Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN: 00859464007212 Lot Number: All Lot Numbers Serial Number: All Serial Numbers
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Turbett Surgical, Inc.
Reason for Recall:
Pod may have weld failures breaching the sterile barrier and result in the use of unsterile instruments leading to infection or other major injuries.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Turbett Surgical Container, TS1000: Indicated for enclosing other medical devices that are to be sterilized by a healthcare provider

Product Codes/Lot Numbers:

GTIN: 00859464007212 Lot Number: All Lot Numbers Serial Number: All Serial Numbers

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1038-2023