Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indicated for the fixation of fractures, non-unions, arthrodeses, and osteotomies of bones appropriate for the size of the device.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot: TSL003191, TSL003637, TSL004101, TSL004395
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Trilliant Surgical Ltd.
- Reason for Recall:
- Four lots of Tiger Screw Sterilization Tray Base Assemblies were released to the field with a printing error where k-wire description markings indicate diameters of 0.45 and 0.35 instead of 0.045" and 0.035", respectively.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Tiger Screw Sterilization Tray Base Assembly Model 210-00-001 Revision N, O The product is a component of a system indicated for the fixation of fractures, non-unions, arthrodeses, and osteotomies of bones appropriate for the size of the device.
Product Codes/Lot Numbers:
Lot: TSL003191, TSL003637, TSL004101, TSL004395
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0960-2017
Related Recalls
3.5mm Aresenal Screw Drill Bit , Ref: 330-35-001, Qty. 70, Rx Only, Non-Sterile, UDI: (01)00812926029696 - Product Usage: A drill bit facilitates implant insertion
Trilliant Surgical
Labeling Error; specifically, drill bits from lots # TSL008495A are marked with the lot # of TSL008495. This could lead to parts being misidentified and impact device traceability.
8mm HTR Sterile Hammer Toe Reaming Kit, Material: Stainless Steel, Sterile: R, PX only, UDI: (01)00812926022345(17)231225 (10)TSL007202
Trilliant Surgical
The firm has identified that parts from the kit, lot: TSL007202, have the potential to contain the incorrect contents of the 8mm HTR Sterile Hammer Toe Reaming Kit. The kit should contain (2) reamers total, (1) Concave Reamer and (1) Convex Reamer; there is the potential that kits from the identified lot may have (2) of the same reamer type as opposed to one (1) of each type.
Instructions for Use (IFU) and Surgical Cleaning and Sterilization Protocols updated with revised dry time parameters. This recall of IFU occurred in 2018.