SCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512

Class I - Dangerous

What Should You Do?

  1. Check if you have this product:
    GTIN: 07613327034790 Lot Numbers: 5M6LM1
  2. Do not eat it: Even if it looks and smells fine, do not consume this product.
  3. Throw it away or return it: You can return the product to the store for a full refund.
  4. Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
  5. Report problems: Report any issues to the FDA's Safety Reporting Portal.

⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.

Recall Details

Company:
Howmedica Osteonics Corp.
Reason for Recall:
Product manufactured using UHMWPE raw material over 5 years of age has the potential for elevated levels of oxidation. Oxidation within UHMWPE can have an impact on its material properties.
Classification:
Class I - Dangerous

Dangerous or defective products that predictably could cause serious health problems or death.

Status:
ongoing

Product Information

Full Description:

SCORPIO TS TIB INSERT-Intended for knee replacement Part Number:72-4-7512

Product Codes/Lot Numbers:

GTIN: 07613327034790 Lot Numbers: 5M6LM1

Official Source

Always verify recall information with the official FDA source:

View on FDA.gov

FDA Recall Number: Z-0877-2024

Related Recalls

Class I - Dangerous

Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).

May 19, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

Stryker has identified that Triathlon Universal Baseplate Size 3 (lot UZD9IB), is contained in packaging labeled as Triathlon Universal Baseplate Size 4 (lot TYX7OB), and Triathlon Universal Baseplate Size 4 (lot TYX7OB), is contained in packaging labeled as Triathlon Universal Baseplate Size 3, (lot UZD9IB).

May 19, 2026 Other Medical Devices Nationwide View Details →
Class I - Dangerous

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Feb 2, 2026 Other Medical Devices Nationwide View Details →