R&D CBC-3D Hematology Control: (1) CBC-3D Normal Pack Hematology Control, REF 3D501, containing 10 x 2.0mL Control N; (2) CBC-3D TRI PACK Hematology Control, REF 3D503, containing 4 x 2.0 mL Controls L, N, and H; (3) CBC-3D for Mythic 18 Hematology Control, REF 3D503RX, containing 2 x 2.0 mL each Controls L, N, and H; (4) CBC-3D For MINDRAY Hematology Control, REF 3D506, containing 2 x 3.0 mL each Controls L, N, and H; (5) CBC-3D For MINDRAY Hematology Control, REF 3D506US, containing 2 x 3.0 mL each Controls L, N, and H; (6) CBC-3D TRI PACK Hematology Control, REF 3D507, containing 4 x 2.0 mL each Controls L, N, and H; (7) CBC-3D TRI PACK Hematology Control, REF 3D508, containing 2 x 2.0 m each Controls L, N, and H; (8) CBC-3D For ABACUS, REF D3D02 Hematology Control, containing 2 x 3.0 mL each Controls L, N, and H; (9) CBC-3D For ABACUS Hematology Control, REF D3D04, containing 4 x 3.0 mL each Controls L, N, and H. Loose tubes labeled CBC-3D Hematology Control in 2.0mL and 3.0mL sizes. OEM/Own-label distributor labels: (1) Alere 380 Hemo Control (L-N-H), REF AL35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H; Kit lot #B1122; control lot #B1122N, no UDI (2) diatron Diacon 3 Hematology Control, Normal Level, 3.0 mL tubes, lot #B1122N (3) diatron Diacon 3 Hematology Control, REF DN35002-SET, containing 6 x 3.0 mL Control N, kit lot #B1122, control lot #B1122N (4) diatron Diacon 3 Hematology Control, REF DN35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, Normal control lot #B1122N. (5) Elite 3 HEM Control Normal, REF HEM00014, containing 6 x 3.0 mL Normal controls, kit lot #B1122, control lot #B1122N; (6) Erba H3 CON N, REF HEM00034, 3 mL vials, lot #E1122N, exp. 2023-02-10 (7) Horiba Medical Minotrol 16 Whole Blood Hematology Control, Cat. #5300000277, containing 3 x 2.5mL tubes (1 each Control 1, Control 2 (Normal), and Control 3; Cat. #5300000278, containing 6 x 2.5mL tubes (2 each Control 1, Control 2 (Normal), and Control 3; and Cat. #5300000283, containing 12 x 2.5mL tubes (4 each Control 1, Control 2 (Normal), and Control 3; kit lot #MX438, exp. 2023-01-05, Control 2 lot #MX438N. (8) 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-912-3D; and 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-914-3D. (9) mindray BC-3D, P/N 040-001423-00. (10) MINDRAY BC-3D CONTROL, REF BC3D01, containing 1 x 3.0 mL Controls L, N, and H; REF BC3D02, containing 2 x 3.0 mL Controls L, N, and H; and REF BC3D6N, containing 6 x 3.0 mL Control N; kit lot #B1122, control lot #B1122N. (11) MINDRAY BC-3D CONTROL, REF BC3D02US, containing 2 x 3.0 mL Level L, N, and H, kit lot number B1122, control lot #B1122N. (12) Nihon Kohden MEK-3D Hematology Control, REF 3DN03, containing MEK-3DN x 3 tubes; REF 3DN06, containing MEK-3DN x 6 tubes; and REF 3DN12, containing MEK-3DN x 12 tubes, kit lot #B22Y, control lot #B22YN. (13) Nihon Kohden MEK-3DN Normal Hematology Control tubes, 2.0 mL, lot #B22YN. (14) Orphee MYT-3D, REF MYT301, containing 2 x 2.5 mL Control N; REF MYT302, containing 2 x 2.5 mL each Controls L, N, and H; REF MYT304N, containing 4 x 2.5 mL Control N; REF 306N, containing 6 x 2.5 mL Control N; kit lot #B1122, control lot #B1122N; Myt-3D VET, REF MYT303-VET, containing 1 x 2.5 mL each Control L, N, and H. (15) Siemens ADVIA 360 Control, REF 11170853, containing 2 x 3.0 mL each Controls Low, Normal & High, kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) URIT HQ-3DIFF, REF 5411000306, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, control lot #B1122N; URIT HQ-3DIFF Hematology Control N tubes, 3.0 mL, lot #B1122N. (17) Woodley Haematology Control, REF WD1154, containing 6 x 2.0 mL Control N; and REF WD1154A, containing 2 x 2.0 mL each Controls L, N, and H, kit lot number B1122, control lot #B1122N;
Class I - DangerousWhat Should You Do?
- Check if you have this product: R&D labeled product lot numbers: (Products 1-9): Lot B1122, exp. date 2023-02-05; UDI numbers for the following catalog numbers: 3D501 - UDI 10815762020091; 3D503 - UDI 00815762020117; 3D503RX - UDI 00815762020124; 3D506 - UDI N/A OUS only; 3D506US - UDI 00815762020162; 3D507 - UDI 00815762020179; 3D508 - UDI 00815762020186; D3D02 - UDI 00815762020230; and D3D04 - UDI 00815762020247. OEM/Own-label distributor labeled product lot numbers (Products 1-17) (1) Kit lot #B1122; control lot #B1122N. (2) Lot #B1122N. (3) Kit lot #B1122, control lot #B1122N. (4) Kit lot #B1122, control lot #B1122N. (5) Kit lot #B1122, control lot #B1122N. (6) Lot #E1122N. (7) Kit lot #MX438, Control 2 lot #MX438N. (8) Lot #2022-912-3D and Lot #2022-914-3D. (9) P/N 040-001423-00. (10) Kit lot #B1122, control lot #B1122N. (11) Kit lot number B1122, control lot #B1122N. (12) Kit lot #B22Y, control lot #B22YN. (13) Lot #B22YN. (14) Kit lot #B1122, control lot #B1122N. (15) Kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) Kit lot #B1122, control lot #B1122N. (17) Kit lot number B1122, control lot #B1122N;
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
โ ๏ธ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- R & D Systems, Inc.
- Reason for Recall:
- For one lot of CBC-3D Hematology Control, the Normal Level may exhibit hemolysis or deterioration due to microbial contamination.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
R&D CBC-3D Hematology Control: (1) CBC-3D Normal Pack Hematology Control, REF 3D501, containing 10 x 2.0mL Control N; (2) CBC-3D TRI PACK Hematology Control, REF 3D503, containing 4 x 2.0 mL Controls L, N, and H; (3) CBC-3D for Mythic 18 Hematology Control, REF 3D503RX, containing 2 x 2.0 mL each Controls L, N, and H; (4) CBC-3D For MINDRAY Hematology Control, REF 3D506, containing 2 x 3.0 mL each Controls L, N, and H; (5) CBC-3D For MINDRAY Hematology Control, REF 3D506US, containing 2 x 3.0 mL each Controls L, N, and H; (6) CBC-3D TRI PACK Hematology Control, REF 3D507, containing 4 x 2.0 mL each Controls L, N, and H; (7) CBC-3D TRI PACK Hematology Control, REF 3D508, containing 2 x 2.0 m each Controls L, N, and H; (8) CBC-3D For ABACUS, REF D3D02 Hematology Control, containing 2 x 3.0 mL each Controls L, N, and H; (9) CBC-3D For ABACUS Hematology Control, REF D3D04, containing 4 x 3.0 mL each Controls L, N, and H. Loose tubes labeled CBC-3D Hematology Control in 2.0mL and 3.0mL sizes. OEM/Own-label distributor labels: (1) Alere 380 Hemo Control (L-N-H), REF AL35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H; Kit lot #B1122; control lot #B1122N, no UDI (2) diatron Diacon 3 Hematology Control, Normal Level, 3.0 mL tubes, lot #B1122N (3) diatron Diacon 3 Hematology Control, REF DN35002-SET, containing 6 x 3.0 mL Control N, kit lot #B1122, control lot #B1122N (4) diatron Diacon 3 Hematology Control, REF DN35004-KIT, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, Normal control lot #B1122N. (5) Elite 3 HEM Control Normal, REF HEM00014, containing 6 x 3.0 mL Normal controls, kit lot #B1122, control lot #B1122N; (6) Erba H3 CON N, REF HEM00034, 3 mL vials, lot #E1122N, exp. 2023-02-10 (7) Horiba Medical Minotrol 16 Whole Blood Hematology Control, Cat. #5300000277, containing 3 x 2.5mL tubes (1 each Control 1, Control 2 (Normal), and Control 3; Cat. #5300000278, containing 6 x 2.5mL tubes (2 each Control 1, Control 2 (Normal), and Control 3; and Cat. #5300000283, containing 12 x 2.5mL tubes (4 each Control 1, Control 2 (Normal), and Control 3; kit lot #MX438, exp. 2023-01-05, Control 2 lot #MX438N. (8) 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-912-3D; and 2022 3rd Test Event, CBC PLT 3 PART DIFF 2022-914-3D. (9) mindray BC-3D, P/N 040-001423-00. (10) MINDRAY BC-3D CONTROL, REF BC3D01, containing 1 x 3.0 mL Controls L, N, and H; REF BC3D02, containing 2 x 3.0 mL Controls L, N, and H; and REF BC3D6N, containing 6 x 3.0 mL Control N; kit lot #B1122, control lot #B1122N. (11) MINDRAY BC-3D CONTROL, REF BC3D02US, containing 2 x 3.0 mL Level L, N, and H, kit lot number B1122, control lot #B1122N. (12) Nihon Kohden MEK-3D Hematology Control, REF 3DN03, containing MEK-3DN x 3 tubes; REF 3DN06, containing MEK-3DN x 6 tubes; and REF 3DN12, containing MEK-3DN x 12 tubes, kit lot #B22Y, control lot #B22YN. (13) Nihon Kohden MEK-3DN Normal Hematology Control tubes, 2.0 mL, lot #B22YN. (14) Orphee MYT-3D, REF MYT301, containing 2 x 2.5 mL Control N; REF MYT302, containing 2 x 2.5 mL each Controls L, N, and H; REF MYT304N, containing 4 x 2.5 mL Control N; REF 306N, containing 6 x 2.5 mL Control N; kit lot #B1122, control lot #B1122N; Myt-3D VET, REF MYT303-VET, containing 1 x 2.5 mL each Control L, N, and H. (15) Siemens ADVIA 360 Control, REF 11170853, containing 2 x 3.0 mL each Controls Low, Normal & High, kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) URIT HQ-3DIFF, REF 5411000306, containing 2 x 3.0 mL each Controls L, N, and H, kit lot #B1122, control lot #B1122N; URIT HQ-3DIFF Hematology Control N tubes, 3.0 mL, lot #B1122N. (17) Woodley Haematology Control, REF WD1154, containing 6 x 2.0 mL Control N; and REF WD1154A, containing 2 x 2.0 mL each Controls L, N, and H, kit lot number B1122, control lot #B1122N;
Product Codes/Lot Numbers:
R&D labeled product lot numbers: (Products 1-9): Lot B1122, exp. date 2023-02-05; UDI numbers for the following catalog numbers: 3D501 - UDI 10815762020091; 3D503 - UDI 00815762020117; 3D503RX - UDI 00815762020124; 3D506 - UDI N/A OUS only; 3D506US - UDI 00815762020162; 3D507 - UDI 00815762020179; 3D508 - UDI 00815762020186; D3D02 - UDI 00815762020230; and D3D04 - UDI 00815762020247. OEM/Own-label distributor labeled product lot numbers (Products 1-17) (1) Kit lot #B1122; control lot #B1122N. (2) Lot #B1122N. (3) Kit lot #B1122, control lot #B1122N. (4) Kit lot #B1122, control lot #B1122N. (5) Kit lot #B1122, control lot #B1122N. (6) Lot #E1122N. (7) Kit lot #MX438, Control 2 lot #MX438N. (8) Lot #2022-912-3D and Lot #2022-914-3D. (9) P/N 040-001423-00. (10) Kit lot #B1122, control lot #B1122N. (11) Kit lot number B1122, control lot #B1122N. (12) Kit lot #B22Y, control lot #B22YN. (13) Lot #B22YN. (14) Kit lot #B1122, control lot #B1122N. (15) Kit lot #B0822, GTIN 00630414600833, control lot #B1122N. (16) Kit lot #B1122, control lot #B1122N. (17) Kit lot number B1122, control lot #B1122N;
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0999-2023
Related Recalls
The Instructions for Use (IFU, Part Number 750095) provided with the affected lots by the firm contains an error associated with product s CE marking. The CE mark contained in the IFU is followed by the Notified Body designation, British Standards Institution number 2797 (BSI-2797) for product inserts 750095.28 and 750095.29. The erroneous label CE2797 were in the IFU insert revision 750095.28 for the 408 affected lots that were shipped. No product was packed with erroneous insert revision 750095.29. Since the product is currently CE marked as a Class I - self-certified product under the EU In Vitro Diagnostics Directive, the Notified Body designation should not be specified. As part of the corrective action, the firm has notified the customers who received the products with erroneous labeling, and is conducting the replacements with the correct product insert containing the Insert Revision 750095.30
R&D Systems, Inc. received two complaints that the standard signal was low and controls were not within the specification range provided with the kit. Investigation confirmed low standard signal and controls running out of specification for kits returned from the customer and the retained kits.
The Human sTfR Quantikine IVD ELISA Kit, catalog # DTFR1, lot P180151 was packed with the incorrect control data card number 750228.4.