VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port (refurbished), Cat. #W0441R; (7) dc7700 8-port, Cat. #W0447; (8) dc7700 8-port (refurbished), Cat. #W0447R; (9) dc7800, Cat. #W0449; and (10) dc7800 (refurbished), Cat. #W0449R. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Manufactured from September, 2004, to present.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Biomerieux Inc
- Reason for Recall:
- There is a potential to link an isolate result to the wrong patient and then upload the results to the LIS.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
VITEK 2 XL computer system when equipped with the following Hewlett-Packard PC models: (1) rp5700, Cat. #W0452; (2) rp5700 (refurbished), Cat. #W0452R; (3) rp5800 XPE, Cat. 413642; (4) rp5800 WES7, Cat. #413862; (5) dc7100 8-port, Cat. #W0441; (6) dc7100 8-port (refurbished), Cat. #W0441R; (7) dc7700 8-port, Cat. #W0447; (8) dc7700 8-port (refurbished), Cat. #W0447R; (9) dc7800, Cat. #W0449; and (10) dc7800 (refurbished), Cat. #W0449R. VITEK 2 is an automated system designed for the identification and antimicrobial susceptibility testing of bacteria and yeast.
Product Codes/Lot Numbers:
Manufactured from September, 2004, to present.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0990-2013
Related Recalls
Potential for false resistant results for Colistin (cs02n) with multi-drug-resistant Pseudomonas aeruginosa and Acinetobacter baumannii complex used in specific cards
Ceftriaxone (cro02n) concentration errors associated with specific recently manufactured test kits which would have the potential for for false susceptible results. The risk is present for isolates that have a result of MIC=0.5, 1, 2.
For users with MYLA V4.8.X / V4.9 that use VITEK MS to identify organisms as part of their workflow, AST filter rules that have been activated are not always being applied to AST results when sending the results to the clinician.