The Siemens Symbia T series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patients physiological (SPECT) and anatomical (CT) images. Software: The syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities.
Class I - DangerousWhat Should You Do?
- Check if you have this product: 1) Symbia T: Product Number: 10275007; Serial Numbers: 1003 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1259 ; 1260 ; 1261 ; 1262 ; 1263 ; 1264 ; 1265 ; 1266 ; 1267 ; 1268 ; 1269 ; 1270 ; 1271 ; 1272 ; 1273 ; 1275 ; 1276 ; 1277 ; 1278 ; 1279 ; 1280 ; 1281 ; 1282 ; 1283 ; 1284 ; 1285 ; 1286 ; 1287 ; 1288 ; 1289 ; 1290 ; 1291 ; 1292 ; 1293 ; 1294 ; 1295 ; 1296 ; 1297 ; 1298 ; 1299 ; 1300 ; 1301 ; 1302 ; 1303 ; 1304 ; 1305 ; 1306 ; 1307 ; 1308 ; 1309 ; 1310 ; 1311 ; 1312 ; 1313 ; 1314 ; 1315 ; 1317 ; 1318 ; 1319 ; 1320 ; 1321 ; 1322 ; 1323 ; 1324 ; 1325 ; 1326 ; 1327 ; 1328 ; 1329 ; 1330 ; 1331 ; 1332 ; 1333 ; 1334 ; 1335 ; 1336 ; 1337 ; 1338 ; 1339 ; 1340 ; 1341 ; 1342 ; 1343 ; 1344 ; 1345 ; 1346 ; 1347 ; 1348 ; 1349 ; 1350 ; 1351 ; 1352 ; 1353 ; 1354 ; 1355 ; 1356 ; 1357 ; 1358 ; 1359 ; 1360 ; 1361 ; 1362 ; 1363 ; 1364 ; 1365 ; 1366 ; 1367 ; 1368 ; 1369 ; 1370 ; 1371 ; 1372 ; 1373 ; 1374 ; 1375 ; 1376 ; 1377 ; 1378 ; 1379 ; 1380 ; 1381 ; 1382 ; 1383 ; 1384 ; 1385 ; 1386 ; 1387 ; 1388 ; 1389 ; 1390 ; 1391 ; 1392 ; 1393 ; 1394 ; 1395 ; 1396 ; 1397 ; 1398 ; 1399 ; 1404 ; 68282 ; 2) Symbia T2; Product Number: 10275008; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1566 ; 3) Symbia T6; Product Number: 10257009; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 4) Symbia T16; Product Number: 10275010; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1259 ; 1260 ; 1261 ; 1262 ; 1263 ; 1264 ; 1265 ; 1266 ; 1267 ; 1271 ; 1272 ; 1274 ; 1644 ; 5) Symbia-T Series (SPECT/CT); Product Number: 8717733; Serial Numbers: 1000 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1208 ; 1209 ; 701001
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Medical Solutions USA, Inc.
- Reason for Recall:
- Siemens Medical Solutions USA, Inc. has received reports of unintended radial motion of the detectors. This unintended motion, should it occur, may cause the patient compression related injuries.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
The Siemens Symbia T series is intended for use by appropriately trained health care professionals to aid in detecting, localizing, diagnosing, staging and restaging of lesions, tumors, disease and organ function for the evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The images produced by the system can also be used by the physician to aid in radiotherapy treatment planning and interventional radiology procedures. SPECT: The SPECT component is intended to detect or image the distribution of radionuclides in the body or organ (physiology), using the following techniques: planar imaging, whole body imaging, and tomographic imaging for isotopes with energies up to 588 keV. CT: The CT component is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data (anatomy) from either the same axial plane taken at different angles or spiral planes taken at different angles. SPECT+CT: The SPECT and CT components used together acquire SPECT/CT images. The SPECT images can be corrected for attenuation with the CT images, and can be combined (image registration) to merge the patients physiological (SPECT) and anatomical (CT) images. Software: The syngo MI Applications software is a display and analysis package intended to aid the clinician in the assessment and quantification of pathologies in images produced from SPECT, PET, CT, and other imaging modalities.
Product Codes/Lot Numbers:
1) Symbia T: Product Number: 10275007; Serial Numbers: 1003 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1259 ; 1260 ; 1261 ; 1262 ; 1263 ; 1264 ; 1265 ; 1266 ; 1267 ; 1268 ; 1269 ; 1270 ; 1271 ; 1272 ; 1273 ; 1275 ; 1276 ; 1277 ; 1278 ; 1279 ; 1280 ; 1281 ; 1282 ; 1283 ; 1284 ; 1285 ; 1286 ; 1287 ; 1288 ; 1289 ; 1290 ; 1291 ; 1292 ; 1293 ; 1294 ; 1295 ; 1296 ; 1297 ; 1298 ; 1299 ; 1300 ; 1301 ; 1302 ; 1303 ; 1304 ; 1305 ; 1306 ; 1307 ; 1308 ; 1309 ; 1310 ; 1311 ; 1312 ; 1313 ; 1314 ; 1315 ; 1317 ; 1318 ; 1319 ; 1320 ; 1321 ; 1322 ; 1323 ; 1324 ; 1325 ; 1326 ; 1327 ; 1328 ; 1329 ; 1330 ; 1331 ; 1332 ; 1333 ; 1334 ; 1335 ; 1336 ; 1337 ; 1338 ; 1339 ; 1340 ; 1341 ; 1342 ; 1343 ; 1344 ; 1345 ; 1346 ; 1347 ; 1348 ; 1349 ; 1350 ; 1351 ; 1352 ; 1353 ; 1354 ; 1355 ; 1356 ; 1357 ; 1358 ; 1359 ; 1360 ; 1361 ; 1362 ; 1363 ; 1364 ; 1365 ; 1366 ; 1367 ; 1368 ; 1369 ; 1370 ; 1371 ; 1372 ; 1373 ; 1374 ; 1375 ; 1376 ; 1377 ; 1378 ; 1379 ; 1380 ; 1381 ; 1382 ; 1383 ; 1384 ; 1385 ; 1386 ; 1387 ; 1388 ; 1389 ; 1390 ; 1391 ; 1392 ; 1393 ; 1394 ; 1395 ; 1396 ; 1397 ; 1398 ; 1399 ; 1404 ; 68282 ; 2) Symbia T2; Product Number: 10275008; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1566 ; 3) Symbia T6; Product Number: 10257009; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 4) Symbia T16; Product Number: 10275010; Serial Numbers: 1001 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1031 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1047 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1069 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1123 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1207 ; 1208 ; 1209 ; 1210 ; 1211 ; 1212 ; 1213 ; 1214 ; 1215 ; 1216 ; 1217 ; 1218 ; 1219 ; 1220 ; 1221 ; 1222 ; 1223 ; 1224 ; 1225 ; 1226 ; 1227 ; 1228 ; 1229 ; 1230 ; 1231 ; 1232 ; 1233 ; 1234 ; 1235 ; 1236 ; 1237 ; 1238 ; 1239 ; 1240 ; 1241 ; 1242 ; 1243 ; 1244 ; 1245 ; 1246 ; 1247 ; 1248 ; 1249 ; 1250 ; 1251 ; 1252 ; 1253 ; 1254 ; 1255 ; 1256 ; 1257 ; 1258 ; 1259 ; 1260 ; 1261 ; 1262 ; 1263 ; 1264 ; 1265 ; 1266 ; 1267 ; 1271 ; 1272 ; 1274 ; 1644 ; 5) Symbia-T Series (SPECT/CT); Product Number: 8717733; Serial Numbers: 1000 ; 1002 ; 1003 ; 1004 ; 1005 ; 1006 ; 1007 ; 1008 ; 1009 ; 1010 ; 1011 ; 1012 ; 1013 ; 1014 ; 1015 ; 1016 ; 1017 ; 1018 ; 1019 ; 1020 ; 1021 ; 1022 ; 1023 ; 1024 ; 1025 ; 1026 ; 1027 ; 1028 ; 1029 ; 1030 ; 1032 ; 1033 ; 1034 ; 1035 ; 1036 ; 1037 ; 1038 ; 1039 ; 1040 ; 1041 ; 1042 ; 1043 ; 1044 ; 1045 ; 1046 ; 1048 ; 1049 ; 1050 ; 1051 ; 1052 ; 1053 ; 1054 ; 1055 ; 1056 ; 1057 ; 1058 ; 1059 ; 1060 ; 1061 ; 1062 ; 1063 ; 1064 ; 1065 ; 1066 ; 1067 ; 1068 ; 1070 ; 1071 ; 1072 ; 1073 ; 1074 ; 1075 ; 1076 ; 1077 ; 1078 ; 1079 ; 1080 ; 1081 ; 1082 ; 1083 ; 1084 ; 1085 ; 1086 ; 1087 ; 1088 ; 1089 ; 1090 ; 1091 ; 1092 ; 1093 ; 1094 ; 1095 ; 1096 ; 1097 ; 1098 ; 1099 ; 1100 ; 1101 ; 1102 ; 1103 ; 1104 ; 1105 ; 1106 ; 1107 ; 1108 ; 1109 ; 1110 ; 1111 ; 1112 ; 1113 ; 1114 ; 1115 ; 1116 ; 1117 ; 1118 ; 1119 ; 1120 ; 1121 ; 1122 ; 1124 ; 1125 ; 1126 ; 1127 ; 1128 ; 1129 ; 1130 ; 1131 ; 1132 ; 1133 ; 1134 ; 1135 ; 1136 ; 1137 ; 1138 ; 1139 ; 1140 ; 1141 ; 1142 ; 1143 ; 1144 ; 1145 ; 1146 ; 1147 ; 1148 ; 1149 ; 1150 ; 1151 ; 1152 ; 1153 ; 1154 ; 1155 ; 1156 ; 1157 ; 1158 ; 1159 ; 1160 ; 1161 ; 1162 ; 1163 ; 1164 ; 1165 ; 1166 ; 1167 ; 1168 ; 1169 ; 1170 ; 1171 ; 1172 ; 1173 ; 1174 ; 1175 ; 1176 ; 1177 ; 1178 ; 1179 ; 1180 ; 1181 ; 1182 ; 1183 ; 1184 ; 1185 ; 1186 ; 1187 ; 1188 ; 1189 ; 1190 ; 1191 ; 1192 ; 1193 ; 1194 ; 1195 ; 1196 ; 1197 ; 1198 ; 1199 ; 1200 ; 1201 ; 1202 ; 1203 ; 1204 ; 1205 ; 1206 ; 1208 ; 1209 ; 701001
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0912-2015
Related Recalls
MAMMOMAT Inspiration;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Fusion;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.
MAMMOMAT Revelation;
Siemens Medical Solutions USA
There were instances where the operator table was sold together with a bus-installation kit. The operator table is not intended and not released for bus installations as it was designed for stationary operation only.