Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    N/A
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
AGFA Healthcare Corp.
Reason for Recall:
It is possible to exceed the maximum allowed Air Kerma rate (AKR) of 88mGy/min for pulsed fluoro exams. This can happen in case the framerate is increased on the NX workstation when ABS is not enabled.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Brand name: AGFA Digital Radiography X- Ray system DR 800 with MUSICA Dynamic Common Name: DR 800

Product Codes/Lot Numbers:

N/A

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0923-2026

Related Recalls

A customer reported that when using an NX workstation with software version NX 3.0.8950 software and selecting the affected patient/exam from closed exams, initially the wrong image was linked to the exam and appeared. After a short time the wrong image was replaced by the correct image, however the wrong image was used for transmitting to PACS.

Dec 21, 2016 Diagnostic Equipment Nationwide View Details →

A customer experienced when using IMPAX CV Reporting software, specifically, when building a NIV report, the NIV Cardio report was showing incorrect findings for Aneurysm.

Dec 22, 2016 Other Medical Devices Nationwide View Details →