Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model Number: 08N53-002
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Abbott Molecular, Inc.
Reason for Recall:
The firm identified the following three potential issues which may have impacted ADUs installed on distributed Alinity m Systems. 1. Systems which had their ADUs serviced may have an incorrect board value (calibration data file) referenced causing a misconfiguration of the calibration. 2. Inability of Maintenance and Diagnostics 2300 (Amp Detect Optical Calibrations) to recognize error handling and thermal hold errors during the calibration process, potentially leading to a misconfiguration of the calibration. 3. Under specific conditions during the ADU manufacturing process, an incorrect calibration may potentially have been configured.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Alinity m System. The Alinity m System is designed to automate the steps for nucleic acid testing, which includes sample processing, amplification, detection, and data reduction.

Product Codes/Lot Numbers:

Model Number: 08N53-002

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1009-2023

Related Recalls

There were reports of an increase in reactive negative controls and false positive results.

Apr 28, 2025 Implants & Prosthetics Nationwide View Details β†’

There were reports of an increase in reactive negative controls and false positive results.

Apr 28, 2025 Diagnostic Equipment Nationwide View Details β†’

The design of the Alinity m System Solutions drawer enclosure may allow liquid (liquid waste or system solutions) originating in the System Solutions drawer to flow beyond the footprint of the instrument and into the walking-path of the user.

Mar 20, 2025 Diagnostic Equipment Nationwide View Details β†’