AF w/AS 550x 0.5 7 ml/hr, 5ml 30 LO CPNB, Product code MVAX30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Class I - DangerousWhat Should You Do?
- Check if you have this product: 74F1700085 74F1703357
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Arrow International Inc
- Reason for Recall:
- Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
AF w/AS 550x 0.5 7 ml/hr, 5ml 30 LO CPNB, Product code MVAX30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump
Product Codes/Lot Numbers:
74F1700085 74F1703357
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0941-2019
Related Recalls
Reports have been received regarding open seals on the packaging.
Arrow FiberOptix Intra-Aortic Balloon Catheter Kit: a) REF IAB-05830-LWS; b) REF IAB-05840-LWS; c) REF IAB-05850-LWS
ARROW INTERNATIONAL
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.
Arrow UltraFlex Intra-Aortic Balloon Catheter Kit, a) REF IAB-06830-U; b) REF IAB-06840-U; c) REF IAB-06850-U
ARROW INTERNATIONAL
Teleflex received reports indicating an infrequent condition that, when not identified and corrected promptly, could result in serious health consequences. The issue may manifest as: -failure of the intra-aortic balloon to completely inflate over its full length -damaged or broken central lumen in the segment contained within the balloon -helium loss or blood in the helium pathway.