Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
Class I - DangerousWhat Should You Do?
- Check if you have this product: Part # UDI 6677200 7325710001349 Serial Numbers 1202-8027 Part # UDI 6677300 7325710001363 Serial Numbers 1058-1201 Part # UDI 6677400 7325710001318 Serial Numbers 2034-8446 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Getinge Usa Sales Inc
- Reason for Recall:
- Due to a software bug, under certain conditions, pressure cannot be built up resulting in no ventilation. If gas delivery is stopped, a sustained decrease in delivered O2 concentration may lead to hypoventilation and hypoxia.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Getinge Flow-i Anesthesia Systems Flow-i C20, Part no. 6677200 Flow-i C30, Part no. 6677300 Flow-i C40, Part no. 6677400
Product Codes/Lot Numbers:
Part # UDI 6677200 7325710001349 Serial Numbers 1202-8027 Part # UDI 6677300 7325710001363 Serial Numbers 1058-1201 Part # UDI 6677400 7325710001318 Serial Numbers 2034-8446 Software versions 4.8.0, 4.8.1, 4.8.2, and 4.8.3.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0959-2023
Related Recalls
Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.
Getinge Usa Sales
Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay