Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
Class I - DangerousWhat Should You Do?
- Check if you have this product: affects all units.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Olympus Scientific Solutions Americas
- Reason for Recall:
- It was discovered under rare circumstances the LED warning light circuit could fail. If this happens, while the instrument is emitting X-rays during at test, the LED warning lights will not illuminate. The test runs normally and will execute as before, but the light will not be on.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Olympus Scientific Solutions Americas Corporation (OSSA) VANTA¿ XRF Analyzer. Model Vanta VCR and Vanta VMR Analytical X-ray systems
Product Codes/Lot Numbers:
affects all units.
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-1036-2017
Related Recalls
Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System, X-Ray Fluorescence
Olympus Scientific Solutions Americas
Instruments may, under certain specific circumstances listed below, fail to stop when a test is initiated and no sample is present (proximity shut down fails to end test).
Olympus Scientific Solutions Americas Corporation (OSSA) Model: VANTA Analytical X-Ray System
Olympus Scientific Solutions Americas
Possible unintentional x-ray emission after users attempt early termination of the sequence.
Trigger lock option does not always prevent x-ray generation. Specifically if the unit is in test mode or repeat test mode then the trigger lock may not prevent x-ray initiation when the trigger is pulled. The correction is via repair by a software update.