Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063
Class I - DangerousWhat Should You Do?
- Check if you have this product: All product manufactured after December 10, 2017; Serial Numbers greater than or equal to N17606406
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Cardinal Health 200, LLC
- Reason for Recall:
- The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range (¿ 0.2¿C for Genius 2 and ¿ 0.3¿C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2¿C for Genius 2 or +0.3¿C for Genius 3.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Genius 2 Tympanic Thermometer OEM Tympanic Item Code: 303063
Product Codes/Lot Numbers:
All product manufactured after December 10, 2017; Serial Numbers greater than or equal to N17606406
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0993-2020