SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
Class I - DangerousWhat Should You Do?
- Check if you have this product: GTIN: 07613327033885 Lot Numbers: EPXH, DYAE, XVK2, PVP9, LJ0K, 12R3, N8P6, T355, 8HDN, A3EV, 5LKW, R04Y, RKT1,D06K, 01JR, 3PA4, 0AYL, W424, 53T1, Y7X2,
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Howmedica Osteonics Corp.
- Reason for Recall:
- Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
SCORPIO U-DOME PATELLA -Intended for knee replacement Part Number: 73-3508
Product Codes/Lot Numbers:
GTIN: 07613327033885 Lot Numbers: EPXH, DYAE, XVK2, PVP9, LJ0K, 12R3, N8P6, T355, 8HDN, A3EV, 5LKW, R04Y, RKT1,D06K, 01JR, 3PA4, 0AYL, W424, 53T1, Y7X2,
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0895-2024
Related Recalls
TRI TS BASEPLATE SIZE 3. Part Number: 5521-B-300.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
TRI TS BASEPLATE SIZE 4. Part Number: 5521-B-400.
Howmedica Osteonics
Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
Howmedica Osteonics
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.