Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchored Lateral Retractable Drills are used with the Timberline MPF System to establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Model 8630-0102 Anchored Lateral Retractable Drill lot TU00049: February 25, 2013 and TU00155: July 31, 2013
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Biomet Spine, LLC
Reason for Recall:
The outer diameter of the Drill shaft is oversized; thereby, resulting in interference fit when inserting the Drill into the Fixed or Variable Sleeve Assembly.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Anchored Lateral Retractable Drill, Timberline IPF IFU IFU8621-0113, Model 8630-0102. Timberline Anchored Lateral Retractable Drills are used with the Timberline MPF System to establish access into the vertebrae for the screws utilized to anchor the plate to the vertebrae.

Product Codes/Lot Numbers:

Model 8630-0102 Anchored Lateral Retractable Drill lot TU00049: February 25, 2013 and TU00155: July 31, 2013

Distribution:

Distributed in: PR, NY, AZ, TN, TX, CA

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0982-2015

Related Recalls

Zimmer Biomet Spine initiated a recall of certain Solitaire Ti Spacers because they may have a mislabeled lordotic angle of 12degrees when the angle is actually 6 degrees.

Feb 12, 2016 Infusion Pumps Nationwide View Details →