Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003); Multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments. Software utilized in multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments.

Class I - Dangerous
🏥 Medical Devices Recalled: November 20, 2014 Phadia US Diagnostic Equipment Nationwide

What Should You Do?

  1. Check if you have this product:
    Part Number 12-3800-01
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Phadia US Inc
Reason for Recall:
During an investigation of instrument logs it was determined that In specific circumstances involving multiple steps, a rack sequencing error may occur. This will result in a mismatch between the sample ID and the test result reported for all subsequent sample racks in that run.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Phadia 1000 Instrument (introduced as UniCAP 1000), Article number: 12-3800-01 (All instrument software versions since launch in 2003); Multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments. Software utilized in multiple ImmunoCAP assays for Allergen Testing, FDA Cleared for use on Phadia 1000 instrument with IU statement-see attachments.

Product Codes/Lot Numbers:

Part Number 12-3800-01

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0947-2015

Related Recalls

Higher reported results when running EliA GliadinDP IgG Well on the instruments Phadia 2500EE compared to running EliA GliadinDP IgG Well on Phadia 250 instrument

May 1, 2023 Surgical Instruments View Details →

Decreased values for EliA ANA Positive Control when using the EliA dsDNA Well lot BFA37/0142 may potentially cause false positive results (increase in patient test results).

Dec 14, 2020 Infusion Pumps Nationwide View Details →