i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Class I - DangerousWhat Should You Do?
- Check if you have this product: All units that have been upgraded to software version 2.8
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Abbott Point Of Care Inc.
- Reason for Recall:
- Issues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)
Product Codes/Lot Numbers:
All units that have been upgraded to software version 2.8
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0926-2018
Related Recalls
i-STAT EG6+ cartridge. List Number: 03P77-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.
i-STAT CG4+ cartridge (white). List Number: 03P85-25.
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG4+ cartridge to reflect updated sample type for lactate and measurement ranges for pH and PO2.
i-STAT CG8+ cartridge. List Number: 03P88-25. 510(k) numbers: K894914, K940918, K223710, K230275, K230285, K230300
Abbott Point Of Care
Lack of a 510(k) premarket clearance for the i-STAT CG8+, EG7+ and EG6+ cartridges to reflect updated sample type for potassium and ionized calcium and measurement ranges for pH and PO2.