BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: W1-6C000006, W1-71000007, W1-73000009, W1-73000010, XP-68000687T, W1-71000008, W1-85000011
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
β οΈ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Mindray DS USA, Inc. dba Mindray North America
- Reason for Recall:
- The gas spring on the front and rear protective cover may fail. There is a low risk of injury to the user resulting from failure of the gas spring. Injury may include impact to the arms, hands or fingers.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
BA-800M Chemistry Analyzer Part Numbers BA81E-PA00002 and BA81E-PA00003, Containing Gas spring YQ-8/18-90-272-180N (Material Code M6T-010001--- (front spring)) and Gas spring YQ6/15-84-235(B-B)-80N (033-000090-00 (rear spring))
Product Codes/Lot Numbers:
Serial Numbers: W1-6C000006, W1-71000007, W1-73000009, W1-73000010, XP-68000687T, W1-71000008, W1-85000011
Distribution:
Distributed in: US, MI, TX
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0920-2020
Related Recalls
Potential for anesthesia leakage.
Potential for anesthesia leakage.
Potential for activation of an abnormal alarm pause.