Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN: 00845384016410 All lots
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- The Anspach Effort, Inc.
- Reason for Recall:
- Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
Adult Craniotome Ref: CRANI-A-G1 Intended for cutting and shaping bone including the spine and cranium
Product Codes/Lot Numbers:
GTIN: 00845384016410 All lots
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0918-2023
Related Recalls
Pediatric Craniotome Ref:CRANI-P-G1
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Adult Craniotome, Large Ref: CRANI-L-G1 Intended for cutting and shaping bone including the spine and cranium
The Anspach Effort
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury