Lifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 Huber type needle set used for the infusion of medical fluids into implantable ports.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot #1210099
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Churchill Medical Systems, Inc.
- Reason for Recall:
- The label on the device states that the item number is CLGY-2210 when the item number on the device label should read CLGY-2010. This is a 20Ga Huber needle set; however, the product label states the pouch contains a 22Ga set.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Lifeguard Infusion Set with Smartsite, 20Ga, 1, CLGY-2010 Huber type needle set used for the infusion of medical fluids into implantable ports.
Product Codes/Lot Numbers:
Lot #1210099
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0914-2013
Related Recalls
4" Trifurcated set with Bionectors and check valves, Product Code: BN-3082CV and BN-3083CV Accessory device used to administer medical fluids
Churchill Medical Systems
Potential for leaking caused by insufficient bond between needleless device and female luer.
4" Microbore T-Port Set, Product Code: BN-448 Accessory device used to administer medical fluids
Churchill Medical Systems
Potential for leaking caused by insufficient bond between needleless device and female luer.
4" Microbore Trifurcated Ext set, Product Code: BN-308 Accessory device used to administer medical fluids
Churchill Medical Systems
Potential for leaking caused by insufficient bond between needleless device and female luer.