IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number (SMN) 10706279 (US); Catalog Number LKPS1, SMN 10380960 and Catalog Number LKPS5, SMN 10380949 (OUS); IVD --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL554EL UK For in vitro diagnostic use with the IMMULITE /IMMULITE 1000/IMMULITE 2000/2000 XPi Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Class I - DangerousWhat Should You Do?
- Check if you have this product: All lots LKPS1/LKPS5 - Lots 443, 445, 446 LKPTS1 - Lot D116
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Siemens Healthcare Diagnostics, Inc.
- Reason for Recall:
- Siemens Healthcare Diagnostics has confirmed that the IMMULITE/ IMMULITE 1000 Systems PSA assays are not meeting the current High-dose Hook Effect expectation in the Instructions for Use.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
IMMULITE/IMMULITE 1000 Systems Prostate-Specific Antigen (PSA) Assay, REF/Catalog Number LKPTS1, Siemens Material Number (SMN) 10706279 (US); Catalog Number LKPS1, SMN 10380960 and Catalog Number LKPS5, SMN 10380949 (OUS); IVD --- Origin: UK Siemens Healthcare Diagnostics Products Ltd. Llanberis, Gwynedd, LL554EL UK For in vitro diagnostic use with the IMMULITE /IMMULITE 1000/IMMULITE 2000/2000 XPi Analyzers for the quantitative measurement of prostate-specific antigen (PSA) in human serum, as an aid in the detection of prostate cancer when used in conjunction with digital rectal examination (DRE) in men aged 50 years or older. This assay is further indicated as an adjunctive test to aid in the management of prostate cancer patients.
Product Codes/Lot Numbers:
All lots LKPS1/LKPS5 - Lots 443, 445, 446 LKPTS1 - Lot D116
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0884-2017
Related Recalls
Food Panel 3 Specific Allergen; Siemens Material Number (SMN): 10385891; Catalog Number: FP3L4;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
IMMULITE 2000 Oak Mix Specific Allergen; Siemens Material Number (SMN): 10386105; Catalog Number: L2KUN6;
Siemens Healthcare Diagnostics
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.