8470-009-040 4.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Lot Number:16038017, Product Number: 8470-009-040
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Stryker Instruments Div. of Stryker Corporation
- Reason for Recall:
- Stryker Instruments is voluntarily recalling specific lots of sterile MIS Toolsteel Burs because the burs may exhibit signs of corrosion.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
8470-009-040 4.0MM Precision Round, 16CM, single use, Sterile The Stryker MIS (Minimally Invasive Surgery) product offering, is used to remove osseous material during surgical cases. Each cutting accessory consists of: " A shank (B) with a notch (A), that fits in the attachment on one end and is connected to the head of the bur on the other end. " A head (C) connected to the shank (B) and removes material using machined flutes. The head is machined directly on the shank/neck of the bur.
Product Codes/Lot Numbers:
Lot Number:16038017, Product Number: 8470-009-040
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0871-2017
Related Recalls
SurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting, displaying, recording and locating the number of RFID-tagged absorbent articles used during surgical procedures.
Stryker Instruments Div. of Stryker
When scanning sponges out after a surgical procedure, an error may allow for a duplicate sponge to be counted without the prompt to make a case note and allowing the case to close. The software update will prevent the occurrence of this error.
Triton Canister System (finished part numbers FG 12009, for 3L canisters, and FG 12019 for 2L canisters) Scanning Label (component part number 11032). A software application used to help estimation of blood loss. Used with blood bearing canisters and accessories.
Stryker Instruments Div. of Stryker
Some Canister Scanning Labels used to calibrate the Triton Canister System have a color variance. Based on a comprehensive analysis, blood loss estimates obtained using these Canister Scanning Labels may result in lower than expected results.
Triton Sponge System (Software), GAUS-3, Software Versions 4.6.5, 4.8.7, 4.9.6
Stryker Instruments Div. of Stryker
Gauss Surgical is updating the Instructions for Use (IFU) for our Triton Sponge System in order to clarify proper technique for handling sponges and the calibration placard.