GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Reason for Recall:
GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

GE Healthcare Discovery NM/CT 670 CZT, Model/Catalog Numbers: 1) H3906AB; System, Tomography, Computed, Emission

Product Codes/Lot Numbers:

1) H3906AB: UDI/DI 840682121446, System ID UX2305NU03, Serial Number CZTV29005; UDI/DI 840682121446, System ID M4166103, Serial Number CZTV29006; UDI/DI 840682121446, System ID 00243NUC22, Serial Number CZTW29008; UDI/DI 840682121446, System ID 00222NUC07, Serial Number CZTW29009; UDI/DI N/A, System ID 604875D670B, Serial Number CZTW29007; UDI/DI N/A, System ID NT6200, Serial Number AD7W38007; UDI/DI N/A, System ID NT6201, Serial Number AD7X38017; UDI/DI N/A, System ID NT6100, Serial Number AP7W37001.

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0895-2025

Related Recalls

Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →
Class I - Dangerous

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Sep 12, 2025 Diagnostic Equipment View Details →