Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad CRT -D SureScan. implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers of AMPLIA & CLARIA Released from Manufacturing as of 22-Nov-2016 UPN Serial 00643169543362 RPA200003H 00643169543362 RPA200004H 00643169543362 RPA200005H 00643169543362 RPA200006H 00643169543362 RPA200008H 00643169543362 RPA200009H 00643169543362 RPA200010H 00643169543362 RPA200011H 00643169543362 RPA200012H 00643169543362 RPA200013H 00643169543362 RPA200014H 00643169543362 RPA200015H 00643169543362 RPA200016H 00643169543362 RPA200017H 00643169543362 RPA200018H 00643169543362 RPA200019H 00643169543362 RPA200021H 00643169543362 RPA200022H 00643169543362 RPA200023H 00643169543362 RPA200024H 00643169543362 RPA200025H 00643169543362 RPA200027H 00643169543362 RPA200028H 00643169543362 RPA200029H 00643169543362 RPA200030H 00643169543362 RPA200031H 00643169543362 RPA200032H 00643169543362 RPA200033H 00643169543362 RPA200034H 00643169543362 RPA200035H 00643169543362 RPA200036H 00643169543362 RPA200037H 00643169543362 RPA200039H 00643169543362 RPA200040H 00643169543362 RPA200041H 00643169543362 RPA200042H 00643169543362 RPA200043H 00643169543362 RPA200044H 00643169543362 RPA200045H 00643169543362 RPA200046H 00643169543362 RPA200047H 00643169543362 RPA200048H 00643169543362 RPA200049H 00643169543362 RPA200050H 00643169543362 RPA200051H 00643169543362 RPA200052H 00643169543362 RPA200053H 00643169543362 RPA200054H 00643169543362 RPA200055H 00643169543362 RPA200056H 00643169543362 RPA200057H 00643169543362 RPA200058H 00643169543362 RPA200059H 00643169543362 RPA200060H 00643169543362 RPA200061H 00643169543362 RPA200062H 00643169543362 RPA200063H 00643169543362 RPA200064H 00643169543362 RPA200065H 00643169543362 RPA200066H 00643169543362 RPA200067H 00643169543362 RPA200068H 00643169543362 RPA200069H 00643169543362 RPA200071H 00643169543362 RPA200072H 00643169543362 RPA200073H 00643169543362 RPA200075H 00643169543362 RPA200076H 00643169543362 RPA200077H 00643169543362 RPA200078H 00643169543362 RPA200079H 00643169543362 RPA200080H 00643169543362 RPA200081H 00643169543362 RPA200082H 00643169543362 RPA200083H 00643169543362 RPA200084H 00643169543362 RPA200085H 00643169543362 RPA200086H 00643169543362 RPA200087H 00643169543362 RPA200088H 00643169543362 RPA200089H 00643169543362 RPA200090H 00643169543362 RPA200091H 00643169543362 RPA200092H 00643169543362 RPA200093H 00643169543362 RPA200094H 00643169543362 RPA200095H 00643169543362 RPA200096H 00643169543362 RPA200097H 00643169543362 RPA200098H 00643169543362 RPA200099H 00643169543362 RPA200100H 00643169543362 RPA200101H 00643169543362 RPA200102H 00643169543362 RPA200103H 00643169543362 RPA200104H 00643169543362 RPA200105H 00643169543362 RPA200106H 00643169543362 RPA200107H 00643169543362 RPA200108H 00643169543362 RPA200109H 00643169543362 RPA200110H 00643169543362 RPA200111H 00643169543362 RPA200112H 00643169543362 RPA200113H 00643169543362 RPA200114H 00643169543362 RPA200115H 00643169543362 RPA200116H 00643169543362 RPA200117H 00643169543362 RPA200118H 00643169543362 RPA200119H 00643169543362 RPA200120H 00643169543362 RPA200121H 00643169543362 RPA200122H 00643169543362 RPA200123H 00643169543362 RPA200124H 00643169543362 RPA200125H 00643169543362 RPA200126H 00643169543362 RPA200127H 00643169543362 RPA200128H 00643169543362 RPA200129H 00643169543362 RPA200130H 00643169543362 RPA200131H 00643169543362 RPA200132H 00643169543362 RPA200133H 00643169543362 RPA200134H 00643169543362 RPA200135H 00643169543362 RPA200136H 00643169543362 RPA200137H 00643169543362 RPA200138H 00643169543362 RPA200139H 00643169543362 RPA200140H 00643169543362 RPA200141H 00643169543362 RPA200142H 00643169543362 RPA200143H 00643169543362 RPA200144H 00643169543362 RPA200145H 00643169543362 RPA200146H 00643169543362 RPA200147H 00643169543362 RPA200148H 00643169543362 RPA200149H 00643169543362 RPA200150H 00643169543362 RPA200151H 00643169543362 RPA200152H 00643169543362 RPA200153H 00643169543362 RPA200154H 00643169543362 RPA200155H 00643169543362 RPA200156H 00643169543362 RPA200157H 00643169543362 RPA200158H 00643169543362 RPA200159H 00643169543362 RPA200160H 00643169543362 RPA200161H 00643169543362 RPA200162H 00643169543362 RPA200164H 00643169543362 RPA200165H 00643169543362 RPA200166H 00643169543362 RPA200167H 00643169543362 RPA200168H 00643169543362 RPA200169H 00643169543362 RPA200170H 00643169543362 RPA200172H 00643169543362 RPA200173H 00643169543362 RPA200174H 00643169543362 RPA200175H 00643169543362 RPA200176H 00643169543362 RPA200177H 00643169543362 RPA200178H 00643169543362 RPA200179H 00643169543362 RPA200180H 00643169543362 RPA200181H 00643169543362 RPA200182H 00643169543362 RPA200183H 00643169543362 RPA200184H 00643169543362 RPA200185H 00643169543362 RPA200186H 00643169543362 RPA200187H 00643169543362 RPA200188H 00643169543362 RPA200189H 00643169543362 RPA200190H 00643169543362 RPA200191H 00643169543362 RPA200192H 00643169543362 RPA200193H 00643169543362 RPA200194H 00643169543362 RPA200195H 00643169543362 RPA200196H 00643169543362 RPA200197H 00643169543362 RPA200198H 00643169543362 RPA200199H 00643169543362 RPA200200H 00643169543362 RPA200201H 00643169543362 RPA200202H 00643169543362 RPA200203H 00643169543362 RPA200204H 00643169543362 RPA200205H 00643169543362 RPA200206H 00643169543362 RPA200207H 00643169543362 RPA200208H 00643169543362 RPA200209H 00643169543362 RPA200210H 00643169543362 RPA200211H 00643169543362 RPA200212H 00643169543362 RPA200213H 00643169543362 RPA200214H 00643169543362 RPA200215H 00643169543362 RPA200216H 00643169543362 RPA200217H 00643169543362 RPA200218H 00643169543362 RPA200219H 00643169543362 RPA200220H 00643169543362 RPA200221H 00643169543362 RPA200222H 00643169543362 RPA200223H 00643169543362 RPA200224H 00643169543362 RPA200225H 00643169543362 RPA200226H 00643169543362 RPA200227H 00643169543362 RPA200228H 00643169543362 RPA200230H 00643169543362 RPA200231H 00643169543362 RPA200232H 00643169543362 RPA200233H 00643169543362 RPA200234H 00643169543362 RPA200235H 00643169543362 RPA200236H 00643169543362 RPA200237H 00643169543362 RPA200238H 00643169543362 RPA200239H 00643169543362 RPA200240H 00643169543362 RPA200241H 00643169543362 RPA200242H 00643169543362 RPA200243H 00643169543362 RPA200244H 00643169543362 RPA200245H 00643169543362 RPA200246H 00643169543362 RPA200247H 00643169543362 RPA200248H 00643169543362 RPA200249H 00643169543362 RPA200250H 00643169543362 RPA200251H 00643169543362 RPA200252H 00643169543362 RPA200253H 00643169543362 RPA200254H 00643169543362 RPA200258H 00643169543362 RPA200259H 00643169543362 RPA200260H 00643169543362 RPA200261H 00643169543362 RPA200262H 00643169543362 RPA200263H 00643169543362 RPA200264H 00643169543362 RPA200265H 00643169543362 RPA200266H 00643169543362 RPA200268H 00643169543362 RPA200269H 00643169543362 RPA200270H 00643169543362 RPA200271H 00643169543362 RPA200272H 00643169543362 RPA200273H 00643169543362 RPA200275H 00643169543362 RPA200276H 00643169543362 RPA200277H 00643169543362 RPA200278H 00643169543362 RPA200279H 00643169543362 RPA200280H 00643169543362 RPA200281H 00643169543362 RPA200282H 00643169543362 RPA200284H 00643169543362 RPA200285H 00643169543362 RPA200287H 00643169543362 RPA200291H 00643169543362 RPA200297H 00643169543362 RPA200298H 00643169543362 RPA200299H 00643169543362 RPA200300H 00643169543362 RPA200301H 00643169543362 RPA200302H 00643169543362 RPA200303H 00643169543362 RPA200304H 00643169543362 RPA200305H 00643169543362 RPA200306H 00643169543362 RPA200307H 00643169543362 RPA200309H 00643169543362 RPA200310H 00643169543362 RPA200311H 00643169543362 RPA200312H 00643169543362 RPA200313H 00643169543362 RPA200314H 00643169543362 RPA200315H 00643169543362 RPA200316H 00643169543362 RPA200317H 00643169543362 RPA200318H 00643169543362 RPA200319H 00643169543362 RPA200320H 00643169543362 RPA200321H 00643169543362 RPA200322H 00643169543362 RPA200324H 00643169543362 RPA200325H 00643169543362 RPA200326H 00643169543362 RPA200327H 00643169543362 RPA200328H 00643169543362 RPA200329H 00643169543362 RPA200330H 00643169543362 RPA200331H 00643169543362 RPA200332H 00643169543362 RPA200333H 00643169543362 RPA200334H 00643169543362 RPA200335H 00643169543362 RPA200336H 00643169543362 RPA200337H 00643169543362 RPA200338H 00643169543362 RPA200339H 00643169543362 RPA200341H 00643169543362 RPA200343H 00643169543362 RPA200344H 00643169543362 RPA200345H 00643169543362 RPA200346H 00643169543362 RPA200348H 00643169543362 RPA200349H 00643169543362 RPA200350H 00643169543362 RPA200351H 00643169543362 RPA200352H 00643169543362 RPA200353H 00643169543362 RPA200354H 00643169543362 RPA200355H 00643169543362 RPA200356H 00643169543362 RPA200357H 00643169543362 RPA200358H 00643169543362 RPA200359H 00643169543362 RPA200360H 00643169543362 RPA200361H 00643169543362 RPA200362H 00643169543362 RPA200363H 00643169543362 RPA200364H 00643169543362 RPA200365H 00643169543362 RPA200366H 00643169543362 RPA200367H 00643169543362 RPA200368H 00643169543362 RPA200369H 00643169543362 RPA200372H 00643169543362 RPA200373H 00643169543362 RPA200374H 00643169543362 RPA200375H 00643169543362 RPA200376H 00643169543362 RPA200377H 00643169543362 RPA200378H 00643169543362 RPA200379H 00643169543362 RPA200380H 00643169543362 RPA200381H 00643169543362 RPA200382H 00643169543362 RPA200383H 00643169543362 RPA200384H 00643169543362 RPA200385H 00643169543362 RPA200386H 00643169543362 RPA200387H 00643169543362 RPA200388H 00643169543362 RPA200389H 00643169543362 RPA200390H 00643169543362 RPA200391H 00643169543362 RPA200392H 00643169543362 RPA200393H 00643169543362 RPA200394H 00643169543362 RPA200395H 00643169543362 RPA200396H 00643169543362 RPA200397H 00643169543362 RPA200398H 00643169543362 RPA200399H 00643169543362 RPA200400H 00643169543362 RPA200401H 00643169543362 RPA200402H 00643169543362 RPA200403H 00643169543362 RPA200404H 00643169543362 RPA200405H 00643169543362 RPA200406H 00643169543362 RPA200407H 00643169543362 RPA200408H 00643169543362 RPA200409H 00643169543362 RPA200410H 00643169543362 RPA200411H 00643169543362 RPA200412H 00643169543362 RPA200413H 00643169543362 RPA200414H 00643169543362 RPA200415H 00643169543362 RPA200416H 00643169543362 RPA200417H 00643169543362 RPA200418H 00643169543362 RPA200419H 00643169543362 RPA200420H 00643169543362 RPA200421H 00643169543362 RPA200422H 00643169543362 RPA200423H 00643169543362 RPA200425H 00643169543362 RPA200426H 00643169543362 RPA200427H 00643169543362 RPA200428H 00643169543362 RPA200429H 00643169543362 RPA200430H 00643169543362 RPA200431H 00643169543362 RPA200432H 00643169543362 RPA200433H 00643169543362 RPA200434H 00643169543362 RPA200435H 00643169543362 RPA200436H 00643169543362 RPA200438H 00643169543362 RPA200439H 00643169543362 RPA200440H 00643169543362 RPA200441H 00643169543362 RPA200442H 00643169543362 RPA200443H 00643169543362 RPA200445H 00643169543362 RPA200446H 00643169543362 RPA200447H 00643169543362 RPA200448H 00643169543362 RPA200450H 00643169543362 RPA200452H 00643169543362 RPA200453H 00643169543362 RPA200454H 00643169543362 RPA200455H 00643169543362 RPA200456H 00643169543362 RPA200457H 00643169543362 RPA200458H 00643169543362 RPA200459H 00643169543362 RPA200460H 00643169543362 RPA200461H 00643169543362 RPA200462H 00643169543362 RPA200463H 00643169543362 RPA200464H 00643169543362 RPA200465H 00643169543362 RPA200466H 00643169543362 RPA200467H 00643169543362 RPA200469H 00643169543362 RPA200470H 00643169543362 RPA200471H 00643169543362 RPA200472H 00643169543362 RPA200473H 00643169543362 RPA200474H 00643169543362 RPA200475H 00643169543362 RPA200476H 00643169543362 RPA200477H 00643169543362 RPA200478H 00643169543362 RPA200479H 00643169543362 RPA200480H 00643169543362 RPA200481H 00643169543362 RPA200482H 00643169543362 RPA200484H 00643169543362 RPA200485H 00643169543362 RPA200486H 00643169543362 RPA200487H 00643169543362 RPA200488H 00643169543362 RPA200489H 00643169543362 RPA200490H 00643169543362 RPA200491H 00643169543362 RPA200492H 00643169543362 RPA200493H 00643169543362 RPA200494H 00643169543362 RPA200496H 00643169543362 RPA200497H 00643169543362 RPA200498H 00643169543362 RPA200499H 00643169543362 RPA200500H 00643169543362 RPA200501H 00643169543362 RPA200503H 00643169543362 RPA200504H 00643169543362 RPA200507H 00643169543362 RPA200509H 00643169543362 RPA200510H 00643169543362 RPA200511H 00643169543362 RPA200512H 00643169543362 RPA200513H 00643169543362 RPA200514H 00643169543362 RPA200515H 00643169543362 RPA200516H 00643169543362 RPA200517H 00643169543362 RPA200518H 00643169543362 RPA200519H 00643169543362 RPA200520H 00643169543362 RPA200522H 00643169543362 RPA200523H 00643169543362 RPA200524H 00643169543362 RPA200525H 00643169543362 RPA200526H 00643169543362 RPA200528H 00643169543362 RPA200529H 00643169543362 RPA200530H 00643169543362 RPA200531H 00643169543362 RPA200533H 00643169543362 RPA200534H 00643169543362 RPA200536H 00643169543362 RPA200537H 00643169543362 RPA200538H 00643169543362 RPA200539H 00643169543362 RPA200540H 00643169543362 RPA200541H 00643169543362 RPA200542H 00643169543362 RPA200543H 00643169543362 RPA200544H 00643169543362 RPA200545H 00643169543362 RPA200546H 00643169543362 RPA200547H 00643169543362 RPA200548H 00643169543362 RPA200550H 00643169543362 RPA200551H 00643169543362 RPA200552H 00643169543362 RPA200553H 00643169543362 RPA200554H 00643169543362 RPA200555H 00643169543362 RPA200556H 00643169543362 RPA200557H 00643169543362 RPA200558H 00643169543362 RPA200559H 00643169543362 RPA200560H 00643169543362 RPA200561H 00643169543362 RPA200562H 00643169543362 RPA200563H 00643169543362 RPA200564H 00643169543362 RPA200565H 00643169543362 RPA200566H 00643169543362 RPA200567H 00643169543362 RPA200568H 00643169543362 RPA200569H 00643169543362 RPA200570H 00643169543362 RPA200571H 00643169543362 RPA200572H 00643169543362 RPA200573H 00643169543362 RPA200574H 00643169543362 RPA200575H 00643169543362 RPA200576H 00643169543362 RPA200577H 00643169543362 RPA200579H 00643169543362 RPA200580H 00643169543362 RPA200581H 00643169543362 RPA200582H 00643169543362 RPA200583H 00643169543362 RPA200584H 00643169543362 RPA200585H 00643169543362 RPA200587H 00643169543362 RPA200588H 00643169543362 RPA200589H 00643169543362 RPA200590H 00643169543362 RPA200591H 00643169543362 RPA200592H 00643169543362 RPA200594H 00643169543362 RPA200595H 00643169543362 RPA200596H 00643169543362 RPA200597H 00643169543362 RPA200598H 00643169543362 RPA200599H 00643169543362 RPA200600H 00643169543362 RPA200601H 00643169543362 RPA200602H 00643169543362 RPA200603H 00643169543362 RPA200604H 00643169543362 RPA200605H 00643169543362 RPA200606H 00643169543362 RPA200607H 00643169543362 RPA200608H 00643169543362 RPA200609H 00643169543362 RPA200610H 00643169543362 RPA200611H 00643169543362 RPA200612H 00643169543362 RPA200613H 00643169543362 RPA200614H 00643169543362 RPA200615H 00643169543362 RPA200616H 00643169543362 RPA200617H 00643169543362 RPA200618H 00643169543362 RPA200619H 00643169543362 RPA200620H 00643169543362 RPA200621H 00643169543362 RPA200622H 00643169543362 RPA200623H 00643169543362 RPA200624H 00643169543362 RPA200625H 00643169543362 RPA200626H 00643169543362 RPA200627H 00643169543362 RPA200628H 00643169543362 RPA200629H 00643169543362 RPA200630H 00643169543362 RPA200631H 00643169543362 RPA200632H 00643169543362 RPA200633H 00643169543362 RPA200634H 00643169543362 RPA200635H 00643169543362 RPA200636H 00643169543362 RPA200637H 00643169543362 RPA200638H 00643169543362 RPA200639H 00643169543362 RPA200640H 00643169543362 RPA200641H 00643169543362 RPA200642H 00643169543362 RPA200643H 00643169543362 RPA200644H 00643169543362 RPA200645H 00643169543362 RPA200646H 00643169543362 RPA200647H 00643169543362 RPA200648H 00643169543362 RPA200649H 00643169543362 RPA200651H 00643169543362 RPA200652H 00643169543362 RPA200653H 00643169543362 RPA200654H 00643169543362 RPA200655H 00643169543362 RPA200656H 00643169543362 RPA200657H 00643169543362 RPA200658H 00643169543362 RPA200660H 00643169543362 RPA200662H 00643169543362 RPA200663H 00643169543362 RPA200665H 00643169543362 RPA200666H 00643169543362 RPA200667H 00643169543362 RPA200668H 00643169543362 RPA200669H 00643169543362 RPA200670H 00643169543362 RPA200671H 00643169543362 RPA200672H 00643169543362 RPA200673H 00643169543362 RPA200674H 00643169543362 RPA200675H 00643169543362 RPA200676H 00643169543362 RPA200680H 00643169543362 RPA200681H 00643169543362 RPA200682H 00643169543362 RPA200683H 00643169543362 RPA200685H 00643169543362 RPA200686H 00643169543362 RPA200688H 00643169543362 RPA200689H 00643169543362 RPA200690H 00643169543362 RPA200691H 00643169543362 RPA200692H 00643169543362 RPA200693H 00643169543362 RPA200694H 00643169543362 RPA200696H 00643169543362 RPA200697H 00643169543362 RPA200698H 00643169543362 RPA200700H 00643169543362 RPA200701H 00643169543362 RPA200702H 00643169543362 RPA200703H 00643169543362 RPA200704H 00643169543362 RPA200705H 00643169543362 RPA200706H 00643169543362 RPA200707H 00643169543362 RPA200708H 00643169543362 RPA200709H 00643169543362 RPA200710H 00643169543362 RPA200711H 00643169543362 RPA200712H 00643169543362 RPA200713H 00643169543362 RPA200714H 00643169543362 RPA200715H 00643169543362 RPA200716H 00643169543362 RPA200717H 00643169543362 RPA200718H 00643169543362 RPA200719H 00643169543362 RPA200720H 00643169543362 RPA200721H 00643169543362 RPA200723H 00643169543362 RPA200724H 00643169543362 RPA200725H 00643169543362 RPA200726H 00643169543362 RPA200727H 00643169543362 RPA200728H 00643169543362 RPA200730H 00643169543362 RPA200731H 00643169543362 RPA200732H 00643169543362 RPA200733H 00643169543362 RPA200734H 00643169543362 RPA200735H 00643169543362 RPA200736H 00643169543362 RPA200737H 00643169543362 RPA200741H 00643169543362 RPA200742H 00643169543362 RPA200744H 00643169543362 RPA200746H 00643169543362 RPA200747H 00643169543362 RPA200748H 00643169543362 RPA200749H 00643169543362 RPA200750H 00643169543362 RPA200751H 00643169543362 RPA200752H 00643169543362 RPA200753H 00643169543362 RPA200754H 00643169543362 RPA200755H 00643169543362 RPA200756H 00643169543362 RPA200757H 00643169543362 RPA200758H 00643169543362 RPA200759H 00643169543362 RPA200760H 00643169543362 RPA200761H 00643169543362 RPA200762H 00643169543362 RPA200763H 00643169543362 RPA200764H 00643169543362 RPA200765H 00643169543362 RPA200766H 00643169543362 RPA200767H 00643169543362 RPA200768H 00643169543362 RPA200769H 00643169543362 RPA200770H 00643169543362 RPA200771H 00643169543362 RPA200772H 00643169543362 RPA200773H 00643169543362 RPA200774H 00643169543362 RPA200775H 00643169543362 RPA200776H 00643169543362 RPA200777H 00643169543362 RPA200778H 00643169543362 RPA200779H 00643169543362 RPA200780H 00643169543362 RPA200781H 00643169543362 RPA200782H 00643169543362 RPA200783H 00643169543362 RPA200784H 00643169543362 RPA200785H 00643169543362 RPA200787H 00643169543362 RPA200788H 00643169543362 RPA200789H 00643169543362 RPA200790H 00643169543362 RPA200791H 00643169543362 RPA200792H 00643169543362 RPA200795H 00643169543362 RPA200796H 00643169543362 RPA200798H 00643169543362 RPA200800H 00643169543362 RPA200801H 00643169543362 RPA200806H 00643169543362 RPA200810H 00643169543362 RPA200827H 00643169543362 RPA200839H 00643169543362 RPA200867H 00643169543362 RPA200870H 00643169543362 RPA200913H 00643169543362 RPA200923H 00643169543379 RPB600086S 00643169543379 RPB600087S 00643169543379 RPB600088S 00643169543379 RPB600089S 00643169543379 RPB600090S 00643169543379 RPB600091S 00643169543379 RPB600092S 00643169543379 RPB600093S 00643169543379 RPB600094S 00643169543379 RPB600095S 00643169543379 RPB600096S 00643169543379 RPB600097S 00643169543379 RPB600098S 00643169543379 RPB600099S 00643169543379 RPB600100S 00643169543379 RPB600101S 00643169543379 RPB600102S 00643169543379 RPB600103S 00643169543379 RPB600104S 00643169543379 RPB600105S 00643169543379 RPB600106S 00643169543379 RPB600107S 00643169543379 RPB600108S 00643169543379 RPB600109S 00643169543379 RPB600110S 00643169543379 RPB600111S 00643169543379 RPB600112S 00643169543379 RPB600113S 00643169543379 RPB600114S 00643169543379 RPB600115S 00643169543379 RPB600116S 00643169543379 RPB600117S 00643169543379 RPB600118S 00643169543379 RPB600119S 00643169543379 RPB600120S 00643169543379 RPB600121S 00643169543379 RPB600122S 00643169543379 RPB600123S 00643169543379 RPB600124S 00643169543379 RPB600125S 00643169543379 RPB600126S 00643169543379 RPB600127S 00643169543379 RPB600128S 00643169543379 RPB600129S 00643169543379 RPB600130S 00643169543379 RPB600131S 00643169543379 RPB600132S 00643169543379 RPB600133S 00643169543379 RPB600134S 00643169543379 RPB600135S 00643169543379 RPB600136S 00643169543379 RPB600137S 00643169543379 RPB600138S 00643169543379 RPB600139S 00643169543379 RPB600140S 00643169543379 RPB600141S 00643169543379 RPB600142S 00643169543379 RPB600143S 00643169543379 RPB600144S 00643169543379 RPB600145S 00643169543379 RPB600146S 00643169543379 RPB600147S 00643169543379 RPB600148S 00643169543379 RPB600149S 00643169543379 RPB600150S 00643169543379 RPB600151S 00643169543379 RPB600152S 00643169543379 RPB600153S 00643169543379 RPB600154S 00643169543379 RPB600155S 00643169543379 RPB600156S 00643169543379 RPB600157S 00643169543379 RPB600158S 00643169543379 RPB600159S 00643169543379 RPB600160S 00643169543379 RPB600161S 00643169543379 RPB600162S 00643169543379 RPB600163S 00643169543379 RPB600164S 00643169543379 RPB600165S 00643169543379 RPB600166S 00643169543379 RPB600167S 00643169543379 RPB600168S 00643169543379 RPB600169S 00643169543379 RPB600170S 00643169543379 RPB600171S 00643169543379 RPB600172S 00643169543379 RPB600173S 00643169543379 RPB600174S 00643169543379 RPB600175S 00643169543379 RPB600176S 00643169543379 RPB600177S 00643169543379 RPB600178S 00643169543379 RPB600179S 00643169543379 RPB600180S 00643169543379 RPB600181S 00643169543379 RPB600182S 00643169543379 RPB600183S 00643169543379 RPB600184S 00643169543379 RPB600185S 00643169543379 RPB600186S 00643169543379 RPB600187S 00643169543379 RPB600188S 00643169543379 RPB600189S 00643169543379 RPB600190S 00643169543379 RPB600191S 00643169543379 RPB600192S 00643169543379 RPB600193S 00643169543379 RPB600194S 00643169543379 RPB600195S 00643169543379 RPB600196S 00643169543379 RPB600197S 00643169543379 RPB600198S 00643169543379 RPB600199S 00643169543379 RPB600200S 00643169543379 RPB600201S 00643169543379 RPB600202S 00643169543379 RPB600204S 00643169543379 RPB600205S 00643169543379 RPB600206S 00643169543379 RPB600207S 00643169543379 RPB600208S 00643169543379 RPB600209S 00643169543379 RPB600210S 00643169543379 RPB600211S 00643169543379 RPB600212S 00643169543379 RPB600213S 00643169543379 RPB600214S 00643169543379 RPB600215S 00643169543379 RPB600216S 00643169543379 RPB600217S 00643169543379 RPB600218S 00643169543379 RPB600219S 00643169543379 RPB600220S 00643169543379 RPB600221S 00643169543379 RPB600222S 00643169543379 RPB600223S 00643169543379 RPB600224S 00643169543379 RPB600225S 00643169543379 RPB600226S 00643169543379 RPB600227S 00643169543379 RPB600228S 00643169543379 RPB600229S 00643169543379 RPB600230S 00643169543379 RPB600231S 00643169543379 RPB600232S 00643169543379 RPB600233S 00643169543379 RPB600234S 00643169543379 RPB600235S 00643169543379 RPB600236S 00643169543379 RPB600237S 00643169543379 RPB600238S 00643169543379 RPB600239S 00643169543379 RPB600240S 00643169543379 RPB600241S 00643169543379 RPB600242S 00643169543379 RPB600243S 00643169543379 RPB600244S 00643169543379 RPB600245S 00643169543379 RPB600246S 00643169543379 RPB600247S 00643169543379 RPB600248S 00643169543379 RPB600249S 00643169543379 RPB600250S 00643169543379 RPB600251S 00643169543379 RPB600252S 00643169543379 RPB600253S 00643169543379 RPB600254S 00643169543379 RPB600255S 00643169543379 RPB600256S 00643169543379 RPB600257S 00643169543379 RPB600258S 00643169543379 RPB600259S 00643169543379 RPB600260S 00643169543379 RPB600261S 00643169543379 RPB600262S 00643169543379 RPB600263S 00643169543379 RPB600264S 00643169543379 RPB600265S 00643169543379 RPB600266S 00643169543379 RPB600267S 00643169543379 RPB600268S 00643169543379 RPB600269S 00643169543379 RPB600270S 00643169543379 RPB600271S 00643169543379 RPB600272S 00643169543379 RPB600273S 00643169543379 RPB600274S 00643169543379 RPB600275S 00643169543379 RPB600276S 00643169543379 RPB600277S 00643169543379 RPB600278S 00643169543379 RPB600279S 00643169543379 RPB600280S 00643169543379 RPB600282S 00643169543379 RPB600283S 00643169543379 RPB600284S 00643169543379 RPB600285S 00643169543379 RPB600286S 00643169543379 RPB600287S 00643169543379 RPB600288S 00643169543379 RPB600289S 00643169543379 RPB600290S 00643169543379 RPB600291S 00643169543379 RPB600292S 00643169543379 RPB600293S 00643169543379 RPB600294S 00643169543379 RPB600295S 00643169543379 RPB600296S 00643169543379 RPB600297S 00643169543379 RPB600298S 00643169543379 RPB600299S 00643169543379 RPB600300S 00643169543379 RPB600301S 00643169543379 RPB600302S 00643169543379 RPB600303S 00643169543379 RPB600304S 00643169543379 RPB600305S 00643169543379 RPB600306S 00643169543379 RPB600307S 00643169543379 RPB600308S 00643169543379 RPB600309S 00643169543379 RPB600310S 00643169543379 RPB600311S 00643169543379 RPB600312S 00643169543379 RPB600313S 00643169543379 RPB600314S 00643169543379 RPB600315S 00643169543379 RPB600316S 00643169543379 RPB600317S 00643169543379 RPB600318S 00643169543379 RPB600319S 00643169543379 RPB600320S 00643169543379 RPB600321S 00643169543379 RPB600322S 00643169543379 RPB600323S 00643169543379 RPB600324S 00643169543379 RPB600325S 00643169543379 RPB600326S 00643169543379 RPB600327S 00643169543379 RPB600328S 00643169543379 RPB600329S 00643169543379 RPB600330S 00643169543379 RPB600331S 00643169543379 RPB600332S 00643169543379 RPB600333S 00643169543379 RPB600334S 00643169543379 RPB600335S 00643169543379 RPB600336S 00643169543379 RPB600337S 00643169543379 RPB600338S 00643169543379 RPB600339S 00643169543379 RPB600340S 00643169543379 RPB600341S 00643169543379 RPB600342S 00643169543379 RPB600343S 00643169543379 RPB600344S 00643169543379 RPB600345S 00643169543379 RPB600346S 00643169543379 RPB600347S 00643169543379 RPB600348S 00643169543379 RPB600349S 00643169543379 RPB600350S 00643169543379 RPB600351S 00643169543379 RPB600352S 00643169543379 RPB600353S 00643169543379 RPB600354S 00643169543379 RPB600355S 00643169543379 RPB600356S 00643169543379 RPB600357S 00643169543379 RPB600358S 00643169543379 RPB600359S 00643169543379 RPB600360S 00643169543379 RPB600361S 00643169543379 RPB600362S 00643169543379 RPB600363S 00643169543379 RPB600364S 00643169543379 RPB600365S 00643169543379 RPB600366S 00643169543379 RPB600367S 00643169543379 RPB600368S 00643169543379 RPB600369S 00643169543379 RPB600370S 00643169543379 RPB600371S 00643169543379 RPB600372S 00643169543379 RPB600373S 00643169543379 RPB600374S 00643169543379 RPB600375S 00643169543379 RPB600376S 00643169543379 RPB600377S 00643169543379 RPB600378S 00643169543379 RPB600379S 00643169543379 RPB600380S 00643169543379 RPB600381S 00643169543379 RPB600382S 00643169543379 RPB600383S 00643169543379 RPB600384S 00643169543379 RPB600385S 00643169543379 RPB600386S 00643169543379 RPB600387S 00643169543379 RPB600388S 00643169543379 RPB600389S 00643169543379 RPB600390S 00643169543379 RPB600391S 00643169543379 RPB600392S 00643169543379 RPB600393S 00643169543379 RPB600394S 00643169543379 RPB600395S 00643169543379 RPB600396S 00643169543379 RPB600397S 00643169543379 RPB600398S 00643169543379 RPB600399S 00643169543379 RPB600400S 00643169543379 RPB600401S 00643169543379 RPB600402S 00643169543379 RPB600403S 00643169543379 RPB600404S 00643169543379 RPB600405S 00643169543379 RPB600406S 00643169543379 RPB600407S 00643169543379 RPB600408S 00643169543379 RPB600409S 00643169543379 RPB600410S 00643169543379 RPB600411S 00643169543379 RPB600412S 00643169543379 RPB600413S 00643169543379 RPB600414S 00643169543379 RPB600415S 00643169543379 RPB600416S 00643169543379 RPB600417S 00643169543379 RPB600418S 00643169543379 RPB600419S 00643169543379 RPB600420S 00643169543379 RPB600421S 00643169543379 RPB600422S 00643169543379 RPB600423S 00643169543379 RPB600424S 00643169543379 RPB600425S 00643169543379 RPB600426S 00643169543379 RPB600427S 00643169543379 RPB600428S 00643169543379 RPB600429S 00643169543379 RPB600430S 00643169543379 RPB600431S 00643169543379 RPB600432S 00643169543379 RPB600433S 00643169543379 RPB600434S 00643169543379 RPB600435S 00643169543379 RPB600436S 00643169543379 RPB600437S 00643169543379 RPB600438S 00643169543379 RPB600439S 00643169543379 RPB600440S 00643169543379 RPB600441S 00643169543379 RPB600442S 00643169543379 RPB600443S 00643169543379 RPB600444S 00643169543379 RPB600445S 00643169543379 RPB600446S 00643169543379 RPB600447S 00643169543379 RPB600448S 00643169543379 RPB600449S 00643169543379 RPB600450S 00643169543379 RPB600451S 00643169543379 RPB600452S 00643169543379 RPB600453S 00643169543379 RPB600454S 00643169543379 RPB600455S 00643169543379 RPB600456S 00643169543379 RPB600457S 00643169543379 RPB600458S 00643169543379 RPB600459S 00643169543379 RPB600460S 00643169543379 RPB600461S 00643169543379 RPB600462S 00643169543379 RPB600463S 00643169543379 RPB600464S 00643169543379 RPB600465S 00643169543379 RPB600466S 00643169543379 RPB600467S 00643169543379 RPB600468S 00643169543379 RPB600469S 00643169543379 RPB600470S 00643169543379 RPB600471S 00643169543379 RPB600472S 00643169543379 RPB600473S 00643169543379 RPB600474S 00643169543379 RPB600475S 00643169543379 RPB600476S 00643169543379 RPB600477S 00643169543379 RPB600478S 00643169543379 RPB600479S 00643169543379 RPB600480S 00643169543379 RPB600481S 00643169543379 RPB600482S 00643169543379 RPB600483S 00643169543379 RPB600484S 00643169543379 RPB600485S 00643169543379 RPB600486S 00643169543379 RPB600487S 00643169543379 RPB600488S 00643169543379 RPB600489S 00643169543379 RPB600490S 00643169543379 RPB600491S 00643169543379 RPB600492S 00643169543379 RPB600493S 00643169543379 RPB600494S 00643169543379 RPB600495S 00643169543379 RPB600496S 00643169543379 RPB600497S 00643169543379 RPB600498S 00643169543379 RPB600499S 00643169543379 RPB600500S 00643169543379 RPB600501S 00643169543379 RPB600502S 00643169543379 RPB600503S 00643169543379 RPB600504S 00643169543379 RPB600505S 00643169543379 RPB600506S 00643169543379 RPB600507S 00643169543379 RPB600508S 00643169543379 RPB600509S 00643169543379 RPB600510S 00643169543379 RPB600511S 00643169543379 RPB600512S 00643169543379 RPB600513S 00643169543379 RPB600514S 00643169543379 RPB600515S 00643169543379 RPB600516S 00643169543379 RPB600517S 00643169543379 RPB600518S 00643169543379 RPB600519S 00643169543379 RPB600520S 00643169543379 RPB600521S 00643169543379 RPB600522S 00643169543379 RPB600523S 00643169543379 RPB600524S 00643169543379 RPB600525S 00643169543379 RPB600526S 00643169543379 RPB600527S 00643169543379 RPB600528S 00643169543379 RPB600529S 00643169543379 RPB600530S 00643169543379 RPB600531S 00643169543379 RPB600532S 00643169543379 RPB600533S 00643169543379 RPB600534S 00643169543379 RPB600535S 00643169543379 RPB600536S 00643169543379 RPB600537S 00643169543379 RPB600538S 00643169543379 RPB600539S 00643169543379 RPB600540S 00643169543379 RPB600541S 00643169543379 RPB600542S 00643169543379 RPB600543S 00643169543379 RPB600544S 00643169543379 RPB600545S 00643169543379 RPB600546S 00643169543379 RPB600547S 00643169543379 RPB600548S 00643169543379 RPB600549S 00643169543379 RPB600550S 00643169543379 RPB600551S 00643169543379 RPB600552S 00643169543379 RPB600553S 00643169543379 RPB600554S 00643169543379 RPB600555S 00643169543379 RPB600556S 00643169543379 RPB600557S 00643169543379 RPB600558S
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Medtronic Inc., Cardiac Rhythm and Heart Failure
- Reason for Recall:
- During internal testing conducted as part of next generation product development, it was discovered that the AdaptivCRT (aCRT) feature of Amplia MRI and Claria MRI CRT -D devices may appear to be enabled when it is not active.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Amplia MRI" CRT -D SureScan", Amplia MRI" Quad CRT -D SureScan", Claria MRI" CRT -D SureScan", Claria MRI" Quad CRT -D SureScan. implantable medical devices that automatically detect and treat episodes of ventricular fibrillation, ventricular tachycardia, fast ventricular tachycardia, and bradyarrhythmia.
Product Codes/Lot Numbers:
Serial Numbers of AMPLIA & CLARIA Released from Manufacturing as of 22-Nov-2016 UPN Serial 00643169543362 RPA200003H 00643169543362 RPA200004H 00643169543362 RPA200005H 00643169543362 RPA200006H 00643169543362 RPA200008H 00643169543362 RPA200009H 00643169543362 RPA200010H 00643169543362 RPA200011H 00643169543362 RPA200012H 00643169543362 RPA200013H 00643169543362 RPA200014H 00643169543362 RPA200015H 00643169543362 RPA200016H 00643169543362 RPA200017H 00643169543362 RPA200018H 00643169543362 RPA200019H 00643169543362 RPA200021H 00643169543362 RPA200022H 00643169543362 RPA200023H 00643169543362 RPA200024H 00643169543362 RPA200025H 00643169543362 RPA200027H 00643169543362 RPA200028H 00643169543362 RPA200029H 00643169543362 RPA200030H 00643169543362 RPA200031H 00643169543362 RPA200032H 00643169543362 RPA200033H 00643169543362 RPA200034H 00643169543362 RPA200035H 00643169543362 RPA200036H 00643169543362 RPA200037H 00643169543362 RPA200039H 00643169543362 RPA200040H 00643169543362 RPA200041H 00643169543362 RPA200042H 00643169543362 RPA200043H 00643169543362 RPA200044H 00643169543362 RPA200045H 00643169543362 RPA200046H 00643169543362 RPA200047H 00643169543362 RPA200048H 00643169543362 RPA200049H 00643169543362 RPA200050H 00643169543362 RPA200051H 00643169543362 RPA200052H 00643169543362 RPA200053H 00643169543362 RPA200054H 00643169543362 RPA200055H 00643169543362 RPA200056H 00643169543362 RPA200057H 00643169543362 RPA200058H 00643169543362 RPA200059H 00643169543362 RPA200060H 00643169543362 RPA200061H 00643169543362 RPA200062H 00643169543362 RPA200063H 00643169543362 RPA200064H 00643169543362 RPA200065H 00643169543362 RPA200066H 00643169543362 RPA200067H 00643169543362 RPA200068H 00643169543362 RPA200069H 00643169543362 RPA200071H 00643169543362 RPA200072H 00643169543362 RPA200073H 00643169543362 RPA200075H 00643169543362 RPA200076H 00643169543362 RPA200077H 00643169543362 RPA200078H 00643169543362 RPA200079H 00643169543362 RPA200080H 00643169543362 RPA200081H 00643169543362 RPA200082H 00643169543362 RPA200083H 00643169543362 RPA200084H 00643169543362 RPA200085H 00643169543362 RPA200086H 00643169543362 RPA200087H 00643169543362 RPA200088H 00643169543362 RPA200089H 00643169543362 RPA200090H 00643169543362 RPA200091H 00643169543362 RPA200092H 00643169543362 RPA200093H 00643169543362 RPA200094H 00643169543362 RPA200095H 00643169543362 RPA200096H 00643169543362 RPA200097H 00643169543362 RPA200098H 00643169543362 RPA200099H 00643169543362 RPA200100H 00643169543362 RPA200101H 00643169543362 RPA200102H 00643169543362 RPA200103H 00643169543362 RPA200104H 00643169543362 RPA200105H 00643169543362 RPA200106H 00643169543362 RPA200107H 00643169543362 RPA200108H 00643169543362 RPA200109H 00643169543362 RPA200110H 00643169543362 RPA200111H 00643169543362 RPA200112H 00643169543362 RPA200113H 00643169543362 RPA200114H 00643169543362 RPA200115H 00643169543362 RPA200116H 00643169543362 RPA200117H 00643169543362 RPA200118H 00643169543362 RPA200119H 00643169543362 RPA200120H 00643169543362 RPA200121H 00643169543362 RPA200122H 00643169543362 RPA200123H 00643169543362 RPA200124H 00643169543362 RPA200125H 00643169543362 RPA200126H 00643169543362 RPA200127H 00643169543362 RPA200128H 00643169543362 RPA200129H 00643169543362 RPA200130H 00643169543362 RPA200131H 00643169543362 RPA200132H 00643169543362 RPA200133H 00643169543362 RPA200134H 00643169543362 RPA200135H 00643169543362 RPA200136H 00643169543362 RPA200137H 00643169543362 RPA200138H 00643169543362 RPA200139H 00643169543362 RPA200140H 00643169543362 RPA200141H 00643169543362 RPA200142H 00643169543362 RPA200143H 00643169543362 RPA200144H 00643169543362 RPA200145H 00643169543362 RPA200146H 00643169543362 RPA200147H 00643169543362 RPA200148H 00643169543362 RPA200149H 00643169543362 RPA200150H 00643169543362 RPA200151H 00643169543362 RPA200152H 00643169543362 RPA200153H 00643169543362 RPA200154H 00643169543362 RPA200155H 00643169543362 RPA200156H 00643169543362 RPA200157H 00643169543362 RPA200158H 00643169543362 RPA200159H 00643169543362 RPA200160H 00643169543362 RPA200161H 00643169543362 RPA200162H 00643169543362 RPA200164H 00643169543362 RPA200165H 00643169543362 RPA200166H 00643169543362 RPA200167H 00643169543362 RPA200168H 00643169543362 RPA200169H 00643169543362 RPA200170H 00643169543362 RPA200172H 00643169543362 RPA200173H 00643169543362 RPA200174H 00643169543362 RPA200175H 00643169543362 RPA200176H 00643169543362 RPA200177H 00643169543362 RPA200178H 00643169543362 RPA200179H 00643169543362 RPA200180H 00643169543362 RPA200181H 00643169543362 RPA200182H 00643169543362 RPA200183H 00643169543362 RPA200184H 00643169543362 RPA200185H 00643169543362 RPA200186H 00643169543362 RPA200187H 00643169543362 RPA200188H 00643169543362 RPA200189H 00643169543362 RPA200190H 00643169543362 RPA200191H 00643169543362 RPA200192H 00643169543362 RPA200193H 00643169543362 RPA200194H 00643169543362 RPA200195H 00643169543362 RPA200196H 00643169543362 RPA200197H 00643169543362 RPA200198H 00643169543362 RPA200199H 00643169543362 RPA200200H 00643169543362 RPA200201H 00643169543362 RPA200202H 00643169543362 RPA200203H 00643169543362 RPA200204H 00643169543362 RPA200205H 00643169543362 RPA200206H 00643169543362 RPA200207H 00643169543362 RPA200208H 00643169543362 RPA200209H 00643169543362 RPA200210H 00643169543362 RPA200211H 00643169543362 RPA200212H 00643169543362 RPA200213H 00643169543362 RPA200214H 00643169543362 RPA200215H 00643169543362 RPA200216H 00643169543362 RPA200217H 00643169543362 RPA200218H 00643169543362 RPA200219H 00643169543362 RPA200220H 00643169543362 RPA200221H 00643169543362 RPA200222H 00643169543362 RPA200223H 00643169543362 RPA200224H 00643169543362 RPA200225H 00643169543362 RPA200226H 00643169543362 RPA200227H 00643169543362 RPA200228H 00643169543362 RPA200230H 00643169543362 RPA200231H 00643169543362 RPA200232H 00643169543362 RPA200233H 00643169543362 RPA200234H 00643169543362 RPA200235H 00643169543362 RPA200236H 00643169543362 RPA200237H 00643169543362 RPA200238H 00643169543362 RPA200239H 00643169543362 RPA200240H 00643169543362 RPA200241H 00643169543362 RPA200242H 00643169543362 RPA200243H 00643169543362 RPA200244H 00643169543362 RPA200245H 00643169543362 RPA200246H 00643169543362 RPA200247H 00643169543362 RPA200248H 00643169543362 RPA200249H 00643169543362 RPA200250H 00643169543362 RPA200251H 00643169543362 RPA200252H 00643169543362 RPA200253H 00643169543362 RPA200254H 00643169543362 RPA200258H 00643169543362 RPA200259H 00643169543362 RPA200260H 00643169543362 RPA200261H 00643169543362 RPA200262H 00643169543362 RPA200263H 00643169543362 RPA200264H 00643169543362 RPA200265H 00643169543362 RPA200266H 00643169543362 RPA200268H 00643169543362 RPA200269H 00643169543362 RPA200270H 00643169543362 RPA200271H 00643169543362 RPA200272H 00643169543362 RPA200273H 00643169543362 RPA200275H 00643169543362 RPA200276H 00643169543362 RPA200277H 00643169543362 RPA200278H 00643169543362 RPA200279H 00643169543362 RPA200280H 00643169543362 RPA200281H 00643169543362 RPA200282H 00643169543362 RPA200284H 00643169543362 RPA200285H 00643169543362 RPA200287H 00643169543362 RPA200291H 00643169543362 RPA200297H 00643169543362 RPA200298H 00643169543362 RPA200299H 00643169543362 RPA200300H 00643169543362 RPA200301H 00643169543362 RPA200302H 00643169543362 RPA200303H 00643169543362 RPA200304H 00643169543362 RPA200305H 00643169543362 RPA200306H 00643169543362 RPA200307H 00643169543362 RPA200309H 00643169543362 RPA200310H 00643169543362 RPA200311H 00643169543362 RPA200312H 00643169543362 RPA200313H 00643169543362 RPA200314H 00643169543362 RPA200315H 00643169543362 RPA200316H 00643169543362 RPA200317H 00643169543362 RPA200318H 00643169543362 RPA200319H 00643169543362 RPA200320H 00643169543362 RPA200321H 00643169543362 RPA200322H 00643169543362 RPA200324H 00643169543362 RPA200325H 00643169543362 RPA200326H 00643169543362 RPA200327H 00643169543362 RPA200328H 00643169543362 RPA200329H 00643169543362 RPA200330H 00643169543362 RPA200331H 00643169543362 RPA200332H 00643169543362 RPA200333H 00643169543362 RPA200334H 00643169543362 RPA200335H 00643169543362 RPA200336H 00643169543362 RPA200337H 00643169543362 RPA200338H 00643169543362 RPA200339H 00643169543362 RPA200341H 00643169543362 RPA200343H 00643169543362 RPA200344H 00643169543362 RPA200345H 00643169543362 RPA200346H 00643169543362 RPA200348H 00643169543362 RPA200349H 00643169543362 RPA200350H 00643169543362 RPA200351H 00643169543362 RPA200352H 00643169543362 RPA200353H 00643169543362 RPA200354H 00643169543362 RPA200355H 00643169543362 RPA200356H 00643169543362 RPA200357H 00643169543362 RPA200358H 00643169543362 RPA200359H 00643169543362 RPA200360H 00643169543362 RPA200361H 00643169543362 RPA200362H 00643169543362 RPA200363H 00643169543362 RPA200364H 00643169543362 RPA200365H 00643169543362 RPA200366H 00643169543362 RPA200367H 00643169543362 RPA200368H 00643169543362 RPA200369H 00643169543362 RPA200372H 00643169543362 RPA200373H 00643169543362 RPA200374H 00643169543362 RPA200375H 00643169543362 RPA200376H 00643169543362 RPA200377H 00643169543362 RPA200378H 00643169543362 RPA200379H 00643169543362 RPA200380H 00643169543362 RPA200381H 00643169543362 RPA200382H 00643169543362 RPA200383H 00643169543362 RPA200384H 00643169543362 RPA200385H 00643169543362 RPA200386H 00643169543362 RPA200387H 00643169543362 RPA200388H 00643169543362 RPA200389H 00643169543362 RPA200390H 00643169543362 RPA200391H 00643169543362 RPA200392H 00643169543362 RPA200393H 00643169543362 RPA200394H 00643169543362 RPA200395H 00643169543362 RPA200396H 00643169543362 RPA200397H 00643169543362 RPA200398H 00643169543362 RPA200399H 00643169543362 RPA200400H 00643169543362 RPA200401H 00643169543362 RPA200402H 00643169543362 RPA200403H 00643169543362 RPA200404H 00643169543362 RPA200405H 00643169543362 RPA200406H 00643169543362 RPA200407H 00643169543362 RPA200408H 00643169543362 RPA200409H 00643169543362 RPA200410H 00643169543362 RPA200411H 00643169543362 RPA200412H 00643169543362 RPA200413H 00643169543362 RPA200414H 00643169543362 RPA200415H 00643169543362 RPA200416H 00643169543362 RPA200417H 00643169543362 RPA200418H 00643169543362 RPA200419H 00643169543362 RPA200420H 00643169543362 RPA200421H 00643169543362 RPA200422H 00643169543362 RPA200423H 00643169543362 RPA200425H 00643169543362 RPA200426H 00643169543362 RPA200427H 00643169543362 RPA200428H 00643169543362 RPA200429H 00643169543362 RPA200430H 00643169543362 RPA200431H 00643169543362 RPA200432H 00643169543362 RPA200433H 00643169543362 RPA200434H 00643169543362 RPA200435H 00643169543362 RPA200436H 00643169543362 RPA200438H 00643169543362 RPA200439H 00643169543362 RPA200440H 00643169543362 RPA200441H 00643169543362 RPA200442H 00643169543362 RPA200443H 00643169543362 RPA200445H 00643169543362 RPA200446H 00643169543362 RPA200447H 00643169543362 RPA200448H 00643169543362 RPA200450H 00643169543362 RPA200452H 00643169543362 RPA200453H 00643169543362 RPA200454H 00643169543362 RPA200455H 00643169543362 RPA200456H 00643169543362 RPA200457H 00643169543362 RPA200458H 00643169543362 RPA200459H 00643169543362 RPA200460H 00643169543362 RPA200461H 00643169543362 RPA200462H 00643169543362 RPA200463H 00643169543362 RPA200464H 00643169543362 RPA200465H 00643169543362 RPA200466H 00643169543362 RPA200467H 00643169543362 RPA200469H 00643169543362 RPA200470H 00643169543362 RPA200471H 00643169543362 RPA200472H 00643169543362 RPA200473H 00643169543362 RPA200474H 00643169543362 RPA200475H 00643169543362 RPA200476H 00643169543362 RPA200477H 00643169543362 RPA200478H 00643169543362 RPA200479H 00643169543362 RPA200480H 00643169543362 RPA200481H 00643169543362 RPA200482H 00643169543362 RPA200484H 00643169543362 RPA200485H 00643169543362 RPA200486H 00643169543362 RPA200487H 00643169543362 RPA200488H 00643169543362 RPA200489H 00643169543362 RPA200490H 00643169543362 RPA200491H 00643169543362 RPA200492H 00643169543362 RPA200493H 00643169543362 RPA200494H 00643169543362 RPA200496H 00643169543362 RPA200497H 00643169543362 RPA200498H 00643169543362 RPA200499H 00643169543362 RPA200500H 00643169543362 RPA200501H 00643169543362 RPA200503H 00643169543362 RPA200504H 00643169543362 RPA200507H 00643169543362 RPA200509H 00643169543362 RPA200510H 00643169543362 RPA200511H 00643169543362 RPA200512H 00643169543362 RPA200513H 00643169543362 RPA200514H 00643169543362 RPA200515H 00643169543362 RPA200516H 00643169543362 RPA200517H 00643169543362 RPA200518H 00643169543362 RPA200519H 00643169543362 RPA200520H 00643169543362 RPA200522H 00643169543362 RPA200523H 00643169543362 RPA200524H 00643169543362 RPA200525H 00643169543362 RPA200526H 00643169543362 RPA200528H 00643169543362 RPA200529H 00643169543362 RPA200530H 00643169543362 RPA200531H 00643169543362 RPA200533H 00643169543362 RPA200534H 00643169543362 RPA200536H 00643169543362 RPA200537H 00643169543362 RPA200538H 00643169543362 RPA200539H 00643169543362 RPA200540H 00643169543362 RPA200541H 00643169543362 RPA200542H 00643169543362 RPA200543H 00643169543362 RPA200544H 00643169543362 RPA200545H 00643169543362 RPA200546H 00643169543362 RPA200547H 00643169543362 RPA200548H 00643169543362 RPA200550H 00643169543362 RPA200551H 00643169543362 RPA200552H 00643169543362 RPA200553H 00643169543362 RPA200554H 00643169543362 RPA200555H 00643169543362 RPA200556H 00643169543362 RPA200557H 00643169543362 RPA200558H 00643169543362 RPA200559H 00643169543362 RPA200560H 00643169543362 RPA200561H 00643169543362 RPA200562H 00643169543362 RPA200563H 00643169543362 RPA200564H 00643169543362 RPA200565H 00643169543362 RPA200566H 00643169543362 RPA200567H 00643169543362 RPA200568H 00643169543362 RPA200569H 00643169543362 RPA200570H 00643169543362 RPA200571H 00643169543362 RPA200572H 00643169543362 RPA200573H 00643169543362 RPA200574H 00643169543362 RPA200575H 00643169543362 RPA200576H 00643169543362 RPA200577H 00643169543362 RPA200579H 00643169543362 RPA200580H 00643169543362 RPA200581H 00643169543362 RPA200582H 00643169543362 RPA200583H 00643169543362 RPA200584H 00643169543362 RPA200585H 00643169543362 RPA200587H 00643169543362 RPA200588H 00643169543362 RPA200589H 00643169543362 RPA200590H 00643169543362 RPA200591H 00643169543362 RPA200592H 00643169543362 RPA200594H 00643169543362 RPA200595H 00643169543362 RPA200596H 00643169543362 RPA200597H 00643169543362 RPA200598H 00643169543362 RPA200599H 00643169543362 RPA200600H 00643169543362 RPA200601H 00643169543362 RPA200602H 00643169543362 RPA200603H 00643169543362 RPA200604H 00643169543362 RPA200605H 00643169543362 RPA200606H 00643169543362 RPA200607H 00643169543362 RPA200608H 00643169543362 RPA200609H 00643169543362 RPA200610H 00643169543362 RPA200611H 00643169543362 RPA200612H 00643169543362 RPA200613H 00643169543362 RPA200614H 00643169543362 RPA200615H 00643169543362 RPA200616H 00643169543362 RPA200617H 00643169543362 RPA200618H 00643169543362 RPA200619H 00643169543362 RPA200620H 00643169543362 RPA200621H 00643169543362 RPA200622H 00643169543362 RPA200623H 00643169543362 RPA200624H 00643169543362 RPA200625H 00643169543362 RPA200626H 00643169543362 RPA200627H 00643169543362 RPA200628H 00643169543362 RPA200629H 00643169543362 RPA200630H 00643169543362 RPA200631H 00643169543362 RPA200632H 00643169543362 RPA200633H 00643169543362 RPA200634H 00643169543362 RPA200635H 00643169543362 RPA200636H 00643169543362 RPA200637H 00643169543362 RPA200638H 00643169543362 RPA200639H 00643169543362 RPA200640H 00643169543362 RPA200641H 00643169543362 RPA200642H 00643169543362 RPA200643H 00643169543362 RPA200644H 00643169543362 RPA200645H 00643169543362 RPA200646H 00643169543362 RPA200647H 00643169543362 RPA200648H 00643169543362 RPA200649H 00643169543362 RPA200651H 00643169543362 RPA200652H 00643169543362 RPA200653H 00643169543362 RPA200654H 00643169543362 RPA200655H 00643169543362 RPA200656H 00643169543362 RPA200657H 00643169543362 RPA200658H 00643169543362 RPA200660H 00643169543362 RPA200662H 00643169543362 RPA200663H 00643169543362 RPA200665H 00643169543362 RPA200666H 00643169543362 RPA200667H 00643169543362 RPA200668H 00643169543362 RPA200669H 00643169543362 RPA200670H 00643169543362 RPA200671H 00643169543362 RPA200672H 00643169543362 RPA200673H 00643169543362 RPA200674H 00643169543362 RPA200675H 00643169543362 RPA200676H 00643169543362 RPA200680H 00643169543362 RPA200681H 00643169543362 RPA200682H 00643169543362 RPA200683H 00643169543362 RPA200685H 00643169543362 RPA200686H 00643169543362 RPA200688H 00643169543362 RPA200689H 00643169543362 RPA200690H 00643169543362 RPA200691H 00643169543362 RPA200692H 00643169543362 RPA200693H 00643169543362 RPA200694H 00643169543362 RPA200696H 00643169543362 RPA200697H 00643169543362 RPA200698H 00643169543362 RPA200700H 00643169543362 RPA200701H 00643169543362 RPA200702H 00643169543362 RPA200703H 00643169543362 RPA200704H 00643169543362 RPA200705H 00643169543362 RPA200706H 00643169543362 RPA200707H 00643169543362 RPA200708H 00643169543362 RPA200709H 00643169543362 RPA200710H 00643169543362 RPA200711H 00643169543362 RPA200712H 00643169543362 RPA200713H 00643169543362 RPA200714H 00643169543362 RPA200715H 00643169543362 RPA200716H 00643169543362 RPA200717H 00643169543362 RPA200718H 00643169543362 RPA200719H 00643169543362 RPA200720H 00643169543362 RPA200721H 00643169543362 RPA200723H 00643169543362 RPA200724H 00643169543362 RPA200725H 00643169543362 RPA200726H 00643169543362 RPA200727H 00643169543362 RPA200728H 00643169543362 RPA200730H 00643169543362 RPA200731H 00643169543362 RPA200732H 00643169543362 RPA200733H 00643169543362 RPA200734H 00643169543362 RPA200735H 00643169543362 RPA200736H 00643169543362 RPA200737H 00643169543362 RPA200741H 00643169543362 RPA200742H 00643169543362 RPA200744H 00643169543362 RPA200746H 00643169543362 RPA200747H 00643169543362 RPA200748H 00643169543362 RPA200749H 00643169543362 RPA200750H 00643169543362 RPA200751H 00643169543362 RPA200752H 00643169543362 RPA200753H 00643169543362 RPA200754H 00643169543362 RPA200755H 00643169543362 RPA200756H 00643169543362 RPA200757H 00643169543362 RPA200758H 00643169543362 RPA200759H 00643169543362 RPA200760H 00643169543362 RPA200761H 00643169543362 RPA200762H 00643169543362 RPA200763H 00643169543362 RPA200764H 00643169543362 RPA200765H 00643169543362 RPA200766H 00643169543362 RPA200767H 00643169543362 RPA200768H 00643169543362 RPA200769H 00643169543362 RPA200770H 00643169543362 RPA200771H 00643169543362 RPA200772H 00643169543362 RPA200773H 00643169543362 RPA200774H 00643169543362 RPA200775H 00643169543362 RPA200776H 00643169543362 RPA200777H 00643169543362 RPA200778H 00643169543362 RPA200779H 00643169543362 RPA200780H 00643169543362 RPA200781H 00643169543362 RPA200782H 00643169543362 RPA200783H 00643169543362 RPA200784H 00643169543362 RPA200785H 00643169543362 RPA200787H 00643169543362 RPA200788H 00643169543362 RPA200789H 00643169543362 RPA200790H 00643169543362 RPA200791H 00643169543362 RPA200792H 00643169543362 RPA200795H 00643169543362 RPA200796H 00643169543362 RPA200798H 00643169543362 RPA200800H 00643169543362 RPA200801H 00643169543362 RPA200806H 00643169543362 RPA200810H 00643169543362 RPA200827H 00643169543362 RPA200839H 00643169543362 RPA200867H 00643169543362 RPA200870H 00643169543362 RPA200913H 00643169543362 RPA200923H 00643169543379 RPB600086S 00643169543379 RPB600087S 00643169543379 RPB600088S 00643169543379 RPB600089S 00643169543379 RPB600090S 00643169543379 RPB600091S 00643169543379 RPB600092S 00643169543379 RPB600093S 00643169543379 RPB600094S 00643169543379 RPB600095S 00643169543379 RPB600096S 00643169543379 RPB600097S 00643169543379 RPB600098S 00643169543379 RPB600099S 00643169543379 RPB600100S 00643169543379 RPB600101S 00643169543379 RPB600102S 00643169543379 RPB600103S 00643169543379 RPB600104S 00643169543379 RPB600105S 00643169543379 RPB600106S 00643169543379 RPB600107S 00643169543379 RPB600108S 00643169543379 RPB600109S 00643169543379 RPB600110S 00643169543379 RPB600111S 00643169543379 RPB600112S 00643169543379 RPB600113S 00643169543379 RPB600114S 00643169543379 RPB600115S 00643169543379 RPB600116S 00643169543379 RPB600117S 00643169543379 RPB600118S 00643169543379 RPB600119S 00643169543379 RPB600120S 00643169543379 RPB600121S 00643169543379 RPB600122S 00643169543379 RPB600123S 00643169543379 RPB600124S 00643169543379 RPB600125S 00643169543379 RPB600126S 00643169543379 RPB600127S 00643169543379 RPB600128S 00643169543379 RPB600129S 00643169543379 RPB600130S 00643169543379 RPB600131S 00643169543379 RPB600132S 00643169543379 RPB600133S 00643169543379 RPB600134S 00643169543379 RPB600135S 00643169543379 RPB600136S 00643169543379 RPB600137S 00643169543379 RPB600138S 00643169543379 RPB600139S 00643169543379 RPB600140S 00643169543379 RPB600141S 00643169543379 RPB600142S 00643169543379 RPB600143S 00643169543379 RPB600144S 00643169543379 RPB600145S 00643169543379 RPB600146S 00643169543379 RPB600147S 00643169543379 RPB600148S 00643169543379 RPB600149S 00643169543379 RPB600150S 00643169543379 RPB600151S 00643169543379 RPB600152S 00643169543379 RPB600153S 00643169543379 RPB600154S 00643169543379 RPB600155S 00643169543379 RPB600156S 00643169543379 RPB600157S 00643169543379 RPB600158S 00643169543379 RPB600159S 00643169543379 RPB600160S 00643169543379 RPB600161S 00643169543379 RPB600162S 00643169543379 RPB600163S 00643169543379 RPB600164S 00643169543379 RPB600165S 00643169543379 RPB600166S 00643169543379 RPB600167S 00643169543379 RPB600168S 00643169543379 RPB600169S 00643169543379 RPB600170S 00643169543379 RPB600171S 00643169543379 RPB600172S 00643169543379 RPB600173S 00643169543379 RPB600174S 00643169543379 RPB600175S 00643169543379 RPB600176S 00643169543379 RPB600177S 00643169543379 RPB600178S 00643169543379 RPB600179S 00643169543379 RPB600180S 00643169543379 RPB600181S 00643169543379 RPB600182S 00643169543379 RPB600183S 00643169543379 RPB600184S 00643169543379 RPB600185S 00643169543379 RPB600186S 00643169543379 RPB600187S 00643169543379 RPB600188S 00643169543379 RPB600189S 00643169543379 RPB600190S 00643169543379 RPB600191S 00643169543379 RPB600192S 00643169543379 RPB600193S 00643169543379 RPB600194S 00643169543379 RPB600195S 00643169543379 RPB600196S 00643169543379 RPB600197S 00643169543379 RPB600198S 00643169543379 RPB600199S 00643169543379 RPB600200S 00643169543379 RPB600201S 00643169543379 RPB600202S 00643169543379 RPB600204S 00643169543379 RPB600205S 00643169543379 RPB600206S 00643169543379 RPB600207S 00643169543379 RPB600208S 00643169543379 RPB600209S 00643169543379 RPB600210S 00643169543379 RPB600211S 00643169543379 RPB600212S 00643169543379 RPB600213S 00643169543379 RPB600214S 00643169543379 RPB600215S 00643169543379 RPB600216S 00643169543379 RPB600217S 00643169543379 RPB600218S 00643169543379 RPB600219S 00643169543379 RPB600220S 00643169543379 RPB600221S 00643169543379 RPB600222S 00643169543379 RPB600223S 00643169543379 RPB600224S 00643169543379 RPB600225S 00643169543379 RPB600226S 00643169543379 RPB600227S 00643169543379 RPB600228S 00643169543379 RPB600229S 00643169543379 RPB600230S 00643169543379 RPB600231S 00643169543379 RPB600232S 00643169543379 RPB600233S 00643169543379 RPB600234S 00643169543379 RPB600235S 00643169543379 RPB600236S 00643169543379 RPB600237S 00643169543379 RPB600238S 00643169543379 RPB600239S 00643169543379 RPB600240S 00643169543379 RPB600241S 00643169543379 RPB600242S 00643169543379 RPB600243S 00643169543379 RPB600244S 00643169543379 RPB600245S 00643169543379 RPB600246S 00643169543379 RPB600247S 00643169543379 RPB600248S 00643169543379 RPB600249S 00643169543379 RPB600250S 00643169543379 RPB600251S 00643169543379 RPB600252S 00643169543379 RPB600253S 00643169543379 RPB600254S 00643169543379 RPB600255S 00643169543379 RPB600256S 00643169543379 RPB600257S 00643169543379 RPB600258S 00643169543379 RPB600259S 00643169543379 RPB600260S 00643169543379 RPB600261S 00643169543379 RPB600262S 00643169543379 RPB600263S 00643169543379 RPB600264S 00643169543379 RPB600265S 00643169543379 RPB600266S 00643169543379 RPB600267S 00643169543379 RPB600268S 00643169543379 RPB600269S 00643169543379 RPB600270S 00643169543379 RPB600271S 00643169543379 RPB600272S 00643169543379 RPB600273S 00643169543379 RPB600274S 00643169543379 RPB600275S 00643169543379 RPB600276S 00643169543379 RPB600277S 00643169543379 RPB600278S 00643169543379 RPB600279S 00643169543379 RPB600280S 00643169543379 RPB600282S 00643169543379 RPB600283S 00643169543379 RPB600284S 00643169543379 RPB600285S 00643169543379 RPB600286S 00643169543379 RPB600287S 00643169543379 RPB600288S 00643169543379 RPB600289S 00643169543379 RPB600290S 00643169543379 RPB600291S 00643169543379 RPB600292S 00643169543379 RPB600293S 00643169543379 RPB600294S 00643169543379 RPB600295S 00643169543379 RPB600296S 00643169543379 RPB600297S 00643169543379 RPB600298S 00643169543379 RPB600299S 00643169543379 RPB600300S 00643169543379 RPB600301S 00643169543379 RPB600302S 00643169543379 RPB600303S 00643169543379 RPB600304S 00643169543379 RPB600305S 00643169543379 RPB600306S 00643169543379 RPB600307S 00643169543379 RPB600308S 00643169543379 RPB600309S 00643169543379 RPB600310S 00643169543379 RPB600311S 00643169543379 RPB600312S 00643169543379 RPB600313S 00643169543379 RPB600314S 00643169543379 RPB600315S 00643169543379 RPB600316S 00643169543379 RPB600317S 00643169543379 RPB600318S 00643169543379 RPB600319S 00643169543379 RPB600320S 00643169543379 RPB600321S 00643169543379 RPB600322S 00643169543379 RPB600323S 00643169543379 RPB600324S 00643169543379 RPB600325S 00643169543379 RPB600326S 00643169543379 RPB600327S 00643169543379 RPB600328S 00643169543379 RPB600329S 00643169543379 RPB600330S 00643169543379 RPB600331S 00643169543379 RPB600332S 00643169543379 RPB600333S 00643169543379 RPB600334S 00643169543379 RPB600335S 00643169543379 RPB600336S 00643169543379 RPB600337S 00643169543379 RPB600338S 00643169543379 RPB600339S 00643169543379 RPB600340S 00643169543379 RPB600341S 00643169543379 RPB600342S 00643169543379 RPB600343S 00643169543379 RPB600344S 00643169543379 RPB600345S 00643169543379 RPB600346S 00643169543379 RPB600347S 00643169543379 RPB600348S 00643169543379 RPB600349S 00643169543379 RPB600350S 00643169543379 RPB600351S 00643169543379 RPB600352S 00643169543379 RPB600353S 00643169543379 RPB600354S 00643169543379 RPB600355S 00643169543379 RPB600356S 00643169543379 RPB600357S 00643169543379 RPB600358S 00643169543379 RPB600359S 00643169543379 RPB600360S 00643169543379 RPB600361S 00643169543379 RPB600362S 00643169543379 RPB600363S 00643169543379 RPB600364S 00643169543379 RPB600365S 00643169543379 RPB600366S 00643169543379 RPB600367S 00643169543379 RPB600368S 00643169543379 RPB600369S 00643169543379 RPB600370S 00643169543379 RPB600371S 00643169543379 RPB600372S 00643169543379 RPB600373S 00643169543379 RPB600374S 00643169543379 RPB600375S 00643169543379 RPB600376S 00643169543379 RPB600377S 00643169543379 RPB600378S 00643169543379 RPB600379S 00643169543379 RPB600380S 00643169543379 RPB600381S 00643169543379 RPB600382S 00643169543379 RPB600383S 00643169543379 RPB600384S 00643169543379 RPB600385S 00643169543379 RPB600386S 00643169543379 RPB600387S 00643169543379 RPB600388S 00643169543379 RPB600389S 00643169543379 RPB600390S 00643169543379 RPB600391S 00643169543379 RPB600392S 00643169543379 RPB600393S 00643169543379 RPB600394S 00643169543379 RPB600395S 00643169543379 RPB600396S 00643169543379 RPB600397S 00643169543379 RPB600398S 00643169543379 RPB600399S 00643169543379 RPB600400S 00643169543379 RPB600401S 00643169543379 RPB600402S 00643169543379 RPB600403S 00643169543379 RPB600404S 00643169543379 RPB600405S 00643169543379 RPB600406S 00643169543379 RPB600407S 00643169543379 RPB600408S 00643169543379 RPB600409S 00643169543379 RPB600410S 00643169543379 RPB600411S 00643169543379 RPB600412S 00643169543379 RPB600413S 00643169543379 RPB600414S 00643169543379 RPB600415S 00643169543379 RPB600416S 00643169543379 RPB600417S 00643169543379 RPB600418S 00643169543379 RPB600419S 00643169543379 RPB600420S 00643169543379 RPB600421S 00643169543379 RPB600422S 00643169543379 RPB600423S 00643169543379 RPB600424S 00643169543379 RPB600425S 00643169543379 RPB600426S 00643169543379 RPB600427S 00643169543379 RPB600428S 00643169543379 RPB600429S 00643169543379 RPB600430S 00643169543379 RPB600431S 00643169543379 RPB600432S 00643169543379 RPB600433S 00643169543379 RPB600434S 00643169543379 RPB600435S 00643169543379 RPB600436S 00643169543379 RPB600437S 00643169543379 RPB600438S 00643169543379 RPB600439S 00643169543379 RPB600440S 00643169543379 RPB600441S 00643169543379 RPB600442S 00643169543379 RPB600443S 00643169543379 RPB600444S 00643169543379 RPB600445S 00643169543379 RPB600446S 00643169543379 RPB600447S 00643169543379 RPB600448S 00643169543379 RPB600449S 00643169543379 RPB600450S 00643169543379 RPB600451S 00643169543379 RPB600452S 00643169543379 RPB600453S 00643169543379 RPB600454S 00643169543379 RPB600455S 00643169543379 RPB600456S 00643169543379 RPB600457S 00643169543379 RPB600458S 00643169543379 RPB600459S 00643169543379 RPB600460S 00643169543379 RPB600461S 00643169543379 RPB600462S 00643169543379 RPB600463S 00643169543379 RPB600464S 00643169543379 RPB600465S 00643169543379 RPB600466S 00643169543379 RPB600467S 00643169543379 RPB600468S 00643169543379 RPB600469S 00643169543379 RPB600470S 00643169543379 RPB600471S 00643169543379 RPB600472S 00643169543379 RPB600473S 00643169543379 RPB600474S 00643169543379 RPB600475S 00643169543379 RPB600476S 00643169543379 RPB600477S 00643169543379 RPB600478S 00643169543379 RPB600479S 00643169543379 RPB600480S 00643169543379 RPB600481S 00643169543379 RPB600482S 00643169543379 RPB600483S 00643169543379 RPB600484S 00643169543379 RPB600485S 00643169543379 RPB600486S 00643169543379 RPB600487S 00643169543379 RPB600488S 00643169543379 RPB600489S 00643169543379 RPB600490S 00643169543379 RPB600491S 00643169543379 RPB600492S 00643169543379 RPB600493S 00643169543379 RPB600494S 00643169543379 RPB600495S 00643169543379 RPB600496S 00643169543379 RPB600497S 00643169543379 RPB600498S 00643169543379 RPB600499S 00643169543379 RPB600500S 00643169543379 RPB600501S 00643169543379 RPB600502S 00643169543379 RPB600503S 00643169543379 RPB600504S 00643169543379 RPB600505S 00643169543379 RPB600506S 00643169543379 RPB600507S 00643169543379 RPB600508S 00643169543379 RPB600509S 00643169543379 RPB600510S 00643169543379 RPB600511S 00643169543379 RPB600512S 00643169543379 RPB600513S 00643169543379 RPB600514S 00643169543379 RPB600515S 00643169543379 RPB600516S 00643169543379 RPB600517S 00643169543379 RPB600518S 00643169543379 RPB600519S 00643169543379 RPB600520S 00643169543379 RPB600521S 00643169543379 RPB600522S 00643169543379 RPB600523S 00643169543379 RPB600524S 00643169543379 RPB600525S 00643169543379 RPB600526S 00643169543379 RPB600527S 00643169543379 RPB600528S 00643169543379 RPB600529S 00643169543379 RPB600530S 00643169543379 RPB600531S 00643169543379 RPB600532S 00643169543379 RPB600533S 00643169543379 RPB600534S 00643169543379 RPB600535S 00643169543379 RPB600536S 00643169543379 RPB600537S 00643169543379 RPB600538S 00643169543379 RPB600539S 00643169543379 RPB600540S 00643169543379 RPB600541S 00643169543379 RPB600542S 00643169543379 RPB600543S 00643169543379 RPB600544S 00643169543379 RPB600545S 00643169543379 RPB600546S 00643169543379 RPB600547S 00643169543379 RPB600548S 00643169543379 RPB600549S 00643169543379 RPB600550S 00643169543379 RPB600551S 00643169543379 RPB600552S 00643169543379 RPB600553S 00643169543379 RPB600554S 00643169543379 RPB600555S 00643169543379 RPB600556S 00643169543379 RPB600557S 00643169543379 RPB600558S
Distribution:
Distributed in: US, DC, PR
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0962-2017
Related Recalls
MyCareLink Smart Patient Monitors. It is intended for use with a compatible Medtronic patient implanted heart device. The reader is a portable electronic device that communicates with an implanted heart device
Medtronic Inc., Cardiac Rhythm and Heart Failure
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
MyCareLink Patient Monitors. It is intended for used with a compatible Medtronic patient implanted heart device. The monitor is an external electronic device that interfaces with Medtronic implanted heart devices and a telecommunications connection to transmit stored implanted heart device data to the physician or clinician.
Medtronic Inc., Cardiac Rhythm and Heart Failure
Patients monitored on two (2) or more implanted Medtronic heart devices in the Medtronic CareLink Network may have potential impact on the ability to remotely monitor the patient's heart devices. Potential impacts could lead to missed CareAlert notifications or device reports.
Amplia MRI Quad CRT-0 SureScan Claria MRI Quad CRT-D SureScan Compia MRI Quad CRT-D SureScan
Medtronic Inc., Cardiac Rhythm and Heart Failure
An unexpected transient mode switch behavior was detected by Medtronic during systems testing that affects Amplia' "', Claria'"', and Compia'"' Quad CRT-D MRI devices.