Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    Lot Number: 15STAC072 Exp Date: 2017-12-31
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Fresenius Medical Care Renal Therapies Group, LLC
Reason for Recall:
On the label calcium (Ca) listed in the right upper corner of the label (red background) incorrectly indicates the product contains 2.5 mEq/L Ca. The concentration indicated in the list of constituents represents the actual calcium concentration of the final dialysate, 3.0 mEq/L.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
terminated

Product Information

Full Description:

Fresenius Naturalyte Liquid Acid Concentrate Catalog Number: 08-2301-3 Intended Use: Acid Concentrate for Bicarbonate Dialysis

Product Codes/Lot Numbers:

Lot Number: 15STAC072 Exp Date: 2017-12-31

Distribution:

Distributed in: US, KS, LA, MS, OK, TX

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0887-2016

Related Recalls

Additional Instructions: step-by step handling and inserting the cassette into the Liberty Cycler set. A damaged cassette can cause peritoneal dialysis (PD) solution to leak into the cycler and is a potential risk for contamination.

May 29, 2019 Other Medical Devices Nationwide View Details →