Olympus, Model No. GIF-Q180,flexible scope, large diameter, video gastroscope
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial Numbers: 2501321,2603572,2603751,2707414,2500011,2705546,2604232,2603079,2500780,2808458, 2001720,2002054,2001724,2001723,2501905,2809593,2705417,2500916,2808421,2809413, 2705911,2203846,2808998,2501295,2809000,2705264,2603152,2809444,2706645,2102694, 2203503,2809061,2705453,2706361,2102880,2002142,2500236,2705566,2705242,2500869, 2501511,2603466,2706515,2706908,2706905,2706442,2001193,2501087,2501807,2705697, 2705333,2900664,2501186,2800014,2707065,2705978,2501786,2002139,2002083,2809935, 2809951,2707769,2304352,2707593,2304332,2001807,2808518,2602866,2602846,2706687, 2001864,2808509,2705957,2706072,2705888,2602226,2102504,2501901,2501941,2900354, 2501928,2203680,2500463,2203472,2707126,2603744,2707886,2901047,2603180,2604176, 2707760,2502049,2602102,2706079,2706071,2603287,2602732,2602730,2900859,2705386, 2808456,2705662,2809750,2705605,2705589,2808232,2706404,2500529,2602714,2501738, 2706823,2809590,2604997,2604999,2604989,2808704,2808713,2604264,2706028,2800064, 2808452,2500660,2501776,2808455,2501961,2501854,2809540,2501472,2602442,2603316, 2808333,2707097,2706919,2707574,2501001,2706156,2605035,2501890,2603001,2603053, 2706604,2707840,2501962,2605156,2808499,2304527,2603134,2604932,2102468,2500067, 2501704,2501595,2603135,2705891,2807973,2705461,2603530,2707882,2706068,2807964, 2706013,2705465,2707101,2705475,2707892,2500032,2602610,2501680,2707346,2707374, 2707382,2500458,2500653,2808341,2808894,2603444,2001317,2602502,2604160,2500576, 2706920,2808353,2705694,2705593,2900826,2900764,2604623,2901059
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Integrated Medical Systems Inc
- Reason for Recall:
- If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Olympus, Model No. GIF-Q180,flexible scope, large diameter, video gastroscope
Product Codes/Lot Numbers:
Serial Numbers: 2501321,2603572,2603751,2707414,2500011,2705546,2604232,2603079,2500780,2808458, 2001720,2002054,2001724,2001723,2501905,2809593,2705417,2500916,2808421,2809413, 2705911,2203846,2808998,2501295,2809000,2705264,2603152,2809444,2706645,2102694, 2203503,2809061,2705453,2706361,2102880,2002142,2500236,2705566,2705242,2500869, 2501511,2603466,2706515,2706908,2706905,2706442,2001193,2501087,2501807,2705697, 2705333,2900664,2501186,2800014,2707065,2705978,2501786,2002139,2002083,2809935, 2809951,2707769,2304352,2707593,2304332,2001807,2808518,2602866,2602846,2706687, 2001864,2808509,2705957,2706072,2705888,2602226,2102504,2501901,2501941,2900354, 2501928,2203680,2500463,2203472,2707126,2603744,2707886,2901047,2603180,2604176, 2707760,2502049,2602102,2706079,2706071,2603287,2602732,2602730,2900859,2705386, 2808456,2705662,2809750,2705605,2705589,2808232,2706404,2500529,2602714,2501738, 2706823,2809590,2604997,2604999,2604989,2808704,2808713,2604264,2706028,2800064, 2808452,2500660,2501776,2808455,2501961,2501854,2809540,2501472,2602442,2603316, 2808333,2707097,2706919,2707574,2501001,2706156,2605035,2501890,2603001,2603053, 2706604,2707840,2501962,2605156,2808499,2304527,2603134,2604932,2102468,2500067, 2501704,2501595,2603135,2705891,2807973,2705461,2603530,2707882,2706068,2807964, 2706013,2705465,2707101,2705475,2707892,2500032,2602610,2501680,2707346,2707374, 2707382,2500458,2500653,2808341,2808894,2603444,2001317,2602502,2604160,2500576, 2706920,2808353,2705694,2705593,2900826,2900764,2604623,2901059
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0894-2016
Related Recalls
Olympus, Model No. GIF-H180J, flexible scope, large diameter, video gastroscope
Integrated Medical Systems
If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
Olympus, Model No. CF-Q140L, flexible scope, large diameter, video colonoscope
Integrated Medical Systems
If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.
Olympus, Model No. GIF-Q140, flexible scope, large diameter, video gastroscope
Integrated Medical Systems
If excessive pressure is applied to the distal end of the endoscope, the internal coil pipe assembly at the distal end of the endoscope can become detached and angulation may not operate as intended.