ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Globus Medical, Inc.
Reason for Recall:
Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

Product Codes/Lot Numbers:

UDI-DI 00889095998498 Lots FSB136AA FSB121AA FSB241AA FSB207AA FSB136AA FSB121AA FSB110AA FSB073AA FSA286CA FSA286BA FSA286AA FSA263DA FSA263EA FSA263CA FSA263AA FSA263BA FSA160AA FSA160BA FSA105BA FSA105AA FSA053AA FSY319AA FSY279BA FSY279AA FSY272AA FSY230AA FSY098AA FSY098CA FSY098BA FSY048AA FSX093AA FSX205AA FSX141BA FSX141CA FSX141AA FSX093AA FSX063AA FSX041AA FSX041BA FSW142AA FSW094AA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-1007-2024

Related Recalls