003 1473690009 1473690016 1473690022 1473690029 1473692007 1473692013 1473692019 1473692026 1473996006 1473996013 1474001005 1474001012 1474667004 1474667014 1474668008 1474668015 1476187008 1476188003 1476188011 1477297003 1481015002 1481015011 1484273004 1484273010 1484277003 1484277011 1484284004 1484285002 1484286003 1484286012 1484291007 1484293001 1484293008 1484293018 1484293026 1484295010 1484295020 1484295029 1484320008 1484320022 1484322005 1484323001 1484323011 1484323018 1484323025 1484324005 1484324013 1484325010 1484338007 1484339004 1484339013 1484340006 1484343001 1484344008 1484345004 1484345014 1488074006 1488074024 1488577003 1488577014 1488577020 1488577028 1488599004 1488599017 1488599025 1491487006 1491487014 1491488008 1491488018 1491488024 1491489001 1491489007 1491489013 1491499004 1491499013 1491499023 1491499029 1491501007 1491501018 1494974004 1494983005 1494985005 1494985012 1496165003 1497834004 1497834014 1497834021 1498104001 1498104011 1498104018 1498872005 1498872012 1499080003 1499170001 1499615006 1499615013 1499615028 1499830007 1499830016 1499830023 1511181002 1514717004 1514898002 1515154004 1515154012 1515186006 1515186015 1515334009 1515334018 1515334024 1515687008 1516141002 1516141010 1516141018 1516141024 1516141032 1516630002 1516630012 1516630023 1516641007 1517063005 1517063019 1517063029 1517160004 1517160013 1517329014 1517435003 1517435010 1517845010 1517849005 1517849012 1518008012 1518580006 1518580014 1519185006 1519220005 1519220013 1519260005 1519368004 1519369001 1519369011 1519370006 1519685005 1519685015 1519685024 1519893002 1519908007 1519918002 1519918015 1519942010 1519966009 1519966020 1521503002 1521992001 1521994005 1521994015 1522138012 1522145004 1522145015 1522160007 1522160013 1522241012 1522447008 1522567004 1522567014 1524365002 1524365009 1524366003 1524556005 1524556013 1526681004 1526681012 1526844011 1526848008 1528890011 1538213004 1543262002 1545444007 1547526008 1547629003 1547629015 1551806006 1553665001 1553666007 1553712006 1553712015 1553758012 1556365004 1556365014 1560273010 1560274002 1562392003 1562392012 1562426001 1562573005 1562589002 1562589009 1562591004 1564482007 1566566007 1568632009 1568633005 1575122004 1575125007 1575315014 1575320006 1577574008 1577575008 1577575016 1578331002 1578332004 1578333003 1578543006 1578544003 1578544014 1578810007 1579488002 1579488013 1579615004 1579615014 1579618009 1579619009 1579770007 1579781007 1579967007 1579973002 1580188005 1580291002 1580540012 1580671006 1580675004 1581019008 1581294010 1581296006 1581505001 1581549007 1581647002 1585766008 1585827012 1585831006 1586082005 1586132001 1586134009 1586297009 1586982013 1587881011 1587884006 1588006002 1588487009 1588526001 1588687003 1588774003 1588774014 1590356005 1590360016 1590565010 1590751001 1590769005 1590769013 1591128009 1591279009 1591670002 1592089R011 1592469R017 1593107R005 1593387R002 1593387R015 1593517R002 1593674R014 1594192R002 1594392005 1594558008 1594769007 1594920008 1595287009 1595477005 1595672002 1595698004 1595767014 1596320002 1596540001 1596669001 1596669013 1599015011 1599397012 1601731009 1601735015 1602401001 1604378005 1604607029 1606169001 1607095009 Lot Codes 1354736004 1354737002 1354737008 1354738005 1354739005 1354739013 1354739024 1354740005 1354745002 1354745010 1354798005 1354798011 1354799001 1354799007 1354800004 1354800011 1354801003 1354801010 1354802001 1354802007 1354803003 1354804012 1354806004 1354806010 1354806023 1354806029 1354807006 1354807013 1354808006 1354809003 1354809012 1354810004 1354810011 1354810018 1354810025 1354859006 1354859012 1354859022 1354860002 1354860008 1354861004 1354861011 1354861017 1354861024 1354862001 1354862007 1354862014 1354862022 1354863003 1354863011 1354865014 1354865020 1354865029 1354867008 1354867015 1354867024 1354921001 1354921009 1354922007 1354922013 1354922021 1354922027 1354923005 1354923012 1354923018 1354923024 1354924006 1354924015 1354924021 1354925001 1354925009 1354925015 1354925021 1354926002 1354926008 1354926017 1354926023 1354927003 1354927014 1354927024 1354928003 1354929001 1354929008 1354980001 1354980011 1354981004 1354981013 1354982007 1354982014 1354985005 1355041008 1355041016 1355041023 1355042005 1355042012 1355042018 1355042024 1355043005 1355043013 1355044009 1355044017 1355047001 1355047012 1355102005 1355102011 1355103004 1355103012 1355103018 1355103024 1355104002 1355104013 1355104020 1355105001 1355105010 1355106004 1355106010 1355162001 1355163002 1355163008 1355163015 1355164014 1355165014 1355165023 1355224007 1355224015 1355224021 1355224030 1355225012 1355225024 1355285005 1355285012 1355344007 1355345003 1355345010 1355346002 1355346008 1355346014 1355347007 1355404005 1355404013 1355404021 1355405002 1355405011 1355405025 1355407002 1355407012 1355407018 1355407026 1355467006 1355467013 1355468006 1355469003 1355469010 1355469017 1355469024 1355470006 1355470013 1355470021 1355470027 1355471006 1355471013 1355472007 1355472016 1355472025 1355473004 1355473012 1355473020 1355474003 1355474012 1355475009 1355526001 1355526010 1355527011 1355528007 1355528016 1355528025 1355588005 1355589005 1355589013 1355589025 1355590006 1355590012 1355591005 1355591011 1355592007 1355592015 1355649006 1355650004 1355710006 1355710013 1355710021 1355710028 1355711007 1355771007 1355772014 1355830007 1355830013 1355830021 1355831007 1355831014 1355831024 1355832004 1355832011 1355890003 1355892007 1355929007 1355930004 1355930011 1355931003 1355931010 1355931016 1355931023 1355932002 1355932011 1355932019 1355932027 1355933005 1355934002 1355934008 1355934015 1355934029 1355989009 1355989019 1355989028 1355990008 1356048004 1356048012 1356048019 1356048026 1356049003 1356049010 1356049017 1356049025 1356050004 1356050010 1356053002 1356053010 1356054004 1356054011 1356055002 1356055009 1356056005 1356056014 1356109006 1356109013 1356110005 1356110011 1356192001 1356220006 1356220012 1356220021 1356220028 1356221007 1356221013 1356221020 1356221028 1356222005 1356222015 1356222022 1356281002 1356281008 1356282001 1356282007 1356282016 1356282023 1356282029 1356283008 1356283018 1356498004 1356498010 1356499003 1356499009 1356501003 1356560002 1356560010 1356560021 1356561001 1356561008 1356562006 1356562014 1356562022 1356562028 1356563006 1356563012 1356564005 1356564014 1356564021 1356564028 1356565006 1356565013 1356565020 1356565026 1356566002 1356566011 1356566019 1356566028 1356619003 1356619015 1356619023 1356619034 1356620003 1356621004 1356621010 1356623003 1356623014 1356681002 1356681009 1356681015 1356681021 1356682001 1356682007 1356682013 1356683004 1356683010 1356683016 1356683022 1356683029 1356684007 1356684014 1356685008 1356685015 1356685022 1356742004 1356803004 1356803010 1356804002 1356804008 1356805006 1356805012 1356805018 1356805026 1356806003 1356806009 1356806019 1356806030 1356807008 1356807014 1356807022 1356808002 1356808009 1356808016 1356808023 1356863001 1356863009 1356863018 1356863030 1356864008 1356864014 1356926005 1356926011 1356926018 1356927002 1356927008 1356927016 1356928003 1356928010 1356928019 1356928026 1356929004 1356929016 1356929023 1356987003 1356987011 1356988004 1356988012 1356989004 1356989014 1463073006 1463074003 1463075002 1463075008 1463076007 1463077004 1469330002 1469330010 1469330018 1472915003 1472915011 1472916002 1472916008 1472917001 1472917007 1472917014 1472917021 1472921003 1472921011 1472921025 1472922005 1472924001 1472924011 1472927007 1472928005 1472928012 1473420003 1473420011 1473420017 1473420023 1473690004 1473690010 1473690017 1473690023 1473692001 1473692008 1473692014 1473692020 1473692027 1473996007 1473996014 1474001006 1474001013 1474667005 1474667015 1474668009 1476187001 1476187011 1476188004 1476189002 1477297004 1481015004 1481015012 1484273005 1484273011 1484277004 1484277012 1484284005 1484285004 1484286004 1484286015 1484291009 1484293002 1484293009 1484293019 1484293028 1484295011 1484295021 1484320001 1484320009 1484320023 1484322006 1484323002 1484323012 1484323019 1484323026 1484324006 1484324014 1484325011 1484338010 1484339005 1484339014 1484340009 1484343004 1484344014 1484345005 1484656007 1488074007 1488074025 1488577006 1488577015 1488577021 1488577029 1488599006 1488599018 1488599027 1491487007 1491487015 1491488009 1491488019 1491488025 1491489002 1491489008 1491489014 1491499005 1491499015 1491499024 1491501002 1491501008 1491501020 1494974006 1494983006 1494985006 1494985013 1496165005 1497834005 1497834015 1497834023 1498104005 1498104012 1498104019 1498872007 1498872013 1499080004 1499170002 1499615007 1499615015 1499830002 1499830010 1499830017 1499830025 1511181003 1514717005 1514898003 1515154005 1515154013 1515186009 1515334001 1515334010 1515334019 1515334025 1515687009 1516141004 1516141011 1516141019 1516141027 1516141034 1516630003 1516630015 1516630025 1516641009 1517063008 1517063022 1517063030 1517160005 1517160014 1517412001 1517435004 1517435011 1517845011 1517849006 1518008005 1518580001 1518580007 1518970004 1519185007 1519220006 1519220014 1519260008 1519368005 1519369002 1519369013 1519370008 1519685008 1519685016 1519685026 1519893009 1519908009 1519918003 1519942001 1519942013 1519966010 1520115001 1521503003 1521992006 1521994006 1522138003 1522138013 1522145005 1522160001 1522160008 1522241002 1522241013 1522447009 1522567008 1524364001 1524365003 1524365010 1524366005 1524556007 1524556014 1526681005 1526681013 1526844012 1526848009 1528890014 1538213014 1545444001 1545444008 1547526009 1547629004 1551806001 1551806007 1553665003 1553666012 1553712007 1553758004 1553758014 1556365006 1560273004 1560273011 1560274003 1562392004 1562392013 1562426012 1562573006 1562589003 1562589013 1562591006 1564482008 1566566008 1568632013 1568633006 1575122005 1575125011 1575315015 1575320010 1577574010 1577575009 1578069002 1578331003 1578332007 1578333005 1578543007 1578544005 1578544015 1578810008 1579488003 1579613003 1579615006 1579615015 1579618010 1579619012 1579770012 1579781008 1579967014 1579973003 1580188009 1580291003 1580583001 1580671009 1580675010 1581294003 1581294011 1581296010 1581549001 1581549008 1581647004 1585766010 158
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Endologix
- Reason for Recall:
- Endologix updated the Instructions for Use (IFU) and implemented modifications to the AFX Endovascular AAA System (AFX System).
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
AFX Endovascular AAA System, Endoleak Type: IIIB Common Name: AFX Bifurcated and Accessory Stent Grafts Classification Name: System, Endovascular Graft, Aortic Aneurysm Treatment The AFX Endovascular AAA System is intended for endovascular treatment of patients with abdominal aortic or aortoiliac aneurysms (AAA).
Product Codes/Lot Numbers:
"AFX2 Duraply Lot Codes (July 2015 - Nov. 2016)" Lot Codes 1298056012 1354736007 1354737005 1354738002 1354739002 1354739008 1354739020 1354739027 1354740011 1354745005 1354798002 1354798008 1354798014 1354799004 1354800001 1354800008 1354800014 1354801006 1354801013 1354802004 1354802010 1354804006 1354806001 1354806007 1354806019 1354806026 1354807003 1354807009 1354808003 1354808009 1354809009 1354809015 1354810008 1354810015 1354810022 1354859002 1354859009 1354859016 1354859025 1354860005 1354861001 1354861008 1354861014 1354861021 1354861027 1354862004 1354862011 1354862017 1354862027 1354863007 1354865009 1354865017 1354865025 1354867004 1354867012 1354867020 1354867028 1354921006 1354922002 1354922010 1354922018 1354922024 1354923002 1354923009 1354923015 1354923021 1354924003 1354924012 1354924018 1354924026 1354925005 1354925012 1354925018 1354925024 1354926005 1354926013 1354926020 1354926026 1354927006 1354927019 1354927029 1354928007 1354929005 1354929011 1354980004 1354981001 1354981008 1354982003 1354982010 1354983015 1355041003 1355041011 1355041020 1355042001 1355042008 1355042015 1355042021 1355042027 1355043010 1355044004 1355044013 1355044023 1355047007 1355102002 1355102008 1355102014 1355103009 1355103015 1355103021 1355103029 1355104006 1355104016 1355104024 1355105007 1355106001 1355106007 1355106013 1355162005 1355163005 1355163011 1355164007 1355165006 1355165019 1355224004 1355224011 1355224018 1355224027 1355225006 1355225018 1355285001 1355285008 1355344002 1355344010 1355345007 1355345013 1355346005 1355346011 1355347004 1355347010 1355404009 1355404017 1355404024 1355405005 1355405019 1355405028 1355407005 1355407015 1355407023 1355467003 1355467010 1355468002 1355468009 1355469006 1355469014 1355469021 1355470002 1355470009 1355470016 1355470024 1355471002 1355471009 1355472003 1355472012 1355472021 1355473001 1355473008 1355473015 1355473026 1355474007 1355475003 1355475012 1355526005 1355527004 1355528003 1355528011 1355528021 1355588002 1355588008 1355589009 1355589017 1355590001 1355590009 1355591001 1355591008 1355591014 1355592012 1355649003 1355649009 1355710003 1355710009 1355710017 1355710024 1355711004 1355771003 1355771013 1355830004 1355830010 1355830016 1355830025 1355831011 1355831020 1355831030 1355832008 1355832015 1355891005 1355929003 1355930001 1355930007 1355930014 1355931007 1355931013 1355931020 1355931027 1355932005 1355932014 1355932023 1355933002 1355933011 1355934005 1355934011 1355934022 1355989005 1355989012 1355989023 1355990002 1355990014 1356048009 1356048015 1356048023 1356048030 1356049007 1356049013 1356049020 1356050001 1356050007 1356050014 1356053006 1356053013 1356054007 1356054015 1356055006 1356056002 1356056009 1356109003 1356109010 1356110001 1356110008 1356169003 1356220002 1356220009 1356220017 1356220024 1356221004 1356221010 1356221017 1356221024 1356222001 1356222009 1356222018 1356222027 1356281005 1356281011 1356282004 1356282013 1356282020 1356282026 1356283003 1356283011 1356498001 1356498007 1356498013 1356499006 1356499013 1356501009 1356560006 1356560015 1356560025 1356561005 1356562003 1356562009 1356562019 1356562025 1356563003 1356563009 1356564002 1356564009 1356564018 1356564025 1356565003 1356565010 1356565016 1356565023 1356565029 1356566007 1356566016 1356566023 1356566031 1356619010 1356619019 1356619029 1356619040 1356621001 1356621007 1356622011 1356623010 1356623018 1356681006 1356681012 1356681018 1356681025 1356682004 1356682010 1356683001 1356683007 1356683013 1356683019 1356683026 1356684004 1356684011 1356685003 1356685011 1356685019 1356742001 1356742008 1356803007 1356803014 1356804005 1356805002 1356805009 1356805015 1356805023 1356805029 1356806006 1356806013 1356806025 1356807002 1356807011 1356807017 1356807025 1356808006 1356808013 1356808019 1356808026 1356863005 1356863013 1356863024 1356864004 1356864011 1356926002 1356926008 1356926015 1356926021 1356927005 1356927011 1356927020 1356928006 1356928016 1356928023 1356928029 1356929011 1356929019 1356929029 1356987007 1356987014 1356988009 1356989001 1356989008 1463073003 1463073010 1463074008 1463075005 1463076001 1463076010 1463077007 1469330005 1469330015 1469330024 1472915007 1472915014 1472916005 1472916013 1472917004 1472917011 1472917018 1472917025 1472921007 1472921017 1472921028 1472922009 1472924005 1472927002 1472928002 1472928009 1472928015 1473420007 1473420014 1473420020 1473690001 1473690007 1473690014 1473690020 1473690026 1473692005 1473692011 1473692017 1473692023 1473996004 1473996011 1474001002 1474001010 1474667001 1474667012 1474668005 1474668013 1476187005 1476187014 1476188008 1477297001 1477298004 1481015009 1481015015 1484273008 1484273014 1484277008 1484284001 1484284009 1484286001 1484286010 1484291005 1484291014 1484293005 1484293015 1484293024 1484295006 1484295018 1484295027 1484320006 1484320020 1484322001 1484322010 1484323009 1484323015 1484323023 1484324003 1484324011 1484325006 1484325014 1484338015 1484339010 1484340004 1484340013 1484344003 1484345002 1484345008 1488074001 1488074013 1488074028 1488577011 1488577018 1488577026 1488599002 1488599013 1488599022 1491487004 1491487011 1491488006 1491488016 1491488022 1491488029 1491489005 1491489011 1491491008 1491499009 1491499021 1491499027 1491501005 1491501013 1494974002 1494983003 1494985003 1494985009 1496165001 1496165008 1497834011 1497834019 1497834027 1498104009 1498104015 1498872002 1498872010 1499080001 1499080009 1499615003 1499615011 1499615022 1499830005 1499830014 1499830021 1501763004 1514717001 1514717009 1514898012 1515154010 1515186002 1515186013 1515334007 1515334015 1515334022 1515687005 1515687012 1516141008 1516141015 1516141022 1516141030 1516141041 1516630008 1516630018 1516641001 1516641013 1517063013 1517063026 1517160001 1517160011 1517329010 1517412008 1517435007 1517845005 1517849003 1517849009 1518008008 1518580004 1518580012 1519185002 1519220002 1519220011 1519260002 1519368002 1519368011 1519369008 1519370003 1519685003 1519685013 1519685022 1519685029 1519908005 1519908015 1519918009 1519942007 1519966007 1519966017 1520117003 1521503010 1521994002 1521994011 1522138010 1522145002 1522145013 1522160005 1522160011 1522241009 1522447004 1522567002 1522567011 1524364014 1524365007 1524365014 1524366011 1524556011 1526681002 1526681010 1526844003 1526848003 1528890002 1531324002 1538230009 1545444005 1547526005 1547526013 1547629012 1551806004 1551806013 1553666004 1553712004 1553712011 1553758009 1553760005 1556365012 1560273007 1560273015 1562392001 1562392008 1562393006 1562573002 1562573009 1562589006 1562591002 1564482002 1564482014 1568632004 1568633003 1568633009 1575125003 1575315010 1575320003 1577574005 1577575001 1577575013 1578069005 1578332002 1578333001 1578333015 1578544001 1578544009 1578810005 1578810012 1579488008 1579613014 1579615011 1579618006 1579619005 1579619015 1579781002 1579967001 1579972013 1580162007 1580289012 1580540010 1580601009 1580675002 1581019001 1581294008 1581296003 1581296013 1581549005 1581549015 1581647015 1585827007 1585831003 1586082003 1586090007 1586134005 1586286014 1586982005 1587881006 1587884002 1588001005 1588006015 1588493009 1588526012 1588687009 1588774010 1588778013 1590360012 1590565001 1590587013 1590751011 1590769010 1591097009 1591279005 1591279013 1591670013 1592469R004 1593048R020 1593190r015 1593387R012 1593407R017 1593517R017 1594191R017 1594391011 1594558004 1594769001 1594920006 1595287001 1595477002 1595571001 1595675007 1595767002 1596271006 1596421008 1596540011 1596669010 1599015003 1599354003 1601730005 1601733019 1602400012 1602401014 1604607022 1604740013 1607095005 Lot Codes 1354736001 1354736008 1354737006 1354738003 1354739003 1354739009 1354739022 1354740001 1354740012 1354745006 1354798003 1354798009 1354798015 1354799005 1354800002 1354800009 1354801001 1354801007 1354801014 1354802005 1354803001 1354804009 1354806002 1354806008 1354806020 1354806027 1354807004 1354807010 1354808004 1354809001 1354809010 1354810002 1354810009 1354810016 1354810023 1354859003 1354859010 1354859018 1354859026 1354860006 1354861002 1354861009 1354861015 1354861022 1354861029 1354862005 1354862012 1354862019 1354862029 1354863008 1354865010 1354865018 1354865026 1354867005 1354867013 1354867021 1354867029 1354921007 1354922003 1354922011 1354922019 1354922025 1354923003 1354923010 1354923016 1354923022 1354924004 1354924013 1354924019 1354924028 1354925007 1354925013 1354925019 1354925025 1354926006 1354926014 1354926021 1354926027 1354927012 1354927020 1354927030 1354928009 1354929006 1354929012 1354980006 1354981002 1354981010 1354982004 1354982011 1354985002 1355041006 1355041013 1355041021 1355042002 1355042009 1355042016 1355042022 1355043003 1355043011 1355044005 1355044015 1355044024 1355047009 1355102003 1355102009 1355103001 1355103010 1355103016 1355103022 1355103030 1355104007 1355104018 1355104025 1355105008 1355106002 1355106008 1355106014 1355162006 1355163006 1355163012 1355164008 1355165007 1355165020 1355224005 1355224013 1355224019 1355224028 1355225008 1355225021 1355285003 1355285009 1355344003 1355345001 1355345008 1355345014 1355346006 1355346012 1355347005 1355404002 1355404011 1355404018 1355404025 1355405007 1355405021 1355405030 1355407007 1355407016 1355407024 1355467004 1355467011 1355468003 1355469001 1355469007 1355469015 1355469022 1355470004 1355470010 1355470019 1355470025 1355471003 1355471011 1355472004 1355472014 1355472023 1355473002 1355473010 1355473016 1355473027 1355474010 1355475004 1355475013 1355526006 1355527006 1355528004 1355528014 1355528022 1355588003 1355589002 1355589010 1355589019 1355590003 1355590010 1355591002 1355591009 1355592005 1355592013 1355649004 1355649010 1355710004 1355710010 1355710019 1355710026 1355711005 1355771004 1355772010 1355830005 1355830011 1355830017 1355831002 1355831012 1355831021 1355832001 1355832009 1355890001 1355891010 1355929005 1355930002 1355930009 1355930015 1355931008 1355931014 1355931021 1355931028 1355932006 1355932015 1355932024 1355933003 1355933014 1355934006 1355934013 1355934026 1355989006 1355989014 1355989025 1355990006 1355990015 1356048010 1356048017 1356048024 1356049001 1356049008 1356049014 1356049021 1356050002 1356050008 1356050015 1356053007 1356053015 1356054009 1356054016 1356055007 1356056003 1356056011 1356109004 1356109011 1356110003 1356110009 1356169008 1356220003 1356220010 1356220018 1356220026 1356221005 1356221011 1356221018 1356221025 1356222002 1356222010 1356222019 1356222028 1356281006 1356281013 1356282005 1356282014 1356282021 1356282027 1356283004 1356283012 1356498002 1356498008 1356498014 1356499007 1356499014 1356501013 1356560008 1356560018 1356560026 1356561006 1356562004 1356562010 1356562020 1356562026 1356563004 1356563010 1356564003 1356564011 1356564019 1356564026 1356565004 1356565011 1356565018 1356565024 1356565030 1356566008 1356566017 1356566025 1356619001 1356619011 1356619021 1356619031 1356620001 1356621002 1356621008 1356622012 1356623011 1356623020 1356681007 1356681013 1356681019 1356681026 1356682005 1356682011 1356683002 1356683008 1356683014 1356683020 1356683027 1356684005 1356684012 1356685005 1356685012 1356685020 1356742002 1356803001 1356803008 1356803015 1356804006 1356805003 1356805010 1356805016 1356805024 1356806001 1356806007 1356806017 1356806026 1356807004 1356807012 1356807018 1356807026 1356808007 1356808014 1356808020 1356808027 1356863006 1356863016 1356863027 1356864005 1356864012 1356926003 1356926009 1356926016 1356926022 1356927006 1356927014 1356928001 1356928008 1356928017 1356928024 1356928030 1356929013 1356929020 1356987001 1356987008 1356988002 1356988010 1356989002 1356989011 1463073004 1463074001 1463074009 1463075006 1463076004 1463077001 1463077008 1469330007 1469330016 1472915001 1472915009 1472915015 1472916006 1472916014 1472917005 1472917012 1472917019 1472917026 1472921008 1472921019 1472921029 1472922010 1472924006 1472927004 1472928003 1472928010 1473420001 1473420009 1473420015 1473420021 1473690002 1473690008 1473690015 1473690021 1473690027 1473692006 1473692012 1473692018 1473692025 1473996005 1473996012 1474001003 1474001011 1474667002 1474667013 1474668006 1474668014 1476187007 1476188001 1476188010 1477297002 1481015001 1481015010 1484273001 1484273009 1484277001 1484277009 1484284003 1484284010 1484286002 1484286011 1484291006 1484291015 1484293006 1484293017 1484293025 1484295008 1484295019 1484295028 1484320007 1484320021 1484322002 1484322015 1484323010 1484323017 1484323024 1484324004 1484324012 1484325009 1484338003 1484339003 1484339012 1484340005 1484340014 1484344004 1484345003 1484345010 1488074003 1488074022 1488577001 1488577012 1488577019 1488577027 1488599003 1488599014 1488599023 1491487005 1491487013 1491488007 1491488017 1491488023 1491488030 1491489006 1491489012 1491499003 1491499010 1491499022 1491499028 1491501006 1491501014 1494974003 1494983004 1494985004 1494985011 1496165002 1497834001 1497834013 1497834020 1497834028 1498104010 1498104016 1498872004 1498872011 1499080002 1499080011 1499615005 1499615012 1499615027 1499830006 1499830015 1499830022 1501763005 1514717003 1514717011 1515154001 1515154011 1515186005 1515186014 1515334008 1515334017 1515334023 1515687007 1515687014 1516141009 1516141016 1516141023 1516141031 1516630001 1516630009 1516630021 1516641003 1516641014 1517063015 1517063027 1517160002 1517160012 1517329011 1517435002 1517435008 1517845006 1517849004 1517849011 1518008009 1518580005 1518580013 1519185004 1519220003 1519220012 1519260003 1519368003 1519368012 1519369010 1519370004 1519685004 1519685014 1519685023 1519893001 1519908006 1519918001 1519918010 1519942009 1519966008 1519966018 1521503001 1521503012 1521994003 1521994013 1522138011 1522145003 1522145014 1522160006 1522160012 1522241011 1522447007 1522567003 1522567013 1524365001 1524365008 1524365015 1524556003 1524556012 1526681003 1526681011 1526844007 1526848005 1528890007 1531324004 1543262001 1545444006 1547526006 1547526014 1547629014 1551806005 1551806015 1553666005 1553712005 1553712013 1553758010 1553760006 1556365013 1560273008 1560274001 1562392002 1562392009 1562393012 1562573003 1562573012 1562589007 1562591003 1564482004 1566566004 1568632007 1568633004 1575122002 1575125006 1575315011 1575320005 1577574007 1577575005 1577575014 1578069008 1578332003 1578333002 1578340013 1578544002 1578544010 1578810006 1578810013 1579488012 1579615001 1579615013 1579618008 1579619008 1579770003 1579781003 1579967002 1579972015 1580162012 1580289015 1580540011 1580601013 1580675003 1581019004 1581294009 1581296004 1581296015 1581549006 1581647001 1585766007 1585827011 1585831005 1586082004 1586090014 1586134006 1586297002 1586982009 1587881009 1587884004 1588001007 1588487008 1588493013 1588526013 1588774001 1588774012 1590356003 1590360013 1590565006 1590587014 1590769003 1590769012 1591097015 1591279006 1591279014 1592089R006 1592469R010 1593107R002 1593190R019 1593387R014 1593517R001 1593517R018 1594192R001 1594391015 1594558005 1594769002 1594920007 1595287007 1595477004 1595571006 1595698003 1595767009 1596271012 1596421009 1596540013 1596669012 1599015008 1599354004 1601731004 1601735004 1602400014 1604378002 1604607028 1606029005 1607095008 Lot Codes 1354736003 1354737001 1354737007 1354738004 1354739004 1354739012 1354739023 1354740003 1354745001 1354745008 1354798004 1354798010 1354798016 1354799006 1354800003 1354800010 1354801002 1354801009 1354801015 1354802006 1354803002 1354804010 1354806003 1354806009 1354806021 1354806028 1354807005 1354807011 1354808005 1354809002 1354809011 1354810003 1354810010 1354810017 1354810024 1354859005 1354859011 1354859019 1354860001 1354860007 1354861003 1354861010 1354861016 1354861023 1354861030 1354862006 1354862013 1354862021 1354863002 1354863010 1354865013 1354865019 1354865028 1354867006 1354867014 1354867022 1354867030 1354921008 1354922004 1354922012 1354922020 1354922026 1354923004 1354923011 1354923017 1354923023 1354924005 1354924014 1354924020 1354924029 1354925008 1354925014 1354925020 1354925028 1354926007 1354926016 1354926022 1354927001 1354927013 1354927023 1354928002 1354928010 1354929007 1354929013 1354980008 1354981003 1354981011 1354982006 1354982012 1354985004 1355041007 1355041015 1355041022 1355042003 1355042010 1355042017 1355042023 1355043004 1355043012 1355044007 1355044016 1355044025 1355047011 1355102004 1355102010 1355103003 1355103011 1355103017 1355103023 1355104001 1355104010 1355104019 1355104026 1355105009 1355106003 1355106009 1355106015 1355163001 1355163007 1355163014 1355164009 1355165009 1355165022 1355224006 1355224014 1355224020 1355224029 1355225009 1355225022 1355285004 1355285011 1355344004 1355345002 1355345009 1355345015 1355346007 1355346013 1355347006 1355404004 1355404012 1355404019 1355405001 1355405010 1355405023 1355407001 1355407009 1355407017 1355407025 1355467005 1355467012 1355468005 1355469002 1355469008 1355469016 1355469023 1355470005 1355470012 1355470020 1355470026 1355471005 1355471012 1355472005 1355472015 1355472024 1355473003 1355473011 1355473019 1355474001 1355474011 1355475007 1355475014 1355526009 1355527008 1355528005 1355528015 1355528023 1355588004 1355589004 1355589011 1355589022 1355590004 1355590011 1355591003 1355591010 1355592006 1355592014 1355649005 1355650002 1355710005 1355710011 1355710020 1355710027 1355711006 1355771006 1355772013 1355830006 1355830012 1355830019 1355831004 1355831013 1355831023 1355832003 1355832010 1355890002 1355892001 1355929006 1355930003 1355930010 1355931002 1355931009 1355931015 1355931022 1355931029 1355932008 1355932018 1355932025 1355933004 1355934001 1355934007 1355934014 1355934027 1355989008 1355989018 1355989026 1355990007 1356048003 1356048011 1356048018 1356048025 1356049002 1356049009 1356049016 1356049024 1356050003 1356050009 1356053001 1356053009 1356054002 1356054010 1356054017 1356055008 1356056004 1356056013 1356109005 1356109012 1356110004 1356110010 1356169012 1356220005 1356220011 1356220019 1356220027 1356221006 1356221012 1356221019 1356221026 1356222004 1356222011 1356222020 1356281001 1356281007 1356281014 1356282006 1356282015 1356282022 1356282028 1356283005 1356283015 1356498003 1356498009 1356499002 1356499008 1356501002 1356560001 1356560009 1356560020 1356560027 1356561007 1356562005 1356562011 1356562021 1356562027 1356563005 1356563011 1356564004 1356564013 1356564020 1356564027 1356565005 1356565012 1356565019 1356565025 1356566001 1356566010 1356566018 1356566027 1356619002 1356619013 1356619022 1356619033 1356620002 1356621003 1356621009 1356622014 1356623013 1356681001 1356681008 1356681014 1356681020 1356681027 1356682006 1356682012 1356683003 1356683009 1356683015 1356683021 1356683028 1356684006 1356684013 1356685006 1356685014 1356685021 1356742003 1356803002 1356803009 1356804001 1356804007 1356805005 1356805011 1356805017 1356805025 1356806002 1356806008 1356806018 1356806027 1356807005 1356807013 1356807021 1356808001 1356808008 1356808015 1356808021 1356808028 1356863007 1356863017 1356863029 1356864006 1356864013 1356926004 1356926010 1356926017 1356927001 1356927007 1356927015 1356928002 1356928009 1356928018 1356928025 1356929002 1356929015 1356929021 1356987002 1356987010 1356988003 1356988011 1356989003 1356989013 1463073005 1463074002 1463075001 1463075007 1463076005 1463077002 1463077010 1469330008 1469330017 1472915002 1472915010 1472916001 1472916007 1472916015 1472917006 1472917013 1472917020 1472921001 1472921010 1472921024 1472922004 1472922011 1472924008 1472927005 1472928004 1472928011 1473420002 1473420010 1473420016 1473420022 1473690003 1473690009 1473690016 1473690022 1473690029 1473692007 1473692013 1473692019 1473692026 1473996006 1473996013 1474001005 1474001012 1474667004 1474667014 1474668008 1474668015 1476187008 1476188003 1476188011 1477297003 1481015002 1481015011 1484273004 1484273010 1484277003 1484277011 1484284004 1484285002 1484286003 1484286012 1484291007 1484293001 1484293008 1484293018 1484293026 1484295010 1484295020 1484295029 1484320008 1484320022 1484322005 1484323001 1484323011 1484323018 1484323025 1484324005 1484324013 1484325010 1484338007 1484339004 1484339013 1484340006 1484343001 1484344008 1484345004 1484345014 1488074006 1488074024 1488577003 1488577014 1488577020 1488577028 1488599004 1488599017 1488599025 1491487006 1491487014 1491488008 1491488018 1491488024 1491489001 1491489007 1491489013 1491499004 1491499013 1491499023 1491499029 1491501007 1491501018 1494974004 1494983005 1494985005 1494985012 1496165003 1497834004 1497834014 1497834021 1498104001 1498104011 1498104018 1498872005 1498872012 1499080003 1499170001 1499615006 1499615013 1499615028 1499830007 1499830016 1499830023 1511181002 1514717004 1514898002 1515154004 1515154012 1515186006 1515186015 1515334009 1515334018 1515334024 1515687008 1516141002 1516141010 1516141018 1516141024 1516141032 1516630002 1516630012 1516630023 1516641007 1517063005 1517063019 1517063029 1517160004 1517160013 1517329014 1517435003 1517435010 1517845010 1517849005 1517849012 1518008012 1518580006 1518580014 1519185006 1519220005 1519220013 1519260005 1519368004 1519369001 1519369011 1519370006 1519685005 1519685015 1519685024 1519893002 1519908007 1519918002 1519918015 1519942010 1519966009 1519966020 1521503002 1521992001 1521994005 1521994015 1522138012 1522145004 1522145015 1522160007 1522160013 1522241012 1522447008 1522567004 1522567014 1524365002 1524365009 1524366003 1524556005 1524556013 1526681004 1526681012 1526844011 1526848008 1528890011 1538213004 1543262002 1545444007 1547526008 1547629003 1547629015 1551806006 1553665001 1553666007 1553712006 1553712015 1553758012 1556365004 1556365014 1560273010 1560274002 1562392003 1562392012 1562426001 1562573005 1562589002 1562589009 1562591004 1564482007 1566566007 1568632009 1568633005 1575122004 1575125007 1575315014 1575320006 1577574008 1577575008 1577575016 1578331002 1578332004 1578333003 1578543006 1578544003 1578544014 1578810007 1579488002 1579488013 1579615004 1579615014 1579618009 1579619009 1579770007 1579781007 1579967007 1579973002 1580188005 1580291002 1580540012 1580671006 1580675004 1581019008 1581294010 1581296006 1581505001 1581549007 1581647002 1585766008 1585827012 1585831006 1586082005 1586132001 1586134009 1586297009 1586982013 1587881011 1587884006 1588006002 1588487009 1588526001 1588687003 1588774003 1588774014 1590356005 1590360016 1590565010 1590751001 1590769005 1590769013 1591128009 1591279009 1591670002 1592089R011 1592469R017 1593107R005 1593387R002 1593387R015 1593517R002 1593674R014 1594192R002 1594392005 1594558008 1594769007 1594920008 1595287009 1595477005 1595672002 1595698004 1595767014 1596320002 1596540001 1596669001 1596669013 1599015011 1599397012 1601731009 1601735015 1602401001 1604378005 1604607029 1606169001 1607095009 Lot Codes 1354736004 1354737002 1354737008 1354738005 1354739005 1354739013 1354739024 1354740005 1354745002 1354745010 1354798005 1354798011 1354799001 1354799007 1354800004 1354800011 1354801003 1354801010 1354802001 1354802007 1354803003 1354804012 1354806004 1354806010 1354806023 1354806029 1354807006 1354807013 1354808006 1354809003 1354809012 1354810004 1354810011 1354810018 1354810025 1354859006 1354859012 1354859022 1354860002 1354860008 1354861004 1354861011 1354861017 1354861024 1354862001 1354862007 1354862014 1354862022 1354863003 1354863011 1354865014 1354865020 1354865029 1354867008 1354867015 1354867024 1354921001 1354921009 1354922007 1354922013 1354922021 1354922027 1354923005 1354923012 1354923018 1354923024 1354924006 1354924015 1354924021 1354925001 1354925009 1354925015 1354925021 1354926002 1354926008 1354926017 1354926023 1354927003 1354927014 1354927024 1354928003 1354929001 1354929008 1354980001 1354980011 1354981004 1354981013 1354982007 1354982014 1354985005 1355041008 1355041016 1355041023 1355042005 1355042012 1355042018 1355042024 1355043005 1355043013 1355044009 1355044017 1355047001 1355047012 1355102005 1355102011 1355103004 1355103012 1355103018 1355103024 1355104002 1355104013 1355104020 1355105001 1355105010 1355106004 1355106010 1355162001 1355163002 1355163008 1355163015 1355164014 1355165014 1355165023 1355224007 1355224015 1355224021 1355224030 1355225012 1355225024 1355285005 1355285012 1355344007 1355345003 1355345010 1355346002 1355346008 1355346014 1355347007 1355404005 1355404013 1355404021 1355405002 1355405011 1355405025 1355407002 1355407012 1355407018 1355407026 1355467006 1355467013 1355468006 1355469003 1355469010 1355469017 1355469024 1355470006 1355470013 1355470021 1355470027 1355471006 1355471013 1355472007 1355472016 1355472025 1355473004 1355473012 1355473020 1355474003 1355474012 1355475009 1355526001 1355526010 1355527011 1355528007 1355528016 1355528025 1355588005 1355589005 1355589013 1355589025 1355590006 1355590012 1355591005 1355591011 1355592007 1355592015 1355649006 1355650004 1355710006 1355710013 1355710021 1355710028 1355711007 1355771007 1355772014 1355830007 1355830013 1355830021 1355831007 1355831014 1355831024 1355832004 1355832011 1355890003 1355892007 1355929007 1355930004 1355930011 1355931003 1355931010 1355931016 1355931023 1355932002 1355932011 1355932019 1355932027 1355933005 1355934002 1355934008 1355934015 1355934029 1355989009 1355989019 1355989028 1355990008 1356048004 1356048012 1356048019 1356048026 1356049003 1356049010 1356049017 1356049025 1356050004 1356050010 1356053002 1356053010 1356054004 1356054011 1356055002 1356055009 1356056005 1356056014 1356109006 1356109013 1356110005 1356110011 1356192001 1356220006 1356220012 1356220021 1356220028 1356221007 1356221013 1356221020 1356221028 1356222005 1356222015 1356222022 1356281002 1356281008 1356282001 1356282007 1356282016 1356282023 1356282029 1356283008 1356283018 1356498004 1356498010 1356499003 1356499009 1356501003 1356560002 1356560010 1356560021 1356561001 1356561008 1356562006 1356562014 1356562022 1356562028 1356563006 1356563012 1356564005 1356564014 1356564021 1356564028 1356565006 1356565013 1356565020 1356565026 1356566002 1356566011 1356566019 1356566028 1356619003 1356619015 1356619023 1356619034 1356620003 1356621004 1356621010 1356623003 1356623014 1356681002 1356681009 1356681015 1356681021 1356682001 1356682007 1356682013 1356683004 1356683010 1356683016 1356683022 1356683029 1356684007 1356684014 1356685008 1356685015 1356685022 1356742004 1356803004 1356803010 1356804002 1356804008 1356805006 1356805012 1356805018 1356805026 1356806003 1356806009 1356806019 1356806030 1356807008 1356807014 1356807022 1356808002 1356808009 1356808016 1356808023 1356863001 1356863009 1356863018 1356863030 1356864008 1356864014 1356926005 1356926011 1356926018 1356927002 1356927008 1356927016 1356928003 1356928010 1356928019 1356928026 1356929004 1356929016 1356929023 1356987003 1356987011 1356988004 1356988012 1356989004 1356989014 1463073006 1463074003 1463075002 1463075008 1463076007 1463077004 1469330002 1469330010 1469330018 1472915003 1472915011 1472916002 1472916008 1472917001 1472917007 1472917014 1472917021 1472921003 1472921011 1472921025 1472922005 1472924001 1472924011 1472927007 1472928005 1472928012 1473420003 1473420011 1473420017 1473420023 1473690004 1473690010 1473690017 1473690023 1473692001 1473692008 1473692014 1473692020 1473692027 1473996007 1473996014 1474001006 1474001013 1474667005 1474667015 1474668009 1476187001 1476187011 1476188004 1476189002 1477297004 1481015004 1481015012 1484273005 1484273011 1484277004 1484277012 1484284005 1484285004 1484286004 1484286015 1484291009 1484293002 1484293009 1484293019 1484293028 1484295011 1484295021 1484320001 1484320009 1484320023 1484322006 1484323002 1484323012 1484323019 1484323026 1484324006 1484324014 1484325011 1484338010 1484339005 1484339014 1484340009 1484343004 1484344014 1484345005 1484656007 1488074007 1488074025 1488577006 1488577015 1488577021 1488577029 1488599006 1488599018 1488599027 1491487007 1491487015 1491488009 1491488019 1491488025 1491489002 1491489008 1491489014 1491499005 1491499015 1491499024 1491501002 1491501008 1491501020 1494974006 1494983006 1494985006 1494985013 1496165005 1497834005 1497834015 1497834023 1498104005 1498104012 1498104019 1498872007 1498872013 1499080004 1499170002 1499615007 1499615015 1499830002 1499830010 1499830017 1499830025 1511181003 1514717005 1514898003 1515154005 1515154013 1515186009 1515334001 1515334010 1515334019 1515334025 1515687009 1516141004 1516141011 1516141019 1516141027 1516141034 1516630003 1516630015 1516630025 1516641009 1517063008 1517063022 1517063030 1517160005 1517160014 1517412001 1517435004 1517435011 1517845011 1517849006 1518008005 1518580001 1518580007 1518970004 1519185007 1519220006 1519220014 1519260008 1519368005 1519369002 1519369013 1519370008 1519685008 1519685016 1519685026 1519893009 1519908009 1519918003 1519942001 1519942013 1519966010 1520115001 1521503003 1521992006 1521994006 1522138003 1522138013 1522145005 1522160001 1522160008 1522241002 1522241013 1522447009 1522567008 1524364001 1524365003 1524365010 1524366005 1524556007 1524556014 1526681005 1526681013 1526844012 1526848009 1528890014 1538213014 1545444001 1545444008 1547526009 1547629004 1551806001 1551806007 1553665003 1553666012 1553712007 1553758004 1553758014 1556365006 1560273004 1560273011 1560274003 1562392004 1562392013 1562426012 1562573006 1562589003 1562589013 1562591006 1564482008 1566566008 1568632013 1568633006 1575122005 1575125011 1575315015 1575320010 1577574010 1577575009 1578069002 1578331003 1578332007 1578333005 1578543007 1578544005 1578544015 1578810008 1579488003 1579613003 1579615006 1579615015 1579618010 1579619012 1579770012 1579781008 1579967014 1579973003 1580188009 1580291003 1580583001 1580671009 1580675010 1581294003 1581294011 1581296010 1581549001 1581549008 1581647004 1585766010 158
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1039-2017
Related Recalls
A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.
Endologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions to or through an existing AFX device. This is a follow-up notification to the one that was sent out in December 2016. The updates are related to the observation of Type III endoleaks.
Endologix is notifying physicians of patient tailored surveillance recommendations as well as general warnings/precautions for interventions to or through an existing AFX device. This is a follow-up notification to the one that was sent out in December 2016. The updates are related to the observation of Type III endoleaks.