Cyberonics VNS Therapy AspireSR, Model 106 Generator, 26-0007-8400/1. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
Class I - DangerousWhat Should You Do?
- Check if you have this product: Serial numbers 36519, 36766, 36770, 36776, 36791, 36794, 36800, 36802, 36821, 36824, 36833, 36835, 36845, 36847, 36849, 36869, 36873, 36932, 36935, 36940, 36947, 36951, 36954, 36959, 36961, 36962, 36963, 36964, 36965, 36972, 36981, 36998, 37003, 37014, 37015, 37016, 37024, 37025, 37052, 37095, 37110, 37111, 37116, 37123, 37149, 37151, 37156, 37181, 37213, 37231, 37234, 37239, 37248, 37250, 37253, 37258, 37263, 37270, 37294, 37296, 37297, 37301, 37302, 37310, 37328, 37329, 37331, 37332, 37334, 37338, 37339, 37351, 37353, 37355, 37359, 37360, 37362, 37365, 37371, 37376, 37378, 37383, 37389, 37391, 37395, 37396, 37397, 37402, 37403, 37405, 37409, 37410, 37411, 37414, 37417, 37420, 37421, 37422, 37424, 37425, 37431, 37435, 37439, 37445, 37446, 37449, 37454, 37455, 37458, 37461, 37464, 37466, 37471, 37472, 37480, 37482, 37487, 37488, 37489, 37490, 37498, 37501, 37502, 37505, 37507, 37508, 37510, 37512, 37530, 37540, 37548, 37551, 37559, 37561, 37563, 37564, 37566, 37567, 37572, 37574, 37576, 37577, 37579, 37585, 37587, 37591, 37601, 37607, 37611, 37615, 37634, 37639, 37649, 37662, 37663, 37671, 37672, 37676, 37677, 37678, 37682, 37683, 37689, 37697, 37706, 37708, 37710, 37720, 37723, 37724, 37727, 37730, 37733, 37736, 37738, 37908, 37909, 37920, 37925, 37941, 38091, 38105, 38126, 38212, 38215, 38437, 38446, 38494, 38614, 38657, 38671, 38990, 39025, 39032, 39057, 39068, 39069, 39079, 39261, 39263, 39286, 39340, 39341, 39390, 39412, 39415, 39420, 39421, 39429, 39455, 39463, 39466, 39467, 39470, 39473, 39475, 39477, 39486, 39490, 39499, 39505, 39511, 39532, 39636, 39664, 39906, 39911, 39953, 39970, 40007, 40017, 40036, 40081, 40104, 40116, 40145, 40149, 40159, 40248, 40270, 40280, 40306, 40314, 40320, 40340, 40345, 40346, 40347, 40353, 40355, 40401, 40404, 40426, 40454, 40481, 40482, 40486, 40491, 40492, 40494, 40499, 40503, 40513, 40516, 40517, 40521, 40522, 40528, 40532, 40533, 40534, 40537, 40538, 40540, 40542, 40544, 40551, 40557, 40563, 40565, 40569, 40570, 40572, 40573, 40574, 40582, 40585, 40590, 40591, 40595, 40611, 40619, 40627, 40633, 40640, 40643, 40664, 40676, 40684, 40692, 40694, 40696, 40708, 40718, 40723, 40731, 40736, 40737, 40741, 40750, 40753, 40755, 40757, 40762, 40767, 40769, 40770, 40772, 40774, 40777, 40778, 40780, 40781, 40783, 40784, 40786, 40788, 40789, 40790, 40792, 40793, 40794, 40795, 40801, 40802, 40804, 40808, 40809, 40822, 40823, 40824, 40830, 40834, 40838, 40839, 40840, 40844, 40845, 40846, 40851, 40853, 40854, 40856, 40857, 40861, 40870, 40871, 40873, 40874, 40879, 40881, 40883, 40888, 40894, 40895, 40912, 40919, 40921, 40927, 40934, 40935, 40942, 40943, 40971, 40983, 40993, 41010, 41032, 41039, 41044, 41048, 41051, 41059, 41060, 41065, 41071, 41075, 41078, 41079, 41082, 41083, 41084, 41088, 41090, 41104, 41108, 41113, 41114, 41118, 41120, 41124, 41130, 41133, 41137, 41146, 41148, 41149, 41150, 41157, 41159, 41160, 41169, 41172, 41173, 41178, 41180, 41182, 41195, 41196, 41199, 41200, 41202, 41207, 41209, 41229, 41240, 41246, 41249, 41250, 41253, 41254, 41255, 41264, 41266, 41271, 41283, 41290, 41299, 41301, 41320, 41321, 41322, 41323, 41324, 41326, 41327, 41328, 41347, 41348, 41351, 41354, 41357, 41359, 41368, 41370, 41374, 41380, 41390, 41394, 41413, 41422, 41468, 41470, 41472, 41483, 41487, 41488, 41496, 41506, 41508, 41517, 41519, 41524, 41540, 41547, 41552, 41555, 41559, 41569, 41577, 41589, 41594, 41598, 41601, 41609, 41612, 41617, 41626, 41627, 41629, 41637, 41640, 41645, 41654, 41656, 41661, 41663, 41665, 41671, 41673, 41692, 41696, 41700, 41706, 41740, 41749, 41753, 41756, 41759, 41768, 41773, 41776, 41779, 41791, 41823, 41832, 41833, 41840, 41852, 41872, 41874, 41875, 41879, 41881, 41885, 41905, 41907, 41911, 41926, 41929, 41931, 41932, 41933, 41936, 41943, 41944, 41946, 41953, 41957, 41973, 41979, 42005, 42021, 42036, 42043, 42046, 42051, 42069, 42133, 42134, 42144, 42150, 42154, 42155, 42160, 42177, 42181, 42182, 42186, 42199, 42202, 42219, 42220, 42223, 42224, 42225, 42233, 42234, 42242, 42247, 42255, 42259, 42267, 42268, 42269, 42274, 42292, 42316, 42326, 42336, 42347, 42353, 42354, 42356, 42361, 42387, 42391, 42392, 42405, 42415, 42417, 42428, 42466, 42504, 42505, 42514, 42523, 42528, 42529, 42537, 42539, 42565, 42567, 42568, 42571, 42572, 42579, 42587, 42588, 42592, 42593, 42594, 42595, 42599, 42600, 42601, 42605, 42606, 42608, 42610, 42611, 42613, 42615, 42619, 42624, 42625, 42627, 42629, 42631, 42632, 42633, 42634, 42635, 42636, 42641, 42642, 42643, 42644, 42645, 42647, 42654, 42657, 42658, 42659, 42664, 42666, 42667, 42681, 42685, 42689, 42696, 42697, 42698, 42702, 42703, 42706, 42711, 42713, 42718, 42719, 42720, 42721, 42722, 42725, 42727, 42733, 42735, 42738, 42739, 42746, 42747, 42750, 42754, 42756, 42758, 42761, 42763, 42767, 42773, 42776, 42778, 42781, 42784, 42785, 42787, 42788, 42789, 42811, 42820, 42834, 42835, 42839, 42840, 42844, 42847, 42861, 42862, 42865, 42866, 42868, 42869, 42871, 42873, 42876, 42877, 42880, 42881, 42887, 42888, 42890, 42892, 42893, 42895, 42900, 42902, 42920, 42934, 42945, 42962, 43003, 43012, 43025, 43031, 43032, 43038, 43095, 43102, 43103, 43107, 43109, 43115, 43118, 43129, 43136, 43140, 43144, 43151, 43157, 43165, 43167, 43169, 43172, 43177, 43178, and 43201. The expiration date range is 9/30/2016-2/28/2017.
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- LivaNova USA Inc
- Reason for Recall:
- This recall is an expansion of Z-3019-2017 and Z-3020-2017, which was initiated to fix the devices premature battery depletion, caused by electrical leakage on the circuit board assemblies of the Models 105 and 106 generators. *Note this recall occurred in November 2018.
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Cyberonics VNS Therapy AspireSR, Model 106 Generator, 26-0007-8400/1. The firm name on the label is CYBERONICS, INC., Houston, TX. Vagus nerve stimulator.
Product Codes/Lot Numbers:
Serial numbers 36519, 36766, 36770, 36776, 36791, 36794, 36800, 36802, 36821, 36824, 36833, 36835, 36845, 36847, 36849, 36869, 36873, 36932, 36935, 36940, 36947, 36951, 36954, 36959, 36961, 36962, 36963, 36964, 36965, 36972, 36981, 36998, 37003, 37014, 37015, 37016, 37024, 37025, 37052, 37095, 37110, 37111, 37116, 37123, 37149, 37151, 37156, 37181, 37213, 37231, 37234, 37239, 37248, 37250, 37253, 37258, 37263, 37270, 37294, 37296, 37297, 37301, 37302, 37310, 37328, 37329, 37331, 37332, 37334, 37338, 37339, 37351, 37353, 37355, 37359, 37360, 37362, 37365, 37371, 37376, 37378, 37383, 37389, 37391, 37395, 37396, 37397, 37402, 37403, 37405, 37409, 37410, 37411, 37414, 37417, 37420, 37421, 37422, 37424, 37425, 37431, 37435, 37439, 37445, 37446, 37449, 37454, 37455, 37458, 37461, 37464, 37466, 37471, 37472, 37480, 37482, 37487, 37488, 37489, 37490, 37498, 37501, 37502, 37505, 37507, 37508, 37510, 37512, 37530, 37540, 37548, 37551, 37559, 37561, 37563, 37564, 37566, 37567, 37572, 37574, 37576, 37577, 37579, 37585, 37587, 37591, 37601, 37607, 37611, 37615, 37634, 37639, 37649, 37662, 37663, 37671, 37672, 37676, 37677, 37678, 37682, 37683, 37689, 37697, 37706, 37708, 37710, 37720, 37723, 37724, 37727, 37730, 37733, 37736, 37738, 37908, 37909, 37920, 37925, 37941, 38091, 38105, 38126, 38212, 38215, 38437, 38446, 38494, 38614, 38657, 38671, 38990, 39025, 39032, 39057, 39068, 39069, 39079, 39261, 39263, 39286, 39340, 39341, 39390, 39412, 39415, 39420, 39421, 39429, 39455, 39463, 39466, 39467, 39470, 39473, 39475, 39477, 39486, 39490, 39499, 39505, 39511, 39532, 39636, 39664, 39906, 39911, 39953, 39970, 40007, 40017, 40036, 40081, 40104, 40116, 40145, 40149, 40159, 40248, 40270, 40280, 40306, 40314, 40320, 40340, 40345, 40346, 40347, 40353, 40355, 40401, 40404, 40426, 40454, 40481, 40482, 40486, 40491, 40492, 40494, 40499, 40503, 40513, 40516, 40517, 40521, 40522, 40528, 40532, 40533, 40534, 40537, 40538, 40540, 40542, 40544, 40551, 40557, 40563, 40565, 40569, 40570, 40572, 40573, 40574, 40582, 40585, 40590, 40591, 40595, 40611, 40619, 40627, 40633, 40640, 40643, 40664, 40676, 40684, 40692, 40694, 40696, 40708, 40718, 40723, 40731, 40736, 40737, 40741, 40750, 40753, 40755, 40757, 40762, 40767, 40769, 40770, 40772, 40774, 40777, 40778, 40780, 40781, 40783, 40784, 40786, 40788, 40789, 40790, 40792, 40793, 40794, 40795, 40801, 40802, 40804, 40808, 40809, 40822, 40823, 40824, 40830, 40834, 40838, 40839, 40840, 40844, 40845, 40846, 40851, 40853, 40854, 40856, 40857, 40861, 40870, 40871, 40873, 40874, 40879, 40881, 40883, 40888, 40894, 40895, 40912, 40919, 40921, 40927, 40934, 40935, 40942, 40943, 40971, 40983, 40993, 41010, 41032, 41039, 41044, 41048, 41051, 41059, 41060, 41065, 41071, 41075, 41078, 41079, 41082, 41083, 41084, 41088, 41090, 41104, 41108, 41113, 41114, 41118, 41120, 41124, 41130, 41133, 41137, 41146, 41148, 41149, 41150, 41157, 41159, 41160, 41169, 41172, 41173, 41178, 41180, 41182, 41195, 41196, 41199, 41200, 41202, 41207, 41209, 41229, 41240, 41246, 41249, 41250, 41253, 41254, 41255, 41264, 41266, 41271, 41283, 41290, 41299, 41301, 41320, 41321, 41322, 41323, 41324, 41326, 41327, 41328, 41347, 41348, 41351, 41354, 41357, 41359, 41368, 41370, 41374, 41380, 41390, 41394, 41413, 41422, 41468, 41470, 41472, 41483, 41487, 41488, 41496, 41506, 41508, 41517, 41519, 41524, 41540, 41547, 41552, 41555, 41559, 41569, 41577, 41589, 41594, 41598, 41601, 41609, 41612, 41617, 41626, 41627, 41629, 41637, 41640, 41645, 41654, 41656, 41661, 41663, 41665, 41671, 41673, 41692, 41696, 41700, 41706, 41740, 41749, 41753, 41756, 41759, 41768, 41773, 41776, 41779, 41791, 41823, 41832, 41833, 41840, 41852, 41872, 41874, 41875, 41879, 41881, 41885, 41905, 41907, 41911, 41926, 41929, 41931, 41932, 41933, 41936, 41943, 41944, 41946, 41953, 41957, 41973, 41979, 42005, 42021, 42036, 42043, 42046, 42051, 42069, 42133, 42134, 42144, 42150, 42154, 42155, 42160, 42177, 42181, 42182, 42186, 42199, 42202, 42219, 42220, 42223, 42224, 42225, 42233, 42234, 42242, 42247, 42255, 42259, 42267, 42268, 42269, 42274, 42292, 42316, 42326, 42336, 42347, 42353, 42354, 42356, 42361, 42387, 42391, 42392, 42405, 42415, 42417, 42428, 42466, 42504, 42505, 42514, 42523, 42528, 42529, 42537, 42539, 42565, 42567, 42568, 42571, 42572, 42579, 42587, 42588, 42592, 42593, 42594, 42595, 42599, 42600, 42601, 42605, 42606, 42608, 42610, 42611, 42613, 42615, 42619, 42624, 42625, 42627, 42629, 42631, 42632, 42633, 42634, 42635, 42636, 42641, 42642, 42643, 42644, 42645, 42647, 42654, 42657, 42658, 42659, 42664, 42666, 42667, 42681, 42685, 42689, 42696, 42697, 42698, 42702, 42703, 42706, 42711, 42713, 42718, 42719, 42720, 42721, 42722, 42725, 42727, 42733, 42735, 42738, 42739, 42746, 42747, 42750, 42754, 42756, 42758, 42761, 42763, 42767, 42773, 42776, 42778, 42781, 42784, 42785, 42787, 42788, 42789, 42811, 42820, 42834, 42835, 42839, 42840, 42844, 42847, 42861, 42862, 42865, 42866, 42868, 42869, 42871, 42873, 42876, 42877, 42880, 42881, 42887, 42888, 42890, 42892, 42893, 42895, 42900, 42902, 42920, 42934, 42945, 42962, 43003, 43012, 43025, 43031, 43032, 43038, 43095, 43102, 43103, 43107, 43109, 43115, 43118, 43129, 43136, 43140, 43144, 43151, 43157, 43165, 43167, 43169, 43172, 43177, 43178, and 43201. The expiration date range is 9/30/2016-2/28/2017.
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0869-2020
Related Recalls
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
Their is a potential that implanted pulse generators may stop delivering therapy due to an internal component issue.