Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Catalog Number MS20783 (Infinity MCable-Dual Hemo)
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Draeger Medical, Inc.
- Reason for Recall:
- The displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.
Product Codes/Lot Numbers:
Catalog Number MS20783 (Infinity MCable-Dual Hemo)
Distribution:
Distributed in: CA, IL, MA
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0939-2013
Related Recalls
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.
Brand Name: Atlan Product Name: Atlan A350XL Model/Catalog Number: 8621600 Software Version: Not Applicable Component: Not applicable
Draeger Medical
The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.