MEDRAD Intego PET Infusion System - Pump Intended to deliver accurate doses of Fluoredeoxyglucose or F-Sodium Fluoride and commonly used flushing solutions to patients during molecular imaging (nuclear medicine) diagnostic procedures. It is also intended to provide effective radiation shielding to medical personnel from Fluorine-18 radiation exposure during nuclear medicine diagnostic procedures.

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Device Listing # D053816
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Bayer Healthcare
Reason for Recall:
Bayer has determined that all current Source Administration Sets (SAS) used with the Medrad Intego Pet Infusion System may produce particulates in radiopharmaceutical (RP) vials when the needle is inserted.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

MEDRAD Intego PET Infusion System - Pump Intended to deliver accurate doses of Fluoredeoxyglucose or F-Sodium Fluoride and commonly used flushing solutions to patients during molecular imaging (nuclear medicine) diagnostic procedures. It is also intended to provide effective radiation shielding to medical personnel from Fluorine-18 radiation exposure during nuclear medicine diagnostic procedures.

Product Codes/Lot Numbers:

Device Listing # D053816

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0927-2017

Related Recalls

PET Infusion Pump

Bayer Healthcare

Class I - Dangerous

The Source Administration Sets (SAS) used with the Medrad Intego PET Infusion System do not meet pressure testing specifications in an occlusion situation when the product is more than one year past manufacturing date.

Sep 30, 2016 Infusion Pumps Nationwide View Details →

CertegraWorkstation when used in conjunction with the Medrad Stellant CT Injection System may exhibit thermal events, such as: smoke and sparks of fire.

Aug 2, 2016 Diagnostic Equipment Nationwide View Details →

Bayer Healthcare is initiating this recall due to complaints that were received from customer sites describing a 4205 error message when the injector is used with a 3T scanner.

Jun 9, 2016 Diagnostic Equipment View Details →