Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Elekta, Inc.
- Reason for Recall:
- The DICOM tags do not contain sufficient information for XVI to identify the different volumes that have the same name.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Elekta Synergy XVI. Radiation Therapy Digital Imager. Product XVI R3.5.1, R4.2.1, and R4.5.1 Used as part of radiation therapy treatment process.
Product Codes/Lot Numbers:
151055, 151700, 151809, 151695, 151479, 151696, 151867, 151704, 151765, 151860, 151024, 151521, 151803, 152217, 151774, 152220, 151626, 151842, 151847, 151446, 151801, 151953, 151664, 151751, 151352, 151298, 151728, 151504, 151981, 151705, 151977, 152123, 152126, 151955, 105982, 151160, 151634, 151954, 151416, 151074, 151587, 151729, 151694, 151965, 151418, 151674, 151259, 151698, 151699, 152043, 151636, 151811, 151568, 151258, 151584, 151423, 151623, 151627, 151507, 151640, 151933, 151340, 151722, 151810, 151051, 151411, 151360, 151595, 151433, 151611, 152008, 151151, 151714, 151653, 151675, 151538, 152031, 151523, 151519, 151432, 151783, 151410, 151371, 151084, 151387, 151978, 151530, 151328, 151689, 151567, 151598, 151760, 151269, 151557, 151558, 151789, 151398, 151632, 152214, 151301, 151441, 151256, 151391, 151038, 151585, 151327, 151326, 151250, 151130, 151828, 151245, 151672, 105803, 151073, 151956, 152232, 151319, 151959, 151960, 151958, 151435, 151779, 151969, 151848, 151875, 151881, 151882, 151736, 152023, 152080, 152079, 152064, 151734, 151763, 152154, 151110, 151363, 151216, 151173, 151678, 151467, 151961, 151178, 152091, 152051, 152115, 152116, 151708, 152176, 152192, 152098, 152200, 152174, 152246, 152364, 152673, 152425, 151156, 151764, 151886, 151076, 151369, 152175, 105941
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0931-2013
Related Recalls
Software may fail to load adapted baseline shift plan during treatment delivery when a communication error occurs.
Locking mechanisms may not properly secure to the Arc system, resulting in shifting of coordinates during procedures.
Due to adapter not being properly aligned or locked in place.