Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN: 04546540400734 Lot Numbers: 42874701, 42912101, 42995601
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Howmedica Osteonics Corp.
Reason for Recall:
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

Scorpio NRG Tibial Brg Insert Assy -Intended for knee replacement Part Number: 82-2-0908

Product Codes/Lot Numbers:

GTIN: 04546540400734 Lot Numbers: 42874701, 42912101, 42995601

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0899-2024

Related Recalls

Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.