Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
Class I - DangerousWhat Should You Do?
- Check if you have this product: serial numbers: ABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482, ABSO-004483, ABSO-004487, ABSO-004488, ABSO-004489, ABSO-004490, ABSO-004491, ABSO-004492, ABSO-004493, ABSO-004494, ABSO-004495, ABSO-004497, ABSO-004499, ABSO-004500, ABSO-004503, ABSO-004504, ABSO-004505, ABSO-004506, ABSO-004507, ABSO-004508, ABSO-004510, ABSO-004511, ABSO-004512, ABSO-004513, ABSO-004526, ABSO-004528, ABSO-004529, ABSO-004530, ABSO-004531, ABSO-004532, ABSO-004533, ABSO-004534, ABSO-004535, ABSO-004536, ABSO-004537, ABSO-004538, ABSO-004539, ABSO-004540, ABSO-004541, ABSO-004542, ABSO-004543, ABSO-004544, ABSO-004545, ABSO-004564, ABSO-004566, ABSO-004567, ABSO-004568, ABSO-004569, ABSO-005883, ABSO-005884, ABSO-005885, ABSO-005886, ABSO-005887, ABSO-005945, ABSO-005947, ABSO-005948, ABSO-005949, ABSO-005950, ABSO-100023, ABSO-100024, ABSO-100038, ABSO-100039, ABSO-100040, ABSO-100041, ABSO-100042, ABSO-100043, ABSO-100070, ABSO-100071, ABSO-100072, ABSO-100073, ABSO-100074, ABSO-100075, ABSO-100076, ABSO-100077, ABSO-100078, ABSO-100079, ABSO-100080, ABSO-100081, ABSO-100082, ABSO-100083, ABSO-100084, ABSO-100085, ABSO-100086, and ABSO-100087. Firm added ABSO 004581 - which was inSpacelaabs inventory, in transit to a Field Service Engineer, at the time the recall was initiated. the nonconforming absorber has been replaced with a conforming part and this absober serial number was added to the US Spacelabs Recall tracking. Five incorrect serialnumbers were included in the original serial numbers provided to FDA. This was a result of a data entry error. An updated US and Rest of the World customer/Serial number was provided to FDA 3/25/2013. The correction to the Rest of the World report is as follows: Incorrect SN Reported to FDA Correct, Nonconforming SN ABSO 005883 ABSO-004514 ABSO 005884 ABSO-004515 ABSO 005885 ABSO-004516 ABSO 005886 ABSO-004517 ABSO 005887 ABSO-004518
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Del Mar Reynolds Medical, Ltd.
- Reason for Recall:
- A defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.
Product Codes/Lot Numbers:
serial numbers: ABSO-004448, ABSO-004458, ABSO-004459, ABSO-004460, ABSO-004461, ABSO-004462, ABSO-004463, ABSO-004465, ABSO-004466, ABSO-004467, ABSO-004468, ABSO-004469, ABSO-004470, ABSO-004471, ABSO-004472, ABSO-004473, ABSO-004474, ABSO-004475, ABSO-004476, ABSO-004477, ABSO-004478, ABSO-004479, ABSO-004480, ABSO-004481, ABSO-004482, ABSO-004483, ABSO-004487, ABSO-004488, ABSO-004489, ABSO-004490, ABSO-004491, ABSO-004492, ABSO-004493, ABSO-004494, ABSO-004495, ABSO-004497, ABSO-004499, ABSO-004500, ABSO-004503, ABSO-004504, ABSO-004505, ABSO-004506, ABSO-004507, ABSO-004508, ABSO-004510, ABSO-004511, ABSO-004512, ABSO-004513, ABSO-004526, ABSO-004528, ABSO-004529, ABSO-004530, ABSO-004531, ABSO-004532, ABSO-004533, ABSO-004534, ABSO-004535, ABSO-004536, ABSO-004537, ABSO-004538, ABSO-004539, ABSO-004540, ABSO-004541, ABSO-004542, ABSO-004543, ABSO-004544, ABSO-004545, ABSO-004564, ABSO-004566, ABSO-004567, ABSO-004568, ABSO-004569, ABSO-005883, ABSO-005884, ABSO-005885, ABSO-005886, ABSO-005887, ABSO-005945, ABSO-005947, ABSO-005948, ABSO-005949, ABSO-005950, ABSO-100023, ABSO-100024, ABSO-100038, ABSO-100039, ABSO-100040, ABSO-100041, ABSO-100042, ABSO-100043, ABSO-100070, ABSO-100071, ABSO-100072, ABSO-100073, ABSO-100074, ABSO-100075, ABSO-100076, ABSO-100077, ABSO-100078, ABSO-100079, ABSO-100080, ABSO-100081, ABSO-100082, ABSO-100083, ABSO-100084, ABSO-100085, ABSO-100086, and ABSO-100087. Firm added ABSO 004581 - which was inSpacelaabs inventory, in transit to a Field Service Engineer, at the time the recall was initiated. the nonconforming absorber has been replaced with a conforming part and this absober serial number was added to the US Spacelabs Recall tracking. Five incorrect serialnumbers were included in the original serial numbers provided to FDA. This was a result of a data entry error. An updated US and Rest of the World customer/Serial number was provided to FDA 3/25/2013. The correction to the Rest of the World report is as follows: Incorrect SN Reported to FDA Correct, Nonconforming SN ABSO 005883 ABSO-004514 ABSO 005884 ABSO-004515 ABSO 005885 ABSO-004516 ABSO 005886 ABSO-004517 ABSO 005887 ABSO-004518
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0968-2013
Related Recalls
Reports of the Canister Carrier breaking during absorbent exchange or during anesthesia machine transit, causing a leak in the breathing circuit, which could cause hypoventilation in the patient.
Spacelabs has received reports of the Arkon Anesthesia Delivery System, Model 99999, shutting down without warning. The power supply has failed in a way that does not trigger the normal on battery alarms.
Reports of inaccurate low flow readings. Monitored inspiratory tidal volume (VTi) and expiratory tidal volume (Vte) measurements from the pediatric flow sensor are reporting out of specification low compared to the actual delivered volumes being administered to the patient.