Philips IntelliVue Module Measurement X2 Model: M3002A
Class I - DangerousWhat Should You Do?
- Check if you have this product: Software Revision: K.21.54 or L.00.96
- Do not eat it: Even if it looks and smells fine, do not consume this product.
- Throw it away or return it: You can return the product to the store for a full refund.
- Seek medical attention if needed: If you've consumed this product and feel unwell, contact your doctor immediately.
- Report problems: Report any issues to the FDA's Safety Reporting Portal.
⚠️ Emergency: If you experience severe symptoms after consuming this product, call 911 or Poison Control at 1-800-222-1222.
Recall Details
- Company:
- Philips Electronics North America Corporation
- Reason for Recall:
- The ST elevation alarm on the Patient Monitor or standalone X2 Measurement Module will not sound when indicated for all chest leads derived using Hexad 12-Lead ECG Monitoring in the Host Monitor
- Classification:
- Class I - Dangerous
Dangerous or defective products that predictably could cause serious health problems or death.
- Status:
- terminated
Product Information
Full Description:
Philips IntelliVue Module Measurement X2 Model: M3002A
Product Codes/Lot Numbers:
Software Revision: K.21.54 or L.00.96
Official Source
Always verify recall information with the official FDA source:
View on FDA.govFDA Recall Number: Z-0854-2016
Related Recalls
Xper Flex Cardio Physiomonitoring System, Software versions 1.5.16.0322 and 1.6.1.0390, Model Numbers 453564241901, 453564241911, 453564483321, 453564483331, 453564621791, 453564621801, 453564669081
Philips Electronics North America
There are 3 issues: 1) Under certain conditions, real time waveforms may not be accurately plotted or displayed. 2) The oxygen saturation (SpO2) numeric value displayed on the device may freeze. 3) In auto cycle mode, if the NIBP communication is lost, the pump would not cycle and the non-invasive blood pressure (NIBP) numeric value displayed on the device would not update.
A problem has been detected in the Philips IntelliVue MX4O that, if it were to occur, could affect the performance of the equipment. The volume of the MX4O speaker in these devices may be diminished or not audible above 4,500 feet when operating in Monitor Mode. These devices may intermittently exhibit Speaker Malfunct INOP messages after the Power On Self-Test. The issue is only apparent when the device is being used in Monitor Mode at an altitude of greater than 4,500 feet.
Automated external defibrillators may not fully meet IPx5 water ingress specification. The device may fail to function should water intrusion occur.