MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Class I - DangerousWhat Should You Do?
- Check if you have this product: Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Shimadzu Medical Systems
- Reason for Recall:
- It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- ongoing
Product Information
Full Description:
MODEL: X-RAY TV SYSSTEM SONIALVISION safire17
Product Codes/Lot Numbers:
Model: SonialVision safire: Generator Number: D150-40 Serial Numbers: 3M5249D41006 3M5249D2A004 3M5249D34001 3M5249D37006 3M5249D32012 3M5249D35002 3M5249D1C013 3M5249D31009 3M5249D29002 3M5249D33003 3M5249D2C004 3M5249D22013
Distribution:
Nationwide distribution
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0869-2022
Related Recalls
MODEL: X-RAY TV SYSTEM SONIALVISION G4
Shimadzu Medical Systems
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
MODEL: X-RAY R/F SYSTEM FLUOROspeed X1
Shimadzu Medical Systems
It was found that the irradiated x-ray may exceed the xray radiation dose rate specified in CFR 1020.32 in some specific case of fluoroscopic mode due to inadequate adjusting criteria in installation.
Digital Angiography System Bransist safire Generator Model D150GC-40
Shimadzu Medical Systems
Due to the inadequate adjusting criteria in installation, the system generator on two Digital Angiography Systems, the irradiated x-ray may exceed the x-ray radiation dose rate in fluoroscopic mode.